Text extracted from the exhibit documents filed with the FCC. Open a document above to read the original.
Device Quick Start Guide Device Diagram....................................................................... Accessories............................................................................... Download the App.................................................................. Pair the Device......................................................................... Preparing to Submit a Test...................................................... Submitting a Test...................................................................... Client Text Notifications........................................................... Device Operating Information............................................... Troubleshooting........................................................................ Intended Use............................................................................. Device Recalibration............................................................... Disclaimers................................................................................. Table of Contents 1 2 3 3 4 5 6 7 8 9 9 9 1 FCC ID: X3ZBTMOD4 3600XXXX Soberlink Connect Device Diagram Camera Mouthpiece port Power button USB-C charging port Status light FCC ID: 2A8ZY-3900 3900XXXX Soberlink Connect ST 2 Accessories Device Case Insert the large end of the cable into the charging base and insert the USB-C (small end) into the port on the bottom of the Soberlink Device. Plug the charging base into an electrical outlet. Device Charger Mouthpiece To protect the Device when not in use, keep it stored in the Device Case. The Mouthpiece can be stored inside the Device Case. Insert the small, flanged end of the Mouthpiece into the Device. Note: Always use a Soberlink Mouthpiece when submitting a test. 3 Download the App Search Soberlink Connect and download the App. Make sure Bluetooth is enabled on the iOS or Android device. Open the Soberlink Connect App and follow the screen prompts. Pair New Device Note: In the Apple App store make sure to download the Soberlink Connect App not the SLBLUE App. 4 Preparing to Submit a Test This document, and any other provided training material, should be fully understood by the user before submitting a test. To ensure proper device function keep it fully charged. To avoid damage or injury caused by the product only use accessories provided by the manufacturer. Find an area free from alcohol, cleaners, solvent vapors, and smoke before submitting a test. Use the Device within the operating temperature range (32°F -113°F). Breathe normally before submitting a test. Remove sunglasses, hats, or other objects covering the face. Wait 20 minutes after eating, drinking, or smoking before submitting a test. Failure to wait may result in inaccurate test results. Rinsing out the mouth with water does not act as a substitute for the 20-minute waiting period. 12 6 5 4 3 2 1 9 8 7 10 11 2 4 3 2 1 WAIT 20 MINUTES NO EATING NO DRINKING NO SMOKING The Device must be activated before submitting a test. To activate call Soberlink at 714-975-7200 or contact your Treatment Provider. 5 Submitting a Test Open the Soberlink Connect App. The App should remain open for the duration of testing. Press the Power button to turn on the Soberlink Device. Follow the prompts on the App, and wait for the green light to flash before submitting a test. When prompted to blow, look directly into the camera, breathe in deeply and deliver a firm, continuous breath for 4 seconds. The Device will click when the test is complete. Note: Do not hold the mouthpiece or obstruct the camera. The test results will take approximately 30 seconds to display on the App and send to the Sober Sky Web Portal. Once the test has successfully sent, close the App. Note: After a positive BAC (Breath Alcohol Concentration) is received, wait 15 minutes before submitting another test. 6 Client Text Notifications Once a schedule is enabled on the Sober Sky Web Portal, Soberlink’s automated system will send text message reminders directly to the Client’s phone. Reminder texts are sent before and during the test window. Missed notification after test window closes Notification of retesting requirement 30574 REMINDER TEXT: This is your 8:00 AM PST test reminder. 30574 MISSED TEST: You have missed 1 or more tests. Send a test now. 30574 RETEST REQUIRED: A test is required at 8:18 AM PST. Note: Upon entry of a mobile number on the Sober Sky Web Portal, the Client will receive a text message and must respond as instructed in order to Opt-in to notifications. 30574 You have requested to receive Soberlink Alerts. Reply with YES to start. 7 Device Operating Information Dimensions4.75” x 2.5” x 1.25” Weight5.3 oz. (150 grams) Sensor TechnologyDeep lung, Dart fuel-cell sensor Battery TypeLithium-ion rechargeable battery, holds 10+ days Power SupplyUSB-C charger, Output: 5V 1A Breath Sample Duration4 seconds Operating Temperature32°F - 113°F (0°C - 45°C) Relative Humidity80% non-condensing up to 87 degrees F (31°C), decreasing linearly to 50% non-condensing at 104 degrees F (40 C°) Min. Ambient Air Pressure0.74 atm Pollution Degree2 Sensor Accuracy.+/- .005 BAC Detection Range.000 - .400% BAC CalibrationEvery 1,500 tests CameraColor, digital, infrared technology Test DataBAC, time of test, and photo ComponentsBreathalyzer, mouthpieces, charger, case ManufacturedISO 9001, 14001, 13485, AS9100, and OHSAS 18001 Warranty1 year limited Technical Specifications The user does not need to perform any tests at start up, the Device performs a self-test when it is powered on and prior to a test submission. The self-test is meant to verify the proper operation of the test storage memory, alcohol fuel cell, and imager. Failure of any of these components indicates that the Device is unable to properly record test data. In this case, an error code is displayed to the user and testing will not proceed. Safety There are no anticipated risks or discomforts utilizing the Device, however, if the De…
Text truncated - open the document above for the full version.
(Insert Company Letterhead) Letter of Authorization and Anti- Drug Statement Applicant Name: Soberlink Healthcare Address: 7711 Center Ave, Ste 400 Huntington Beach CA 92647 Product Name: Connect Model Number: Connect ST FCC ID: We authorize: CETECOM Inc., to act on our behalf on all matters concerning the above named equipment. We declare that CETECOM Inc. is allowed to forward all information related to the approval project to the Federal Communications Commission and discuss any issues concerning the approval application. The applicant must certify that neither the applicant nor any party to the application is subject to a denial of Federal benefits, that include FCC benefits, pursuant to Section 5301 of the Anti-Drug Abuse Act of 1988, 21 U.S.C. § 862 because of a conviction for possession or distribution of a controlled substance. See 47 CFR 1.2002(b) for the definition of a "party" for these purposes. Does the applicant or authorized agent so certify? ☒ Yes I certify that I am authorized to sign this application. All of the statements herein and the exhibits attached hereto, are true and correct to the best of my knowledge and belief. In accepting a Grant of Equipment Authorization as a result of the representations made in this application, the applicant is responsible for (1) labeling the equipment with the exact FCC ID specified in this application, (2) compliance statement labeling pursuant to the applicable rules, and (3) compliance of the equipment with the applicable technical rules. If the applicant is not the actual manufacturer of the equipment, appropriate arrangements have been made with the manufacturer to ensure that production units of this equipment will continue to comply with the FCC's technical requirements. Authorizing an agent to sign this application, is done solely at the applicant's discretion; however, the applicant remains responsible for all statements in this application. If an agent has signed this application on behalf of the applicant, a written letter of authorization which includes information to enable the agent to respond to the above section 5301 (Anti-Drug Abuse) Certification statement has been provided by the applicant. It is understood that the letter of authorization must be submitted to the FCC upon request, and that the FCC reserves the right to contact the applicant directly at any time. Signature of Authorized Person Filing: Title: Dan Rhodes, Vice President Engineering Date: 10/4/2022
Company LOGO Federal Communications Commission Equipment Authorization Division 7435 Oakland Mills Road Columbia, MD 21048 Date: Subject: Request for Permanent Confidentiality FCC ID: To Whom It May Concern: In accordance with §0.457(d) and §0.459 of CFR 47, and KDB 726920, Full Company Name hereby requests permanent confidentiality for documents filed under the following exhibit types: 1. Block Diagram 2. Schematic Diagram 3. Part List 4. Operational Description These items contain detailed system and equipment description and related information about the product which Soberlink Healthcare considers to be proprietary, confidential and a custom design which otherwise would not be released to the general public. Since this design is a basis from which future technological products will evolve, Soberlink Healthcare also feels that this information would be of benefit to its competitors, and that the disclosure of the information in these exhibits would give our competitors an unfair advantage in the market. We are aware that exhibit types cover letters, attestations, label document, test reports, RF exposure information (SAR and MPE reports), cannot be requested to be classified confidential and are thus subject to publication, as will be documents of all other exhibits types for which we have not requested a confidentiality classification. Best Regards, _______________________________ Signature Name: Dan Rhodes Title: Vice President, Engineering Company Name: Soberlink Healthcare Address: 7711 Center Ave, Suite 400, Huntington Beach, CA 92647
FCCID2A8ZY-3900 External Photos FCCID2A8ZY-3900 FCCID2A8ZY-3900 FCCID2A8ZY-3900 FCCID2A8ZY-3900 FCCID2A8ZY-3900 FCCID2A8ZY-3900
Background pantone 426c Logo/Font White
FCCID2A8ZY-3900 Internal Photos FCCID2A8ZY-3900 FCCID2A8ZY-3900 FCCID2A8ZY-3900 FCCID2A8ZY-3900 FCCID2A8ZY-3900 FCCID2A8ZY-3900 FCCID2A8ZY-3900 FCCID2A8ZY-3900 RF Antenna RF chip
A2LA Accredited CETECOM Inc. 411 Dixon Landing Road Milpitas, CA 95035 U.S.A. Phone: + 1 (408) 586 6200 Fax: + 1 (408) 586 6299 E-mail: [email protected] http://www.cetecom.com CETECOM Inc. is a Delaware Corporation with Corporation number: 2905571 © Copyright by CETECOM Radio Frequency Exposure Evaluation Report FOR: Soberlink Healthcare Brand: Connect ST Marketing Name Connect Model Name Connect ST Product Description Breathalyzer FCC ID: 2A8ZY-3900 Applied Rules and Standards: CFR 47 Part 2.1093 FCC KDB 447498 D01 General RF Exposure Guidance v06 Test Report #: SAR_EX_SOBER_012_22001_FCC_Rev2 DATE: 12/6/2022 Test Report #: SAR_EX_SOBER_012_22001_FCC_Rev2 FCC ID: 2A8ZY-3900 Date of Report: 12/6/2022 Page 2 of 8 Contents 1 ASSESSMENT ....................................................................................................................................................... 3 2 ADMINISTRATIVE DATA ....................................................................................................................................... 4 2.1 IDENTIFICATION OF THE TESTING LABORATORY ISSUING THE TEST REPORT ........................................................................... 4 2.2 IDENTIFICATION OF THE CLIENT ................................................................................................................................... 4 2.3 IDENTIFICATION OF THE MANUFACTURER ...................................................................................................................... 4 3 EQUIPMENT UNDER ASSESSMENT ....................................................................................................................... 5 4 FCC EXEMPTION LIMITS FOR ROUTINE EVALUATION ........................................................................................... 6 4.1 FCC SAR TEST EXCLUSIONS ARE SET BY KDB 447498 D01 GENERAL RF EXPOSURE GUIDANCE V06 ....................................... 6 5 SAR EXCLUSION EVALUATION .............................................................................................................................. 7 5.1 STANDALONE .......................................................................................................................................................... 7 6 REVISION HISTORY .............................................................................................................................................. 8 Test Report #: SAR_EX_SOBER_012_22001_FCC_Rev2 FCC ID: 2A8ZY-3900 Date of Report: 12/6/2022 Page 3 of 8 1 Assessment The following device meets the limits of general population uncontrolled exposure specified in CFR 47 Part 2.1093 according to SAR evaluation exclusion requirements specified in FCC regulation as listed in KDB 447498, as it has been evaluated against the standards mentioned above under this section. Responsible for Testing Laboratory: 12/6/2022 Compliance Stoecker, Arndt (Director of Regulatory Services) Date Section Name Signature Responsible for the Report: 12/6/2022 Compliance Ghanma, Issa (EMC Engineer) Date Section Name Signature The test results of this test report relate exclusively to the test item specified in Section3. CETECOM Inc. USA does not assume responsibility for any conclusions and generalizations drawn from the test results with regard to other specimens or samples of the type of the equipment represented by the test item. The test report may only be reproduced or published in full. Reproduction or publication of extracts from the report requires the prior written approval of CETECOM Inc. USA. Test Report #: SAR_EX_SOBER_012_22001_FCC_Rev2 FCC ID: 2A8ZY-3900 Date of Report: 12/6/2022 Page 4 of 8 2 Administrative Data 2.1 Identification of the Testing Laboratory Issuing the Test Report Company Name: CETECOM Inc. Department: Compliance Street Address: 411 Dixon Landing Road City/Zip Code Milpitas, CA 95035 Country USA Telephone: +1 (408) 586 6200 Fax: +1 (408) 586 6299 Director of Regulatory Services: Stoecker, Arndt Responsible Project Manager: Sivaraman, Sangeetha 2.2 Identification of the Client Applicant’s Name: Soberlink Healthcare Street Address: 7711 Center Ave, Ste 400 City/Zip Code Huntington Beach, CA 92647 Country USA 2.3 Identification of the Manufacturer Applicant’s Name: ----Same as Client ---------------------------- Street Address: ----Same as Client ---------------------------- City/Zip Code ----Same as Client ---------------------------- Country ----Same as Client ---------------------------- Test Report #: SAR_EX_SOBER_012_22001_FCC_Rev2 FCC ID: 2A8ZY-3900 Date of Report: 12/6/2022 Page 5 of 8 3 Equipment under Assessment Brand: Connect ST Model No: Connect ST Marketing name: Connect FCC ID: 2A8ZY-3900 Hardware Version: 0001 Software Version: 18.14 Minimum distance of antenna or radiating parts to user 5mm Radios included in the device: Bluetooth Low Energy 4.2: Nordic Chipset name/number: nRF52832 Modulation: GFSK Antenna Information as declared: PCB trace; Gain: 3.63 dBi Co-located Transmitters/ Antennas: Yes_______No Exposure Category: Occupational/ Controlled General Population/ Uncontrolled Device Category: Fixed InstallationMobile PortableMixed Mobile and Portable Power Supply/ Rated Operating Voltage Range: Fixed battery: Low 4.5 V DC, Nominal 5 V DC, High 5.5 V DC Operating Temperature Range: Low 0⁰ C, Nominal 23⁰ C, High +50⁰ C Sample Revision: Prototype Unit;Production Unit;Pre-Production Unit; EUT Dimensions [mm]: 103 x 70 x 156 EUT Diameter: < 60 cm Other______ ___ Note: The information of the EUT specifications in the table above are provided by the client. Test Report #: SAR_EX_SOBER_012_22001_FCC_Rev2 FCC ID: 2A8ZY-3900 Date of Report: 12/6/2022 Page 6 of 8 4 FCC Exemption Limits for Routine Evaluation 4.1 FCC SAR test exclusions are set by KDB 447498 D01 General RF Exposure Guidance v06 KDB 447498 Section: 4.3.1. Standalone SAR test exclusion considerations a) For 100 MHz to 6 GHz and test separation distan…
Text truncated - open the document above for the full version.
A2LA Accredited CETECOM Inc. 411 Dixon Landing Road Milpitas, CA 95035 U.S.A. Phone: + 1 (408) 586 6200 Fax: + 1 (408) 586 6299 E-mail: [email protected] http://www.cetecom.com CETECOM Inc. is a Delaware Corporation with Corporation number: 2905571 © Copyright by CETECOM FCC / ISED Test Report For: Soberlink Healthcare Brand: Connect ST Model Name/Number: Connect ST Marketing Name: Connect Product Description: Breathalyzer FCC ID: 2A8ZY-3900 Applied Rules and Standards: 47 CFR Part 15.247 (DTS) REPORT #: EMC_SOBER-012-22001_FCC_15.247_BLE_DTS_Rev2 DATE: 12/6/2022 Test Report #: EMC_SOBER-012-22001_FCC_15.247_BLE_DTS_Rev2 FCC ID: 2A8ZY-3900 Date of Report 12/6/2022 Page 2 of 82 TABLE OF CONTENTS 1 ASSESSMENT .............................................................................................................................................................. 3 2 ADMINISTRATIVE DATA .............................................................................................................................................. 4 2.1 IDENTIFICATION OF THE TESTING LABORATORY ISSUING THE EMC TEST REPORT ............................................................................. 4 2.2 IDENTIFICATION OF THE CLIENT ............................................................................................................................................. 4 2.3 IDENTIFICATION OF THE MANUFACTURER ................................................................................................................................ 4 3 EQUIPMENT UNDER TEST (EUT) .................................................................................................................................. 5 3.1 EUT SPECIFICATIONS .......................................................................................................................................................... 5 3.2 EUT SAMPLE DETAILS.......................................................................................................................................................... 5 3.3 ACCESSORY EQUIPMENT (AE) DETAILS.................................................................................................................................... 6 3.4 TEST SAMPLE CONFIGURATION ............................................................................................................................................. 6 3.5 MODE OF OPERATION DETAILS .............................................................................................................................................. 6 3.6 JUSTIFICATION FOR WORST CASE MODE OF OPERATION ............................................................................................................ 7 4 SUBJECT OF INVESTIGATION ....................................................................................................................................... 8 5 MEASUREMENT RESULTS SUMMARY ......................................................................................................................... 9 6 MEASUREMENT UNCERTAINTY ................................................................................................................................ 10 6.1 ENVIRONMENTAL CONDITIONS DURING TESTING: ................................................................................................................... 10 6.2 DATES OF TESTING:........................................................................................................................................................... 10 7 MEASUREMENT PROCEDURES .................................................................................................................................. 11 7.1 RADIATED MEASUREMENT ................................................................................................................................................. 11 7.2 RF CONDUCTED MEASUREMENT PROCEDURE ........................................................................................................................ 13 7.3 POWER LINE CONDUCTED MEASUREMENT PROCEDURE ........................................................................................................... 13 8 TEST RESULT DATA ................................................................................................................................................... 14 8.1 DUTY CYCLE .................................................................................................................................................................... 14 8.2 EMISSION BANDWIDTH 6DB AND 99% OCCUPIED BANDWIDTH ................................................................................................ 17 8.3 MAXIMUM PEAK CONDUCTED OUTPUT POWER ....................................................................................................................... 31 8.4 POWER SPECTRAL DENSITY................................................................................................................................................. 38 8.5 BAND EDGE COMPLIANCE .................................................................................................................................................. 45 8.6 FREQUENCY STABILITY ....................................................................................................................................................... 52 8.7 RADIATED TRANSMITTER SPURIOUS EMISSIONS AND RESTRICTED BANDS .................................................................................... 62 8.8 AC POWER LINE CONDUCTED EMISSIONS ............................................................................................................................. 79 9 TEST SETUP PHOTOS ................................................................................................................................................. 81 10 TEST EQUIPMENT AND ANCILLARIES USED FOR TESTING ..........................…
Text truncated - open the document above for the full version.
SOBERLINK www.fieldtheoryinc.com 6-23-22 Purpose of Report: Passive RF Performance and FT matching of EM939000002 (dated 6/20/22) Click to edit Master title style Report Contents Page 2CONFIDENTIAL •FT Match Topology •Gain Performance •Return Loss •Passive Efficiency •1D Radiation Patterns •2D Heat Map & 3D Spherical Radiation Patterns www.fieldtheoryinc.com Soberlink Click to edit Master title style FT Match Topology Page 4CONFIDENTIAL From L44 PadTo Antenna 1pF Capacitor Part #: 04021J1R0PBS Click to edit Master title style Gain Performance Page 5CONFIDENTIAL Gain Performance Frequency2.42 GHz Maximum Gain3.63 dBi Unit Tested: EM939000002 Click to edit Master title style Return Loss Page 6CONFIDENTIAL Frequency 2.42 GHz Return Loss-10.3 dB Smith Chart Impedance (R + XjΩ)31.9 -18.4j Unit Tested: EM939000002 Click to edit Master title style Passive Efficiency Page 7CONFIDENTIAL 0 10 20 30 40 50 60 70 80 90 22.052.12.152.22.252.32.352.42.452.52.552.62.652.72.752.82.852.92.953 Efficiency (%) Frequency (GHz) Frequency2.42 GHz Efficiency68.70% Unit Tested: EM939000002 Click to edit Master title style 1D Radiation Patterns Page 8 CONFIDENTIAL RED BLUEGREEN -20 -15 -10 -5 0 5 0 15 30 45 60 75 90 105 120 135 150 165 180 195 210 225 240 255 270 285 300 315 330 345 -20 -15 -10 -5 0 5 0 15 30 45 60 75 90 105 120 135 150 165 180 195 210 225 240 255 270 285 300 315 330 345 -20 -15 -10 -5 0 5 0 15 30 45 60 75 90 105 120 135 150 165 180 195 210 225 240 255 270 285 300 315 330 345 Click to edit Master title style Heat Map & 3D Spherical Page 9CONFIDENTIAL 2.42 GHz2.42 GHz Unit Tested: EM939000002 Click to edit Master title style Example: How To Read & Compare the Heat Map & 3D Spherical Page 10CONFIDENTIAL This is the back side of the 3D Pattern This edge is the North Pole of the 3D This edge is the South Pole of the 3D This slit represents the elevation edges at Az 0 & 360 degrees. Notice that the Null on the edge is cut in half The advantage of a Heat Map is to show the entire 3D plot in one graph The slit opens 365 degrees clockwise starting on the left side. This causes the left and right side of the graphs to connect with the opposing figure's side Questions? www.fieldtheoryinc.com
CETECOM Inc. 411 Dixon Landing Road Milpitas, CA 95035 U.S.A. Phone: + 1 (408) 586 6200 Fax: + 1 (408) 586 6299 E-mail: [email protected] http://www.cetecom.com CETECOM Inc. is a Delaware Corporation with Corporation number: 2905571 FCC/IC Test Setup Photos For: Soberlink Healthcare Brand: Connect ST Marketing Name: Connect Model #: Connect ST Product Description: Breathalyzer FCC ID: 2A8ZY-3900 Addition to Report(s) #: EMC_SOBER-012-22001_FCC_15.247_BLE_DTS_Rev2 EMC_SOBER-012-22001_FCC_15B_Rev2 DATE: 12/6/2022 Page 2 of 13 1 EUT Photos Page 3 of 13 2 Measurements setup 2.1 Unintentional Radiated emissions 30 MHz to 1 GHz Page 4 of 13 1 GHz to 3 GHz Page 5 of 13 3 GHz to 18 GHz Page 6 of 13 2.2 Radiated spurious emissions 9 kHz – 30 MHz Page 7 of 13 30 MHz to 1 GHz Page 8 of 13 1 GHz to 3 GHz Page 9 of 13 3 GHz to 18 GHz Page 10 of 13 18 – 26 GHz Page 11 of 13 2.3 Conducted measurement RF Conducted emissions Page 12 of 13 AC Conducted emissions Page 13 of 13 3 Revision History Date Report Name Changes to report Report prepared by 10/26/2022 EMC_SOBER_012_22001_FCC_Setup_Photos Initial Version Issa Ghanma 11/4/2022 EMC_SOBER_012_22001_FCC_Setup_Photos_Rev1 Correction to the addition to reports # names on the cover page Issa Ghanma 12/6/2022 EMC_SOBER_012_22001_FCC_Setup_Photos_Rev2 Correction to the addition to reports # names on the cover page Issa Ghanma <<< The End >>>
| # | Rule Parts | Frequency Range | Power Output |
|---|---|---|---|
| 1 | 15C | 2.40 GHz - 2.48 GHz | 3.40 mW |