FCCID.CO- FCC ID Database
HomeCompaniesEquipment ClassesSearch

© 2026 FCC ID Database. All rights reserved.

AboutContactData sourced from FCC public records
  1. Home/
  2. Jiangxi Guokemeixin Healthcare Technologies Co., Ltd./
  3. 2AZGYIOT-THE001

2AZGYIOT-THE001Wireless electronic thermometer Card Reader

Jiangxi Guokemeixin Healthcare Technologies Co., Ltd.
Wireless electronic thermometer Card Reader - FCC ID 2AZGYIOT-THE001 - Jiangxi Guokemeixin Healthcare Technologies Co., Ltd.
Click to zoom

Application Details

Equipment Class
DXX - Part 15 Low Power Communication Device Transmitter
Date of Grant
Apr 24, 2021
Application Purpose
Original Equipment
Date of Application
Apr 24, 2021
Equipment Note
Wireless electronic thermometer Card Reader
Frequency Range
13.56000000 - 13.56000000
Company
Jiangxi Guokemeixin Healthcare Technologies Co., Ltd.
Country
China

Documents & Files

Select a file to view

Users Manual

↗

Cover Letter(s)

↗
↗
↗

External Photos

↗

ID Label/Location Info

↗

Internal Photos

↗

Test Report

↗

Test Setup Photos

↗

Document Text

Text extracted from the exhibit documents filed with the FCC. Open a document above to read the original.

Users Manual

Operation Instructions of Wireless Electronic Thermometer Catalogue: 1. Product Name........................................................................................1 2. Scope of Application...............................................................................1 3. Product Composition..............................................................................1 4. Product Model........................................................................................1 5. Product Specification..............................................................................1 6. Operation Instructions.............................................................................2 7. Main Performance..................................................................................2 8. Precautions.............................................................................................3 9. Classification...........................................................................................3 10. Marking Description...............................................................................3 11. Maintenance .........................................................................................4 12. Environmental Protection................................................................ ... ... 4 13. Electromagnetic Compatibility...............................................................4 14. Storage Conditions................................................................................4 15. Transport Conditions.............................................................................4 16. Manufacture Date....................................................................... ...........4 17. Service Life........................................................................................ .... 4 18. Production License Information, Product Registration Certificate Information ...4 19. Manufacturer Unit.................................................................................. 4 20. Preparation Date of Instructions......................................................... ...5 21. Attachment............................................................................................ 7 [Product Name]: Wireless electronic thermometer [Scope Of Application]: The wireless electronic thermometer is used to measure the body temperature in medical institutions, and different models are suitable for different body temperature measuring parts. [Product Composition]: It consists of a temperature label sensor probe electrode and a card reader. The temperature label sensor probe electrode is mainly composed of an electronic label integrated with a temperature sensor, a medical foam backing and medical adhesive tapes; the card reader can read multiple temperature label sensor probe electrodes, and temperature label sensor probe electrode is disposable medical consumables. [Product Model]: See the Wireless electronic thermometer model list / specification list for details. [Product Specification]: See the Wireless electronic thermometer model list / specification list for details. [Operation Instructions]: 1. Check whether the packaging of temperature label sensor probe electrode is damaged and the probe is damaged or not . Please contact medical staff and be replaced if damaged. 2. Placement of temperature label sensor probe electrode: A1、A2 Model: Tear the tape isolation belt of temperature label sensor probe electrode,place the side of temperature head under the armpit or at the groin and other advised temperature acquisition points, and press hard for five seconds; place antenna section (coil) on chest or other comfortable position. A5 Model: Place column temperature sensing head under armpits ,mouth ,anus and other advised temperature acquisition points and clamp. A6 Model: Tear the Non-woven tape isolation belt of temperature label sensor probe electrode, place black temperature head on the skin surface to be measured and secure with non-woven tape. 3. After a long press on power button of the card reader left side , the buzzer “beep” sounds; the screen of card reader lights up , it turns on successfully and enters the standby state after power-up. Such as low power indicator lights up, it is recommended to change first. Paste the Model The specification of temperature label sensor probe electrode The specification of car reader A1 A140W0 HF A2 A100W0 A5 A125C A6 A135S0 temperature measurement label sensor probe electrodes at least 5 minutes, read data while making the card reader to get close to the torus of temperature label sensor probe electrode. It indicates that reading data is completed successful and completes the measurement, while the card reader displays the temperature data and buzzer “beep” sounds. The card reader enters the standby state after displaying the temperature for 3 seconds . 4. The card reader will turn off after a long press on the power button in standby mode. If it is in standby mode and doesn’t do the measurement in 5 minutes, it will shut down. 5. Interface description of the Wireless electronic thermometer: [Main performance]: 1. Temperature display scope: 35.00°C - 42.00°C 2. Resolution: 0.01°C 3. Margin of Error: ±0.1°C (35.0 - 41.0°C), ±0.2°C (41.0 - 42.0°C), repeat-ability:S≤ 0.1°C; 4. Maximum standby time for measurement : less than 5 minutes; 5. The card reader has the function of storing 5 measurements. 6. Normal operating condition: ambient temperature 5°C - 40°C; relative humidity < 85%; atmospheric pressure 70 - 106 kpa. Storage and transport conditions.The products shall be transported in packing boxes with sufficient strength , to ensure that they do not move between each other , and dmoisture-proof measurement should be taken.Transportation and storage environment temperature should be in 5 ~40°C, relative humidity is not more than 80%, atmospheric pressure in 70~ 106kPa, no corrosive gas and well ventilated room. [Precautio…

Text truncated - open the document above for the full version.

External Photos

SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Page: 1 of 4 Page 1 of 4 External Photos Applicant: Jiangxi Guokemeixin Healthcare Technologies Co., Ltd. Address of Applicant: No#4 Zhongyangdexin Factory ,Fuzhou, China Equipment Under Test (EUT): Product Name: Wireless electronic thermometer Card Reader Model No.: HF SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Page: 2 of 4 Page 2 of 4 SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Page: 3 of 4 Page 3 of 4 SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Page: 4 of 4 Page 4 of 4 End

ID Label/Location Info

W i reless Electronic Thermometer Card Reader Model / Specrf r cat r on: HF Charging Voltage : DC5V = 1A FCC ID 2AZGYIOT-THE001 Manufacturer, Jrangxi Guokemerxrn Healthcare Technologies Co., Ltd 田 回 ((,.,)) Please refer to the instruct i ons for more informat i on Batch No Manufacture Date Expir y Date S/N � NFC � 无拽曦子体温ct ittll 壳.也红,,_,. 名产存,,_畸 ..朽匾幔生严·-如,. ,1U 贰麟掣."沁岱""' Remark: Statement because the product shell is not suitable for placing, The FCC statement on the product and it be placed in the user manual which the device is marketed

Internal Photos

SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Page: 1 of 6 Page 1 of 6 Internal Photos Applicant: Jiangxi Guokemeixin Healthcare Technologies Co., Ltd. Address of Applicant: No#4 Zhongyangdexin Factory ,Fuzhou, China Equipment Under Test (EUT): Product Name: Wireless electronic thermometer Card Reader Model No.: HF SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Page: 2 of 6 Page 2 of 6 SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Page: 3 of 6 Page 3 of 6 SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Page: 4 of 6 Page 4 of 6 SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Page: 5 of 6 Page 5 of 6 SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Page: 6 of 6 Page 6 of 6 End

Test Report

SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Report No.: SHEM210100054901 Page: 1 of 22 Parlam Zhan E&E Section Manager The manufacturer should ensure that all products in series production are in conformity with the product sample detailed in this report. If the product in this report is used in any configurati on other than that detailed in the report, the manufacturer must ensure the new system complies with all relevant standards. Any mention of SGS International Electrical Approvals or testing done by SGS International Electrical Approvals in connection with, distribution or use of the product described in this report must be approved by SGS International Electrical Approvals in writing. TEST REPORT Application No.: SHEM2101000549CR FCC ID: 2AZGYIOT-THE001 Applicant: Jiangxi Guokemeixin Healthcare Technologies Co., Ltd. Address of Applicant: No#4 Zhongyangdexin Factory ,Fuzhou, China Manufacturer: Jiangxi Guokemeixin Healthcare Technologies Co., Ltd. Address of Manufacturer: No#4 Zhongyangdexin Factory ,Fuzhou, China Factory: Jiangxi Guokemeixin Healthcare Technologies Co., Ltd. Address of Factory: No#4 Zhongyangdexin Factory ,Fuzhou, China Equipment Under Test (EUT): EUT Name: Wireless electronic thermometer Card Reader Model No.: HF ¤ Please refer to section 2 of this report which indicates which model was actually tested and which were electrically identical. Standard(s) : 47 CFR Part 15, Subpart C 15.225 Date of Receipt: 2021-01-18 Date of Test: 2021-01-25 to 2021-01-29 Date of Issue: 2021-02-03 Test Result: Pass* * In the configuration tested, the EUT complied with the standards specified above. SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Report No.: SHEM210100054901 Page: 2 of 22 Revision Record Version Description Date Remark 00 Original 2021-02-03 / Authorized for issue by: Bill Wu / Project Engineer Parlam Zhan / Reviewer SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Report No.: SHEM210100054901 Page: 3 of 22 2 Test Summary Radio Spectrum Technical Requirement Item Standard Method Requirement Result Antenna Requirement 47 CFR Part 15, Subpart C 15.225 N/A 47 CFR Part 15, Subpart C 15.203 Pass Radio Spectrum Matter Part 20dB Bandwidth 47 CFR Part 15, Subpart C 15.225 ANSI C63.10 (2013) Section 6.9 47 CFR Part 15, Subpart C 15.215 Pass Emission Mask 47 CFR Part 15, Subpart C 15.225 ANSI C63.10 (2013) Section 6.4 47 CFR Part 15, Subpart C 15.225(a)&(b)&(C ) Pass* (Note) Frequency tolerance 47 CFR Part 15, Subpart C 15.225 ANSI C63.10 (2013) Section 6.8 47 CFR Part 15, Subpart C 15.225(e) Pass Radiated Emissions(9kHz- 30MHz) 47 CFR Part 15, Subpart C 15.225 ANSI C63.10 (2013) Section 6.4 47 CFR Part 15, Subpart C 15.225(d) & 15.209 Pass Radiated Emissions(30MHz- 1GHz) 47 CFR Part 15, Subpart C 15.225 ANSI C63.10 (2013) Section 6.5 47 CFR Part 15, Subpart C 15.225(d) & 15.209 Pass Note: The test level of the fundamental signal is below the limit of general spurious emission, so the test item doesn’t be performed. SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Report No.: SHEM210100054901 Page: 4 of 22 3 Contents Page 1 COVER PAGE ............................................................................................................................................. 1 2 TEST SUMMARY ......................................................................................................................................... 3 3 CONTENTS .................................................................................................................................................. 4 4 GENERAL INFORMATION ......................................................................................................................... 5 4.1 DETAILS OF E.U.T. ................................................................................................................................ 5 4.2 DESCRIPTION OF SUPPORT UNITS........................................................................................................... 5 4.3 MEASUREMENT UNCERTAINTY ................................................................................................................ 5 4.4 TEST LOCATION ..................................................................................................................................... 6 4.5 TEST FACILITY ....................................................................................................................................... 6 4.6 DEVIATION FROM STANDARDS ................................................................................................................ 6 4.7 ABNORMALITIES FROM STANDARD CONDITIONS ....................................................................................... 6 5 EQUIPMENT LIST ....................................................................................................................................... 7 6 RADIO SPECTRUM TECHNICAL REQUIREMENT ................................................................................... 8 6.1 ANTENNA REQUIREMENT ........................................................................................................................ 8 7 RADIO SPECTRUM MATTER TEST RESULTS ........................................................................................ 9 7.1 20DB BANDWIDTH .................................................................................................................................. 9 7.2 EMISSION MASK ................................................................................................................................... 11 7.3 FREQUENCY TOLERANCE ...................................................................................................................... 13 7.4 RADIATED EMISSIONS(9KHZ-30MHZ) ................................................................................................... 15 7.5 RADIATED EMISSIONS(30MHZ-1GHZ) ......................…

Text truncated - open the document above for the full version.

Test Setup Photos

SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Page: 1 of 4 Page 1 of 4 Test Setup Photos-FCC Applicant: Jiangxi Guokemeixin Healthcare Technologies Co., Ltd. Address of Applicant: No#4 Zhongyangdexin Factory ,Fuzhou, China Equipment Under Test (EUT): Product Name: Wireless electronic thermometer Card Reader Model No.: HF SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Page: 2 of 4 Page 2 of 4 1 Contents Page 1 CONTENTS............................................................................................................... 2 2 EQUIPMENT UNDER TEST PICTURES ............................................................ 3 2.1 Radiated Emission Test Setup .......................................................................... 3 SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Page: 3 of 4 Page 3 of 4 2 EQUIPMENT UNDER TEST PICTURES 2.1 Radiated Emission Test Setup Below 30MHz SGS-CSTC Standards Technical Services Co., Ltd. Shanghai Branch Page: 4 of 4 Page 4 of 4 30MHz to 1GHz End

Contact Information

Applicant

Yuanfeng Yu
[email protected]86-18627792499Fax: 86-07948213099

Technical Contact

SGS-CSTC Standards Technical Services Co., Ltd.Parlam Zhan
[email protected]+862161915666

588 West Jindu Road, Xinqiao, Songjiang · ShangHai · China

Non-Technical Contact

SGS-CSTC Standards Technical Services Co., Ltd.Parlam Zhan
[email protected]+862161915666

Test Firm

Compliance Certification Services (Kunshan) Inc.Jessica Chen
[email protected]+ 86-0512-57355888

Technical Specifications

#Rule PartsFrequency RangePower Output
115C13.56 MHz - 13.56 MHz-
Confidentiality
Long Term