Text extracted from the exhibit documents filed with the FCC. Open a document above to read the original.
UltrasonicEndoActivationDevice UserManual ChangzhouBomedentMedicalTechnologyCo.,Ltd. ThankyouverymuchforchoosingtheUltrasonicEndoActivationDeviceof Bomedent Togivefullplaytothefunctionofthisdevice,andoperateandmaintainit correctlyandsafely,pleasereadthisUserManualcarefullybeforeusingit,and keepthisUserManualproperlyforreferenceatanytime Classificationofdevicesafetylevel: Classificationofelectricshockprotectiontype:ClassII Classificationofelectricshockprotectionlevel:ClassB Classificationofliquidentrypreventionlevel:IPX0 Sterilizationordisinfectionmethod:pleaserefertothecleaninganddisinfection section Classifiedbythesafetylevelinthecaseofusingtheproductwithflammable anestheticgasmixedwithairorflammableanestheticgasmixedwithoxygenor nitrousoxide:nottobeusedinthecaseofusingtheproductwithflammable anestheticgasmixedwithairorflammableanestheticgasmixedwithoxygenor nitrousoxide Operationmode:intermittentloadcontinuousoperation Contents 1.ProductIntroduction...............................................................................................................1 1.1BriefIntroduction.............................................................................................................1 1.2Model................................................................................................................................1 1.3ScopeofApplication........................................................................................................1 1.4Contraindications..............................................................................................................1 1.5PrecautionsforUse...........................................................................................................1 2.ProductConfiguration.............................................................................................................3 2.1ExternalStructureofMainFrameofProduct..................................................................3 2.2MainAccessoriesofProduct............................................................................................3 2.3AccessoriesList................................................................................................................4 3.UserInterface..........................................................................................................................4 4.ProductInstallation.................................................................................................................6 4.1InstallationofSiliconeCase.............................................................................................6 4.2InstallationofWorkingTip..............................................................................................6 4.3RemovalofWorkingTip..................................................................................................7 5.ProductUse.............................................................................................................................7 5.1Operation..........................................................................................................................7 5.2Charging............................................................................................................................9 5.3WirelessFootControl.....................................................................................................10 6.Troubleshooting....................................................................................................................12 7.Cleaning,DisinfectionandSterilization...........................................................................13 7.1CleaningandSterilizationofWorkingTip,WrenchandSiliconeCase........................14 8.Storage,MaintenanceandTransportation............................................................................15 9.TechnicalParameters............................................................................................................16 10.SymbolDescription............................................................................................................17 11.EnvironmentalProtection...................................................................................................17 12.After-salesService..............................................................................................................17 13.EMCStatement...................................................................................................................18 1 1.ProductIntroduction 1.1BriefIntroduction UltrasonicEndoActivationDevice(hereinafterreferredtoas“ActorIpro”or “device”)isanauxiliarydevicefordentiststoperformrootcanaltreatment.Itis mainlyusedtoeffectivelycleantherootcanalwiththehelpofultrasoniccavitation, soastoassistdentiststocompleterootcanaltreatment. Thefeaturesofthisdeviceareasfollows: a)Theworkingprocessofthisdeviceisunderfullautomaticcontrolof microchip,whichisconvenientandsimpletooperate; b)Thevibrationamplitudeoftheworkingtipcanbeadjustedsoasnottoharm therootcanal,anditsfrequencycanbedynamicallyadjustedtoensurecleaning efficiency; c)Theworkingtip,siliconecaseandwrenchcanbesterilizedbyhigh temperatureandhighpressure. d)Thisdevicecanbeilluminatedtomakeitclearer. 1.2Model ActorIpro 1.3ScopeofApplication Intendeduse:forcleaningoftherootcanalareaofteeth. User:dentist Useplace:hospitalordentalclinic 1.4Contraindications a)Itiscontraindicatedbyhemophiliacs,patientswithcardiacpacemakeror cochlearimplantanddoctors; b)Itshouldbeusedwithcautiononpatientswithheartdisease,pregnant womenandyoungchildren 1.5PrecautionsforUse PleasereadthisUserManualcarefullybeforeuse. 2 Thisdevicemaynotworkproperlyduetothefollowingenvironmental factors: 1)Portableormovableradiofrequencytransmittersarepresentnearby. 2)Electromagneticinterferencemaycauseabnormaloperationofthisdevice. Aswithallelectronicdevices,thisdevicehaselectromagneticinterference andshouldnotbeusedonpatientswithcardiacpacemaker. Duringoper…
Text truncated - open the document above for the full version.
郔 SGS North America Inc. 620 Old Peachtree Road SUITE 100 Applicant Declaration Suwanee, Georgia 30024 United States Applicant Legal Business Name Changzhou BoMedent Medical Technology Co.,Ltd Address N0.9 Changyang road, West Taihu Science Technology Industrial Park, Changzhou , China. Grantee Code 2AYV6 FCCID 2A YV6ACTORIP Authorized Contact Name Lili Zhang Contact Email [email protected] Contact Phone +86 051988991980 I, the undersigned, certify that I am an authorized signatory for the Applicant and therefore declare; a) in accordance with 47CFR2.911 (d), all of the statements herein and the exhibits attached hereto are true and correct to the best of my knowledge and belief. b ) in accepting a Grant of Equipment Authorization issued by a TCB, under the authority of the FCC, as a result of the representations made in this application, the Applicant is responsible for: (1) labeling the equipment with the exact FCC ID as specified in this application, (2) compt.ance statement labeling pursuant to the applicable rules , (3) compliance of the equipment with the applicable technical rules, c) if the Applicant is not the actual manufacturer of the equipment, appropriate arrangements have been made with the manufacturer to ensu「e that production units of this equipment will continue to comply with the FCC's technical requirements. d) in accordance with 47 CFR 2.909 and KDB394321, the Applicant has read, understood and agrees to ac岱pt that they are the responsible party and agree to abide by their responsibilities as specified under 47 CFR 2.909 and KDB394321. e) in accordance with 1S017065, FCC KDB641163, FCC KDB610077, KDB394321 and RSP-100, the Applicant has read, understood, accepts and agrees to abide by the post market surveillance requirements. (1) the Applicant understands, accepts and agrees that a sample may be requested for surveillance testing. (2) the Applicant shall make provisions to always have a production sample available upon request by SGS, FCC and/or ISED (3) the Applicant shall, upon request by SGS, at the Applicant's expense, provide a production sample of the requested product to SGS, FCC and/or ISED as instructed. The sample shall include all support devices, cables, software, accessories or other hardware or software required for evaluation, review, certification and audit surveillance of products certified by SGS. f) neithe『the的plicant nor any party to the application is subject to a denial of Federal benefits, that includes FCC benefits, pursuant to Section 5301 of the Anti-Drug Abuse Act of 1988, 21 U.S.C. §862 because of a conviction for possession or distribution of a controlled substance. See 47CFR 1.2002(b) fo「the definition of a "party" for these purposes. g) the Applicant has read, understood, accepts and agrees to abide by the SGS North America, lnc.(TCB) terms and conditions. Applicant Signature L也4哼 Print Name: Title: Date: 严/汁8 Applicant Declaration Rev:0.0
<Changzhou BoMedent Medical Technology Co.,Ltd> To: SGS North America Inc. 620 Old Peachtree Road SUITE 100 Suwanee, Georgia United States From: Changzhou BoMedent Medical Technology Co.,Ltd N0.9 Changyang road, West Taihu Science Technology Industrial Par k, Changzhou ,China Regarding: Confidentiality Request regarding application for FCC ID:2AYV6ACTORIP LONG TERM CONFIDENTIALITY Pursuant to 47 CFR Section 0.459 and 0.457 of the commission's rules, the applicant hereby request confidential treatment of the documents listed below, associated with the certification application referenced above. Schematic(s) Block Diagrams Operational Descriptions The documents above contain proprietary information not released to the public. Public disclosu「eof this information may prove harmful to the business of the applicant. : ChangZhou BoMedent Medical Technology Co., Ltd. : Quality Supervisor
< Changzhou BoMedent Medical Technolog y Co.,Ltd> TO: Federal Communication Commission Equipment Authorization Branch 7 435 Oakland Mills Road Columbia, MID 21046 Regarding: 2AYV6ACTORIP To whom it may concern: Date: 2021/218 We, the undersigned, hereby authorize Parlam Zhan to act on our behalf in all manners relating to application for equipment authorization with respect to the FCC ID above, including signing of all documents relating to these matters. Any and all acts carried out by the agent on our behalf shall have the same effect as acts of our own. We, the undersigned, hereby certify that we are not subject to a denial of federal benefits, that includes FCC benefits, pursuant to Section 5301 of the Anti-Drug Abuse Act of 1988, 21 U.S.C. 853(a). Where our agent signs the application for certification on our behalf, I acknowledge that all responsibility for complying with the terms and conditions for Certification, as specified by SGS North America, Inc., still resides with«Changzhou BoMedent Medical Technolog y Co.,Ltd». e from the applicant. Title:Quality Supervisor , Ltd +86 051988991980
SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Page: 1 of 5 Page 1 of 5 External Photos Applicant: ChangZhou BoMedent Medical Technology Co., Ltd. Address of Applicant: No. 9 Changyang Road, West Taihu Science Technology Industrial Park, Changzhou, China Equipment Under Test (EUT): Product Name: Ultrasonic Endo Activation Devices Model No.: Actor I pro SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Page: 2 of 5 Page 2 of 5 SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Page: 3 of 5 Page 3 of 5 SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Page: 4 of 5 Page 4 of 5 SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Page: 5 of 5 Page 5 of 5 End
FCC ID : 2AYV6ACTORIP R emark:It’s hard to put FCC statement on the product due to device’s irregular appearance. FCC statement will be placed in the user manual which the device is marketed.
SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Page: 1 of 8 Page 1 of 8 Internal Photos Applicant: ChangZhou BoMedent Medical Technology Co., Ltd. Address of Applicant: No. 9 Changyang Road, West Taihu Science Technology Industrial Park, Changzhou, China Equipment Under Test (EUT): Product Name: Ultrasonic Endo Activation Devices Model No.: Actor I pro SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Page: 2 of 8 Page 2 of 8 SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Page: 3 of 8 Page 3 of 8 SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Page: 4 of 8 Page 4 of 8 SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Page: 5 of 8 Page 5 of 8 SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Page: 6 of 8 Page 6 of 8 SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Page: 7 of 8 Page 7 of 8 SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Page: 8 of 8 Page 8 of 8 End
SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Report No.: SHEM210100054402 Page: 1 of 7 Parlam Zhan E&E Section Manager The manufacturer should ensure that all products in series production are in conformity with the product sample detailed in this report. If the product in this report is used in any configuration other than that detailed in the report, the manufacturer must ensure the new system complies with all relevant standards. Any mention of SGS International Electrical Approvals or testing done by SGS International Electrical Approvals in connection with, distribution or use of the product described in this report must be approved by SGS International Electrical Approvals in writing. 1 Cover Page RF Exposure REPORT Application No.: SHEM2101000544CR FCC ID: 2AYV6ACTORIP Applicant: ChangZhou BoMedent Medical Technology Co., Ltd. Address of Applicant: No. 9 Changyang Road, West Taihu Science Technology Industrial Park, Changzhou, China Manufacturer: ChangZhou BoMedent Medical Technology Co., Ltd. Address of Manufacturer: No. 9 Changyang Road, West Taihu Science Technology Industrial Park, Changzhou, China Factory: ChangZhou BoMedent Medical Technology Co., Ltd. Address of Factory: No. 9 Changyang Road, West Taihu Science Technology Industrial Park, Changzhou, China Equipment Under Test (EUT): EUT Name: Ultrasonic Endo Activation Devices Model No.: Actor I pro Trade mark: Standard(s) : FCC Rules 47 CFR §2.1093 KDB447498 D01 General RF Exposure Guidance v06 Date of Receipt: 2021-01-15 Date of Test: 2021-03-02 to 2021-04-06 Date of Issue: 2021-04-13 Test Result: Pass* * In the configuration tested, the EUT complied with the standards specified above. SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Report No.: SHEM210100054402 Page: 2 of 7 Revision Record Version Description Date Remark 00 Original 2021-04-13 / Authorized for issue by: Bill Wu / Project Engineer Parlam Zhan /Reviewer SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Report No.: SHEM210100054402 Page: 3 of 7 2 Contents Page 1 COVER PAGE ................................................................................................................................................. 1 2 CONTENTS ..................................................................................................................................................... 3 3 GENERAL INFORMATION........................................................................................................................... 4 3.1 GENERAL DESCRIPTION OF E.U.T. ......................................................................................................... 4 3.2 DETAILS OF E.U.T. .................................................................................................................................. 4 3.3 TEST LOCATION ....................................................................................................................................... 5 3.4 TEST FACILITY .......................................................................................................................................... 5 4 TEST STANDARDS AND LIMITS ............................................................................................................... 6 4.1 FCC RADIOFREQUENCY RADIATION EXPOSURE LIMITS .......................................................................... 6 5 MEASUREMENT AND CALCULATION .................................................................................................... 7 5.1 MAXIMUM TRANSMIT POWER ................................................................................................................... 7 5.2 RF EXPOSURE CALCULATION ................................................................................................................. 7 SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Report No.: SHEM210100054402 Page: 4 of 7 3 General Information 3.1 General Description of E.U.T. Power supply: DC 3.7V 1600mAh rechargeable battery 3.2 Details of E.U.T. Antenna Gain: 0.5dBi (Provided by manufacturer) Antenna Type: Chip Antenna Bluetooth Version: V5.0 LE Data Rate: 1 Mbps Channel Spacing: 2MHz Modulation Type: GFSK Number of Channels: 40 Operation Frequency: 2402MHz to 2480MHz SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Report No.: SHEM210100054402 Page: 5 of 7 3.3 Test Location All tests were performed at: SGS-CSTC Standards Technical Services(Shanghai) Co., Ltd. 588 West Jindu Road, Xinqiao, Songjiang, 201612 Shanghai, China. Tel: +86 21 6191 5666 Fax: +86 21 6191 5678 3.4 Test Facility The test facility is recognized, certified, or accredited by the following organizations: • CNAS (No. CNAS L0599) CNAS has accredited SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. to ISO/IEC 17025:2017 General Requirements for the Competence of Testing and Calibration Laboratories (CNAS-CL01 Accreditation Criteria for the Competence of Testing and Calibration Laborato ries) for the competence in the field of testing. • A2LA (Certificate No. 6332.01) SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. is accredited by the American Association for Laboratory Accreditation(A2LA). • FCC (Designation Number: CN1301) SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. has been recognized by Federal Communications Commission (FCC) as an accredited testing laboratory. • ISED (CAB Identifier: CN0020) SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. EMC Laboratory has been recognized by In novation, Science and Economic Development Canada (ISED) as an accredited testing laboratory. • VCCI (Member No.: 3061) The 3m Semi-anechoic chamber and Shielded Room of SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. has been registered in accordance with the Regulations for Voluntary Control Measures with Registration No.: R-13868, C-14336, T- 12221, G-10830 respectively. SGS-CSTC Standards Technical Services Co…
Text truncated - open the document above for the full version.
SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Appendix A for SHEM210100054401 1.6dB Bandwidth Test Mode Test Channel EBW[MHz] Limit Verdict BLE 2402 0.69 0.5 PASS BLE 2440 0.69 0.5 PASS BLE 2480 0.70 0.5 PASS SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch 6dB Bandwidth_BLE_2402 6dB Bandwidth_BLE_2440 6dB Bandwidth_BLE_2480 SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch 2.Occupied Bandwidth Test Mode Test Channel OBW[MHz] Limit[MHz] Verdict BLE 2402 1.06 --- PASS BLE 2440 1.08 --- PASS BLE 2480 1.09 --- PASS SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Occupied Bandwidth_BLE_2402 Occupied Bandwidth_BLE_2440 Occupied Bandwidth_BLE_2480 SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch 3.Maximum peak conducted output power Test Mode Test Channel Power[dBm] Limit[dBm] Verdict BLE 2402 -8.52 30 PASS BLE 2440 -9.03 30 PASS BLE 2480 -11.06 30 PASS SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Maximum peak conducted output power_BLE_2402 Maximum peak conducted output power_BLE_2440 Maximum peak conducted output power_BLE_2480 SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch 4.Maximum Peak power spectral density Test Mode Test Channel PSD[dBm/3kHz] Limit[dBm/3kHz] Verdict BLE 2402 -26.79 8.00 PASS BLE 2440 -27.07 8.00 PASS BLE 2480 -28.89 8.00 PASS SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Maximum Peak power spectral density_BLE_2402 Maximum Peak power spectral density_BLE_2440 Maximum Peak power spectral density_BLE_2480 SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch 5.Band-edge for RF Conducted Emissions Test Mode Test Channel Carrier Power[dBm] Max. Spurious Level [dBm] Limit [dBm] Verdict BLE 2402 -8.61 -48.11 -28.61 PASS BLE 2480 -11.26 -63.42 -31.26 PASS SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Band-edge for RF Conducted Emissions_BLE_2402_Hopping Off Band-edge for RF Conducted Emissions_BLE_2480_Hopping Off SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch 6.RF Conducted Spurious Emissions Test Mode Test Channel StartFre [MHz] StopFre [MHz] RBW [kHz] VBW [kHz] Pref[dBm] Max. Level [dBm] Limit [dBm] Verdict BLE 2402 30 10000 100 300 -8.76 -44.39 <-28.76 PASS BLE 2402 10000 26000 100 300 -8.757 -43.935 <- 28.757 PASS BLE 2440 30 10000 100 300 -9.20 -52.13 <-29.20 PASS BLE 2440 10000 26000 100 300 -9.204 -44.210 <- 29.204 PASS BLE 2480 30 10000 100 300 -11.28 -39.82 <-31.28 PASS BLE 2480 10000 26000 100 300 -11.282 -44.456 <- 31.282 PASS SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch RF Conducted Spurious Emissions_BLE_2402 SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch RF Conducted Spurious Emissions_BLE_2440 SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch RF Conducted Spurious Emissions_BLE_2480 SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch
SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Report No.: SHEM210100054401 Page: 1 of 35 Parlam Zhan E&E Section Manager The manufacturer should ensure that all products in series production are in conformity with the product sample detailed in this report. If the product in this report is used in any configuration other than that detailed in the report, the manufacturer must ensure the new system complies with all relevant standards. Any mention of SGS International Electrical Approvals or testing done by SGS International Electrical Approvals in connection with, distribution or use of the product described in this report must be approved by SGS International Electrical Approvals in writing. TEST REPORT Application No.: SHEM2101000544CR FCC ID: 2AYV6ACTORIP Applicant: ChangZhou BoMedent Medical Technology Co., Ltd. Address of Applicant: No. 9 Changyang Road, West Taihu Science Technology Industrial Park, Changzhou, China Manufacturer: ChangZhou BoMedent Medical Technology Co., Ltd. Address of Manufacturer: No. 9 Changyang Road, West Taihu Science Technology Industrial Park, Changzhou, China Factory: ChangZhou BoMedent Medical Technology Co., Ltd. Address of Factory: No. 9 Changyang Road, West Taihu Science Technology Industrial Park, Changzhou, China Equipment Under Test (EUT): EUT Name: Ultrasonic Endo Activation Devices Model No.: Actor I pro Trade mark: Standard(s) : 47 CFR Part 15, Subpart C 15.247 Date of Receipt: 2021-01-15 Date of Test: 2021-03-02 to 2021-04-06 Date of Issue: 2021-04-13 Test Result: Pass* * In the configuration tested, the EUT complied with the standards specified above. SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Report No.: SHEM210100054401 Page: 2 of 35 Revision Record Version Description Date Remark 00 Original 2021-04-13 / Authorized for issue by: Bill Wu / Project Engineer Parlam Zhan / Reviewer SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Report No.: SHEM210100054401 Page: 3 of 35 2 Test Summary Radio Spectrum Technical Requirement Item Standard Method Requirement Result Antenna Requirement 47 CFR Part 15, Subpart C 15.247 N/A 47 CFR Part 15, Subpart C 15.203 & 15.247(b)(4) Pass Radio Spectrum Matter Part Item Standard Method Requirement Result Minimum 6dB Bandwidth 47 CFR Part 15, Subpart C 15.247 ANSI C63.10 (2013) Section 11.8.1 47 CFR Part 15, Subpart C 15.247a(2) Pass Conducted Peak Output Power 47 CFR Part 15, Subpart C 15.247 ANSI C63.10 (2013) Section 11.9.1 47 CFR Part 15, Subpart C 15.247(b)(3) Pass Power Spectrum Density 47 CFR Part 15, Subpart C 15.247 ANSI C63.10 (2013) Section 11.10.2 47 CFR Part 15, Subpart C 15.247(e) Pass Conducted Band Edges Measurement 47 CFR Part 15, Subpart C 15.247 ANSI C63.10 (2013) Section 11.13.3.2 47 CFR Part 15, Subpart C 15.247(d) Pass Conducted Spurious Emissions 47 CFR Part 15, Subpart C 15.247 ANSI C63.10 (2013) Section 11.11 47 CFR Part 15, Subpart C 15.247(d) Pass Radiated Emissions which fall in the restricted bands 47 CFR Part 15, Subpart C 15.247 ANSI C63.10 (2013) Section 6.10.5 47 CFR Part 15, Subpart C 15.205 & 15.209 Pass Radiated Spurious Emissions 47 CFR Part 15, Subpart C 15.247 ANSI C63.10 (2013) Section 6.4,6.5,6.6 47 CFR Part 15, Subpart C 15.205 & 15.209 Pass SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Report No.: SHEM210100054401 Page: 4 of 35 3 Contents Page 1 COVER PAGE.............................................................................................................................................. 1 2 TEST SUMMARY ......................................................................................................................................... 3 3 CONTENTS .................................................................................................................................................. 4 4 GENERAL INFORMATION ......................................................................................................................... 5 4.1 DETAILS OF E.U.T.................................................................................................................................. 5 4.2 DESCRIPTION OF SUPPORT UNITS ........................................................................................................... 5 4.3 POWER LEVEL SETTING USING IN TEST .................................................................................................... 5 4.4 MEASUREMENT UNCERTAINTY ................................................................................................................ 6 4.5 TEST LOCATION ..................................................................................................................................... 7 4.6 TEST FACILITY ....................................................................................................................................... 7 4.7 DEVIATION FROM STANDARDS ................................................................................................................ 7 4.8 ABNORMALITIES FROM STANDARD CONDITIONS ....................................................................................... 7 5 EQUIPMENT LIST ....................................................................................................................................... 8 6 RADIO SPECTRUM TECHNICAL REQUIREMENT ................................................................................... 9 6.1 ANTENNA REQUIREMENT ........................................................................................................................ 9 7 RADIO SPECTRUM MATTER TEST RESULTS ...................................................................................... 10 7.1 MINIMUM 6DB BANDWIDTH ................................................................................................................... 10 7.2 CONDUCTED PEAK OUTPUT POWER ..................................................................................................... 11 7.3 POW…
Text truncated - open the document above for the full version.
SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Page: 1 of 4 Page 1 of 4 Test Setup Photos-FCC Applicant: ChangZhou BoMedent Medical Technology Co., Ltd. Address of Applicant: No. 9 Changyang Road, West Taihu Science Technology Industrial Park, Changzhou, China Equipment Under Test (EUT): Product Name: Ultrasonic Endo Activation Devices Model No.: Actor I pro SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Page: 2 of 4 Page 2 of 4 1 Contents Page 1 CONTENTS............................................................................................................... 2 2 EQUIPMENT UNDER TEST PICTURES ............................................................ 3 2.1 Radiated Emission Test Setup .......................................................................... 3 2.2 RF Conducted Test Setup ................................................................................. 4 SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Page: 3 of 4 Page 3 of 4 2 EQUIPMENT UNDER TEST PICTURES 2.1 Radiated Emission Test Setup 30MHz to 1GHz Above 1GHz SGS-CSTC Standards Technical Services (Shanghai) Co., Ltd. Page: 4 of 4 Page 4 of 4 Zoom: 2.2 RF Conducted Test Setup End
588 West Jindu Road, Xinqiao, Songjiang · ShangHai · China
| # | Rule Parts | Frequency Range | Power Output |
|---|---|---|---|
| 1 | 15C | 2.40 GHz - 2.48 GHz | 100.00 µW |