2A65I-GC866995ORVA Sensor

Picosys Sdn Bhd
ORVA Sensor - FCC ID 2A65I-GC866995 - Picosys Sdn Bhd
Click to zoom

Application Details

Equipment Class
DTS - Digital Transmission System
Date of Grant
May 01, 2023
Application Purpose
Original Equipment
Date of Application
May 01, 2023
Equipment Note
ORVA Sensor
Frequency Range
2402.00000000 - 2480.00000000
Company
Picosys Sdn Bhd
Country
Malaysia

Documents & Files

Select a file to view

Users Manual

↗

Attestation Statements

↗
↗

Cover Letter(s)

↗
↗

External Photos

↗

ID Label/Location Info

↗

Internal Photos

↗

Operational Description

↗

RF Exposure Info

↗

Test Report

↗
↗

Test Setup Photos

↗

Document Text

Text extracted from the exhibit documents filed with the FCC. Open a document above to read the original.

Users Manual

User Manual V1.7 Picosys Sdn Bhd Page 1 of 5 ORVA SENSOR Product Description Online Rehab Virtual Assist ant (ORVA) is developed to reduce the barrier-of-e ntry for Physiotherapy by helping patients rehabilitate and recover at home without having to incur costly bills, travel and waiting times. ORVA achieves its goals with high precision sensors along with smartphone, Internet of Things, Cloud and Artificial Intelligence technologies to automate the capturing, transmission, storage and analytics of patient rehabilitation data. ORVA is a wearable device designed for non-intrusive, risk-free, patient movement monitoring. ORVA is placed on the limbs of patients to track and record movement in three dimensions. Coordinates and angles are automatically calculated and recorded, sent to the smartphone via Bluetooth Low Energy (BLE), and uploaded to the cloud for review, analysis and subsequent physician-direction recommendations and actions. General Specifications Model ORVA Version 1.70 Dimension 33mm x 23mm x 12mm Weight 8.0g Operating Period 60 Minutes Charging Period 20 Minutes Charging Power 5.0V, maximum 0.5A (USB charger compatible) Charging Port Micro USB Battery Capacity 3.7V 25mAh Wireless Module NRF52832-QFAA Bluetooth Antenna MOLEX 47948-0001 Specification ARM® Cortex®-M4F 32-bit processor running at 64 MHz 2.4GHz Bluetooth Low Energy v4.2 3-axis gyroscope 3-axis accelerometer 3-axis magnetometer User Manual V1.7 Picosys Sdn Bhd Page 2 of 5 Environmental and Physical Specifications Operating Temperature 0°C to 40°C (32°F to 104°F) Storage Temperature 0°C to 60°C (32°F to 140°F) Operating Humidity 10% to 80% RH non-condensing Storage Humidity 5% to 80% RH non-condensing Safety Information •Keep the device away from water, fire, high humidity or hot environments. •Do not attempt to disassemble, repair, or modify the device. •Dry sanitization/cleaning only with 99+% isopropyl alcohol (IPA). Do not charge the device while using. Connect device with Micro-USB cable to charge the device. Contact the manufacturer if you have any question. Disposal User has the choice to give this product to a competent recycling organization or to return it to the retailer. Transportation Instruction No specific transportation requirement and instruction. User Manual V1.7 Picosys Sdn Bhd Page 3 of 5 User Guide 1.Insert the straps (sold separately) for the ORVA sensors, as shown below. 2.Turn on all ORVA sensors by moving the switch from left to right (CHECK). 3.Enable Bluetooth on your smartphone. ORVA is Connects over Bluetooth Low Energy (BLE) version 4.0 and above automatically. 4.Download the Stay@home Rehab app from Google’s Playstore (ONLY for Android). 5.Launch the Stay@home Rehab app to connect the ORVA sensors. 6.Follow the instructions in the app. When prompted by the app to place the sensors, the USB charging port of the sensor must be pointed toward the user, as shown below. User Manual V1.7 Picosys Sdn Bhd Page 4 of 5 LED Indicators The states of the sensor is represented by three LEDs, as described in the table below. State Red LED Yellow LED Orange LED Disconnected Alternating Red and Yellow Off Connected Blinking Off Off Sensing Off Blinking Off Low battery Blinking Off Charging On Frequently Asked Question (FAQ) Q1. Why is the Red/Yellow LED not lit after device is turned ok? The Red/Yellow LED should be lit when the device is working properly. If the Red/Yellow LED is not lit, please try the following: A: Ensure the device is fully charge. A: Ensure the device is turned on. Q2: Why is the device not connected to Stay@Home Rehab app? The Yellow LED is blinking when the device is in sensing mode, Yellow LED will be off and Red LED will be blinking when it connected to Stay@Home Rehab app. A: Ensure the smartphone’s Bluetooth is turned on. A: Ensure the device is within 10m range of the smartphone. Warning This device complies with Part 15 Class B of the FCC Rules. Operation is subject to the following conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Caution: Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. User Manual V1.7 Picosys Sdn Bhd Page 5 of 5 Note: This equipment complies with Class B digital device, pursuant to part 15 of the FCC Rules, that provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is unlikely to cause harmful interference. This product is designed, tested, and classified for their intended electromagnetic environment. These electromagnetic environment classifications generally refer to the following harmonized definitions: • Class B products are intended for use in residential/domestic environments but may also be used in non-residential/non-domestic environments. Warranty Statement PS (PICOSYS SDN BHD) warrants its products against defects in workmanship and materials for a period of 12 months from the date of sale by PS to its initial customer (regardless of any subsequent sale of the products). This warranty is void if the product is misused, altered, tampered with or is installed or used in a manner that is inconsistent with PS's written recommendations, specifications and/or instructions, or fails to perform due to normal wear and tear. PS is not liable for special, indirect, incidental, consequential or other damages including, but not limited to, loss, damage, personal injury, or any other expense directly or indirectly arising from the use of or inability to use its products either separately or in combination with other pro…

Text truncated - open the document above for the full version.

Attestation Statements

PICOSYS SDN BHD Co. No. 201801012033 (1274049-H) SS-02-26 SKYPOD Square, Persiaran Puchong Jaya Selatan, Pusat Bandar Puchong, 47100 Puchong, Selangor, Malaysia. CFR 47 §2.911 (d)(5) Attestation Date: 02-15-2023 FCC ID: 2A65I-GC866995 To: Federal Communications Commission 7435 Oakland Mills Road Columbia, MD 21046 USA 1. Pisosys Sdn Bhd (“the applicant”) certifies that the equipment for which authorization is sought is not “covered” equipment prohibited from receiving an equipment authorization pursuant to section 2.903 of the FCC rules. 2. Pisosys Sdn Bhd (“the applicant”) certifies that, as of the date of the filing of the application, the applicant is not identified on the Covered List as an entity producing “covered” equipment and is not an affiliate or subsidiary of an entity identified on the Covered List. Sincerely yours, Name:Wong Wei Siung Job Title:Director Company Name: Picosys Sdn Bhd

Attestation Statements

PICOSYS SDN BHD Co. No. 201801012033 (1274049-H) SS-02-26, SKY-POD Square, Persiaran Puchong Jaya Selatan, Pusat Bandar Puchong, 47100 Puchong, Selangor, Malaysia. V02072023 Federal Communications Commission Authorization and Evaluation Division 7435 Oakland Mills Road Columbia, MD 21046 USA Date: 03/24/2023 Ref: Attestation Statement Part 2.911(d)(7) Filing Validity Period: Permanent FCC ID: 2A65I-GC866995 Manufacturer Name PicoSys Sdn Bhd Address SS-02-26 SkyPod Square Persiaran Puchong Jaya Selatan Pusat Bandar Puchong, Puchong 47100, Selangor Malaysia certifies that, as of the dates of this application, Americas Compliance Consulting LLC dba iCertifi “iCertifi” is our designated U.S. agent for service of process for the above referenced FCC ID. We accept to maintain an agent for no less than one year after the grantee has terminated all marketing and importation or the conclusion of any commission-related proceeding involving the equipment. The US Agent for Service of Process is aware of and agrees to comply with the requirements outlined in the FCC Equipment Authorization Program, Report and Order FCC 22-84, and clause § 2.911. iCertifi accepts, as of the date of the filing of the application, the obligation of the designated U.S. agent for service process for the above: Designated U.S. Agent Information: Name Americas Compliance Consulting, LLC dba iCertifi FCC FRN 0033399411 Address 1001 SW Disk Drive, Suite 250 Bend, Oregon 97702 USA Contact Person Paul Preston Telephone (866) 885-4575 Email [email protected] Sincerely, PicoSys Sdn Bhd Americas Compliance Consulting LLC dba iCertifi Wong Wei Siung Paul Preston

Cover Letter(s)

PICOSYS SDN BHD Co. No. 201801012033 (1274049-H) SS-02-26 Skypod Square, Persiaran Puchong Jaya Selatan, Pusat Bandar Puchong, 47100 Puchong, Selangor, Malaysia. Date: 3/28/2022 Federal Communications Commission 7435 Oakland Mills Road Columbia MD 21046 To whom it may concern: I, the undersigned, hereby authorize UL Verification Services Inc. to act on our behalf in all manners relating to application for equipment authorization, including signing of all documents relating to these matters. Any and all acts carried out by UL Verification Services Inc. on our behalf shall have the same effect as acts of our own. I, the undersigned, hereby certify that we are not subject to a denial of federal benefits, that includes FCC benefits, pursuant to Section 5301 of the Anti-Drug Abuse Act of 1988, 21 U.S.C. 853(a). In authorizing UL Verification Services Inc. As our representative, we still recognize that we are responsible to: a) comply with the relevant provisions of the certification program; b) make all necessary arrangements for the conduct of the evaluation, including provision for examining documentation and access to all areas, records (including internal audit reports) and personnel for the purposes of evaluation (e.g. testing, inspection, assessment, surveillance, reassessment) and resolution of complaints; c) Make claims regarding certification only in respect of the scope for which certification has been granted; d) Do not use our product certification in such a manner as to bring the Certification Division into disrepute and not make any statement regarding our product certification which the Certification Division may consider misleading or unauthorized; e) Upon suspension or cancellation of certification, discontinue use of all advertising matter that contains any reference thereto and return any certification documents as required by the Certification Division; f) Use certification only to indicate the products are certified as being in conformity with specified standards; g) Endeavor to ensure that no certificate or report nor any part thereof is used in a misleading manner; h) ensure that any reference to our product certification in communication media such as documents, brochures or advertising, complies with the requirements of the Certification Division; i) keep a record of all complaints made known to us relating to the product’s compliance with requirements of the relevant standard and to make these records available to the when requested; j) take appropriate action with respect to such complaints and any deficiencies found in products or services that affect compliance with the requirements for certification; k) Document the actions taken. This authorization is valid until further written notice from the applicant. Sincerely Yours, Signed by: Company Name: Picosys Sdn. Bhd. Contact Name: Wong Wei Siung Title of Person: Director Address: SS-02-26 Skypod Square, Persiaran Puchong Jaya Selatan, Pusat Bandar Puchong, 47100 Puchong, Selangor, Malaysia.

Cover Letter(s)

PICOSYS SDN BHD Co. No. 201801012033 (1274049-H) SS-02-26, SKYPOD Square, Persiaran Puchong Jaya Selatan, Pusat Bandar Puchong, 47100 Puchong, Selangor, Malaysia. Federal Communications Commission 7435 Oakland Mills Road Columbia MD 21046 Subject: Request for Confidentiality FCC ID: 2A65I-GC866995 To Whom It May Concern: Pursuant to the provisions of Sections 0.457 and 0.459 of the Commission’s rules (47 CFR §§ 0.457, 0.459), we are requesting the Commission to withhold the following attachments as confidential document from public disclosure indefinitely. Schematic Diagram Block Diagram Above mentioned document contains detailed system and equipment description are considered as proprietary information in operation of the equipment. The public disclosure of above documents might be harmful to our company and would give competitor an unfair advantage in the market. It is our understanding that all measurement test reports, FCC ID label format and correspondent during certification review process cannot be granted as confidential documents and that information will be available for public review once the grant of equipment authorization is issued. Best Regards Company Name: Picosys Sdn. Bhd. Contact Name: Wong Wei Siung Title of Person: Director Address: SS-02-26 Skypod Square, Persiaran Puchong Jaya Selatan, Pusat Bandar Puchong, 47100 Puchong, Selangor, Malaysia. Date: 3/28/2022

External Photos

PICOSYS SDN BHD Page 1 of 3 EXT ER NA L PH O T O G R A PH S FCC ID: 2A65I-GC866995 Brand : Picosys Model : ORVA Photo 1: ORVA Sensor Label Photo 2: Top View PICOSYS SDN BHD Page 2 of 3 EXT ER NA L PH O T O G R A PH S FCC ID: 2A65I-GC866995 Photo 3: Rear View Photo 3: Side View (Right) PICOSYS SDN BHD Page 3 of 3 EXT ER NA L PH O T O G R A PH S FCC ID: 2A65I-GC866995 Photo 4: Side View (Left)

ID Label/Location Info

PICOSYS SDN BHD Page 1 of 3 EXT ER NA L PH O T O G R A PH S FCC ID: 2A65I-GC866995 Brand : Picosys Model : ORVA Photo 1: ORVA Sensor Label Photo 2: Top View

Internal Photos

PICOSYS SDN BHD Page 1 of 4 INT ER NA L PH OT O G R A PH S FCC ID: 2A65I-GC866995 Brand : Picosys Model : ORVA Top View Bottom View PICOSYS SDN BHD Page 2 of 4 INT ER NA L PH OT O G R A PH S FCC ID: 2A65I-GC866995 Zoomed Top View with Antenna Processor Antenna PICOSYS SDN BHD Page 3 of 4 INT ER NA L PH OT O G R A PH S FCC ID: 2A65I-GC866995 Zoomed Bottom View PICOSYS SDN BHD Page 4 of 4 INT ER NA L PH OT O G R A PH S FCC ID: 2A65I-GC866995 Complete View

Operational Description

PICOSYS SDN BHD OPER A T IONA L D E SC R IP T ION Brand : Picosys Model : ORVA Operational Description Product Name Online Rehab Virtual Assistance Brand Name PICOSYS Model Name ORVA Operating Frequency Bluetooth LE : 2402 MHz to 2480 MHz Type of Modulation Gaussian Frequency Shift Keying (GFSK) No of Channel 40 Maximum Tune Up Power 4 dB ± 1 dB Data Rate 1 Mbps , 2 Mbps Module Information Brand Nordic Semiconductor Model nRF52832 Modulation GFSK Functional Bluetooth 5.0 Additional Information Transmitter Circuit Wireless Technology Brand Name Model Name Antenna Type Maximum Gain Surface Mount BLE Molex 0479480001 SMT MID Chip Antenna 3.7 dBi

RF Exposure Info

SIRIM QAS International Sdn. Bhd. (Company No.: 199601037981 (410334-X)) No. 1, Persiaran Dato’ Menteri, Section 2, P.O. Box 7035, 40700 Shah Alam, Selangor Darul Ehsan, Malaysia Tel: 603-55446169 www.sirim-qas.com.my SIRIM QAS International Sdn. Bhd. (Company No.: 199601037981 (410334-X)) No. 1, Persiaran Dato’ Menteri, Section 2, P.O. Box 7035, 40700 Shah Alam, Selangor Darul Ehsan, Malaysia Tel: 603-55446169 www.sirim-qas.com.my SAR EXEMPTION EVALUATION REPORT REPORT NO.: 2022RE0156 (FCC ID 2A65I-GC866995) PAGE: 1 OF 6 This Test Report refers only to samples submitted by the applicant to SIRIM QAS International Sdn. Bhd. and tested by SIRIM QAS International Sdn. Bhd. This Test Report shall not be reproduced, except in full and shall not be used for any purpose by any means or forms (including but not limited to advertising purposes) without written approval from the Head of Quality, Occupational Safety and Health & Environment (QOSHE), SIRIM QAS International Sdn. Bhd. Please refer to the last page of this Test Report for Conditions Relating to the Use of Test Report. Approved Signatories, ..................................... (ZUL HJ. JAAFAR) Group Leader RF Laboratory .................................................. (ZARISMAIL ABD RAHMAN) Head RF & EMC Testing Section Testing Services Department THIS TEST REPORT IS ISSUED IN SECURED PDF SOFTCOPY Applicant : PICOSYS SDN BHD (1274049-H) SS-02-26 Sky-Pod Square Persiaran Puchong Jaya Selatan Pusat Bandar Puchong 47100 Puchong Selangor, Malaysia Manufacturer : PICOSYS SDN BHD (1274049-H) Product : Online Rehab Virtual Assistance (ORVA) Reference Standard / Method of test : KDB 447498 D01 General RF Exposure Guidance v06 Description of Sample : Brand Name : PICOSYS Model : ORVA Issued Date : 09 MARCH 2023 REPORT NO.: 2022RE0156 (FCC ID: 2A65I-GC866995) PAGE: 2 OF 6 This Test Report refers only to samples submitted by the applicant to SIRIM QAS International Sdn. Bhd. and tested by SIRIM QAS International Sdn. Bhd. This Test Report shall not be reproduced, except in full and shall not be used for any purpose by any means or forms (including but not limited to advertising purposes) without written approval from the Head of Quality, Occupational Safety and Health & Environment (QOSHE), SIRIM QAS International Sdn. Bhd. Please refer to the last page of this Test Report for Conditions Relating to the Use of Test Report 1.0 EUT PROFILE The following data is based on the information provided by applicant. Type of Product : Combined Equipment Plug in radio Device  Stand Alone Mode of :  Continuous Transmission Burst Intermittent Signal Transmission Techniques : FHSS  Non FHSS Type of Modulation : Gaussian Frequency Shift Keying (GFSK) Operating Voltage : 3 Vdc RF Power Output : 4 dBm EUT Antenna Gain : 3.7 dBi Operating Frequency Range : Bluetooth Low Energy (BLE) 2402 MHz – 2480 MHz Software : J-Link RTT Viewer V7.80a REPORT NO.: 2022RE0156 (FCC ID: 2A65I-GC866995) PAGE: 3 OF 6 This Test Report refers only to samples submitted by the applicant to SIRIM QAS International Sdn. Bhd. and tested by SIRIM QAS International Sdn. Bhd. This Test Report shall not be reproduced, except in full and shall not be used for any purpose by any means or forms (including but not limited to advertising purposes) without written approval from the Head of Quality, Occupational Safety and Health & Environment (QOSHE), SIRIM QAS International Sdn. Bhd. Please refer to the last page of this Test Report for Conditions Relating to the Use of Test Report 2.0 Specific Absorb Rate (SAR) Evaluation The 1-g and 10-g SAR test exclusion thresholds for 100 MHz to 6 GHz at test separation distances ≤ 50 mm and the formula below: √f(GHz) * (max. power of channel, including tune-up tolerance, mW) (min. test separation distance, mm) · where • f(GHz) is the RF channel transmit frequency in GHz • Power and distance are rounded to the nearest mW • The result is rounded to one decimal place for comparison • The values 3.0 and 7.5 are referred to as numeric thresholds SAR Test Exclusion Threshold (STET) : i) STET ≤ 3.0 for 1-g SAR ii) STET ≤ 7.5 for 10-g extremity SAR The test exclusions are applicable only when the minimum test separation distance is ≤ 50 mm and for transmission frequencies between 100 MHz and 6 GHz. REPORT NO.: 2022RE0156 (FCC ID: 2A65I-GC866995) PAGE: 4 OF 6 This Test Report refers only to samples submitted by the applicant to SIRIM QAS International Sdn. Bhd. and tested by SIRIM QAS International Sdn. Bhd. This Test Report shall not be reproduced, except in full and shall not be used for any purpose by any means or forms (including but not limited to advertising purposes) without written approval from the Head of Quality, Occupational Safety and Health & Environment (QOSHE), SIRIM QAS International Sdn. Bhd. Please refer to the last page of this Test Report for Conditions Relating to the Use of Test Report Antenna to Front (Refer Figure 1.0) Frequency (GHz) Output Power Separation distance (mm) Calculated Threshold Exclusion Threshold (1-g SAR) SAR Test Exclusion dBm mW 2.402 5 3.1622 4 0.9802 3 YES 2.440 5 3.1622 4 0.9879 3 YES 2.480 5 3.1622 4 0.9959 3 YES Note: A distance of 5 mm is applied to determine the SAR test exclusion due to the minimum test separation distance is < 5mm,. Figure 1.0 : Antenna to Front 4 mm REPORT NO.: 2022RE0156 (FCC ID: 2A65I-GC866995) PAGE: 5 OF 6 This Test Report refers only to samples submitted by the applicant to SIRIM QAS International Sdn. Bhd. and tested by SIRIM QAS International Sdn. Bhd. This Test Report shall not be reproduced, except in full and shall not be used for any purpose by any means or forms (including but not limited to advertising purposes) without written approval from the Head of Quality, Occupational Safety and Health & Environment (QOSHE), SIRIM QAS International Sdn. Bhd. Please refer to the last page of this Test Report for Conditions Relating to the Use of Test Report Antenna to back (Refer Figure 2.0) …

Text truncated - open the document above for the full version.

Test Report

SIRIM QAS International Sdn. Bhd. (Company No.: 199601037981 (410334-X)) No. 1, Persiaran Dato’ Menteri, Section 2, P.O. Box 7035, 40700 Shah Alam, Selangor Darul Ehsan, Malaysia Tel: 603-55446169 www.sirim-qas.com.my SIRIM QAS International Sdn. Bhd. (Company No.: 199601037981 (410334-X)) No. 1, Persiaran Dato’ Menteri, Section 2, P.O. Box 7035, 40700 Shah Alam, Selangor Darul Ehsan, Malaysia Tel: 603-55446169 www.sirim-qas.com.my TEST REPORT REPORT NO.: 2022RE0119 (FCC ID:2A65I-GC866995) PAGE: 1 OF 32 This Test Report refers only to samples submitted by the applicant to SIRIM QAS International Sdn. Bhd. and tested by SIRIM QAS International Sdn. Bhd. This Test Report shall not be reproduced, except in full and shall not be used for any purpose by any means or forms (including but not limited to advertising purposes) without written approval from the Head of Quality, Occupational Safety and Health & Environment (QOSHE), SIRIM QAS International Sdn. Bhd. Please refer to the last page of this Test Report for Conditions Relating to the Use of Test Report. Approved Signatories, ..................................... (ZUL HJ. JAAFAR) Group Leader RF Laboratory .................................................. (ZARISMAI L ABD RAHMAN) Head RF & EMC Testing Section Testing Services Department THIS TEST REPORT IS ISSUED IN SECURED PDF SOFTCOPY Applicant : PICOSYS SDN BHD ( 1274049-H) SS-02-26 Sky-Pod Square Persiaran Puchong Jaya Selatan Pusat Bandar Puchong 47100 Puchong Selangor, Malaysia Manufacturer : PICOSYS SDN BHD (1274049-H) Product : Online Rehab Virtual Assistance Reference Standard / Method of test : FCC Subpart C 15.247 / [1]ANSI C63.10-2013 [2]KDB 558074 v05r02 Description of Sample Date Received of Complete Application Job No. Overall Test Result Issued Date : Brand Name : PICOSYS Model : ORVA : 07 FEBRUARY 2023 : J20221410015 : The test results for the submitted test samples as described in this test report complied with the requirements of the above reference standard. : 09 MARCH 2023 REPORT NO.: 2022RE0119 (FCC ID:2A65I-GC866995) PAGE: 2 OF 35 This Test Report refers only to samples submitted by the applicant to SIRIM QAS International Sdn. Bhd. and tested by SIRIM QAS International Sdn. Bhd. This Test Report shall not be reproduced, except in full and shall not be used for any purpose by any means or forms (including but not limited to advertising purposes) without written approval from the Head of Quality, Occupational Safety and Health & Environment (QOSHE), SIRIM QAS International Sdn. Bhd. Please refer to the last page of this Test Report for Conditions Relating to the Use of Test Report TABLE OF CONTENTS NOTES: 1.All tests were carried out at ambient temperature of +15ºC to + 35ºC and relative humidity of 20% to 75%. Ambient temperature (minimum) : 24 ºC Ambient temperature (maximum) : 25 ºC Relative Humidity (minimum) : 49 % Relative Humidity (maximum) : 50 % 2.All the reported uncertainty is calculated using coverage factor k= 1.96, which gives a level of confidence of approximately 95%. 3.Test Facility: FCC Designation No: MY0005 SIRIM QAS International Sdn. Bhd. Block 11 Sirim Complex No. 1 Persiaran Dato’ Menteri, Section 2, 40700 Shah Alam, Selangor Darul Ehsan, Malaysia 4.Date of test sample(s) received : a)1 st submission: 07 February 2023 5.Date of sample(s) tested: a)1 st test: 07 February 2023 Tested by : Hafizah Zainal Abiddin Reviewed by : Zul Hj. Jaafar No. Contents Page 1.0 SUMMARY OF RESULTS 3 2.0 INTRODUCTION 4 3.0 EQUIPMENT UNDER TEST DESCRIPTION 5 3.0 TEST PROCEDURES AND RESULTS 6 4.0 TEST INSTRUMENT 34 CONDITIONS RELATING TO THE USE OF SIRIM QQS INTERNATIONAL TEST REPORT 35 REPORT NO.: 2022RE0119 (FCC ID:2A65I-GC866995) PAGE: 3 OF 35 This Test Report refers only to samples submitted by the applicant to SIRIM QAS International Sdn. Bhd. and tested by SIRIM QAS International Sdn. Bhd. This Test Report shall not be reproduced, except in full and shall not be used for any purpose by any means or forms (including but not limited to advertising purposes) without written approval from the Head of Quality, Occupational Safety and Health & Environment (QOSHE), SIRIM QAS International Sdn. Bhd. Please refer to the last page of this Test Report for Conditions Relating to the Use of Test Report 1.0 SUMMARY OF RESULTS The summary of the test results as follows: FCC Part Section Test Description Result 15.247 (a) (2) Occupied Bandwidth PASS 15.247 (b) (3) Output Power PASS 15.247 (e) Power Spectral Density PASS 15.247 (d) Conducted Band Edge PASS Conducted Out of Band Emission 15.207 AC Conducted Emission *NOT APPLICABLE 15.209 Radiated Spurious Emission PASS Radiated Band Edge *Note: a) Not Applicable for battery operated product REPORT NO.: 2022RE0119 (FCC ID:2A65I-GC866995) PAGE: 4 OF 35 This Test Report refers only to samples submitted by the applicant to SIRIM QAS International Sdn. Bhd. and tested by SIRIM QAS International Sdn. Bhd. This Test Report shall not be reproduced, except in full and shall not be used for any purpose by any means or forms (including but not limited to advertising purposes) without written approval from the Head of Quality, Occupational Safety and Health & Environment (QOSHE), SIRIM QAS International Sdn. Bhd. Please refer to the last page of this Test Report for Conditions Relating to the Use of Test Report 3.0 EQUIPMENT UNDER TEST DESCRIPTION The following data is based on the information provided by applicant. Type of Product : Combined Equipment Plug in radio Device  Stand Alone Mode of :  Continuous Transmission Burst Intermittent Signal transmission techniques : Frequency Hopping Spread Spectrum (FHSS)  Non FHSS Type of Modulation : Gaussian Frequency Shift Keying (GFSK) Operating Voltage : 3 Vdc RF Output Power : 4 dBm EUT Antenna Gain : 3.7 dBi Operating Frequency Range : Bluetooth Low Energy (BLE) 2402 MHz – 2480 MHz Software : J-Link RTT Viewer V7.80a REPORT NO.: 2022RE0119 (FCC ID:2A65I-GC866995) PAGE: 5 OF 35 This Test R…

Text truncated - open the document above for the full version.

Test Report

This document was generated on 01/09/2023 PLEASE CHECK WWW.MOLEX.COM FOR LATEST PART INFORMATION Part Number: 0479480001 Status: Active Overview:Internet of Things (IoT) Antennas: Wi-Fi, Bluetooth, Zigbee Description:2.4GHz SMT MID Chip Antenna Documents: 3D Model (PDF)Application Specification AS-47948-001-001 (PDF) HFSS ModelPackaging Specification PK-47948-001-001 (PDF) 3D ModelDatasheet (PDF) Drawing (PDF)RoHS Certificate of Compliance (PDF) Product Specification PS-47948-001-001 (PDF) General Product FamilyAntennas Series47948 Component TypeSurface Mount Device Antenna FunctionSignal OverviewInternet of Things (IoT) Antennas: Wi-Fi, Bluetooth, Zigbee Product Name2.4GHz SMT MID Chip ProtocolBLE, BT, Thread, Wi-Fi, Wireless Hart, Zigbee TaxonomyAntennas TypeWi-Fi Antenna UPC884982642621 Physical Cable LengthN/A Length3.00mm Mounting StyleSurface Mount Net Weight0.042/g Packaging TypeEmbossed Tape on Reel PolarizationLinear Radiation PatternOmnidirectional Thickness4.00mm Width3.00mm Electrical Band#1 F_End (MHz)2483.5 Band#1 F_Start (MHz)2400 Electrical ConnectivitySMD Peak Gain (dBi)3.7 @ 2.4 GHz Return Loss - S11 (dB)< -6 Total Efficiency>70% Material Info Reference - Drawing Numbers Application SpecificationAS-47948-001-001 Packaging SpecificationPK-47948-001-001 Product SpecificationPS-47948-001-001 Sales DrawingSD-47948-001-001 Series image - Reference only EU ELV Not Relevant EU RoHSChina RoHS Compliant REACH SVHC Not Contained Per - D(2022)4187-DC (10 June 2022) Halogen-Free Status Low-Halogen For more information, please visit Contact US China ROHSGreen Image ELVNot Relevant RoHS PhthalatesNot Contained Search Parts in this Series 47948 Series This document was generated on 01/09/2023 PLEASE CHECK WWW.MOLEX.COM FOR LATEST PART INFORMATION

Test Setup Photos

PICOSYS SDN BHD Page 1 of 4 T EST SET UP FCC ID: 2A65I-GC866995 Brand : Picosys Model : ORVA Photo 1: Measurement below 1 GHz PICOSYS SDN BHD Page 2 of 4 T EST SET UP FCC ID: 2A65I-GC866995 Photo 2: Measurement above 1 GHz ( 1GHz to 18 GHz) PICOSYS SDN BHD Page 3 of 4 T EST SET UP FCC ID: 2A65I-GC866995 Photo 3: Measurement above 1 GHz (18 GHz to 40 GHz) PICOSYS SDN BHD Page 4 of 4 T EST SET UP FCC ID: 2A65I-GC866995 Photo 4: Conducted Measurement

Contact Information

Applicant

Wei Siung Wong
[email protected]+60122965908Fax: +60386029830

Technical Contact

Picosys Sdn BhdWei Siung Wong
[email protected]+603-86029830

SS-02-26 Skypod Square, Persiaran Puchong Jaya Sel · Puchong · Malaysia

Test Firm

SIRIM QAS International Sdn BhdNorhayati Md. Jonit
[email protected]+603-55446172

Technical Specifications

#Rule PartsFrequency RangePower Output
115C2.40 GHz - 2.48 GHz2.80 mW
Confidentiality
Long Term
Grant Notes
Output power listed is conducted.