FCCID.CO- FCC ID Database
HomeCompaniesEquipment ClassesSearch

© 2026 FCC ID Database. All rights reserved.

AboutContactData sourced from FCC public records
  1. Home/
  2. SENS Innovation ApS/
  3. 2BFOF-A016

2BFOF-A016Discrete wearable activity sensor for healthcare and research

SENS Innovation ApS

Application Details

Equipment Class
DTS - Digital Transmission System
Date of Grant
Nov 24, 2025
Application Purpose
Original Equipment
Equipment Note
Discrete wearable activity sensor for healthcare and research
Frequency Range
2402.00000000 - 2480.00000000
Company
SENS Innovation ApS
Country
Denmark

Documents & Files

Select a file to view

Attestation Statements

↗
↗

Cover Letter(s)

↗
↗
↗

ID Label/Location Info

↗
↗

RF Exposure Info

↗

Test Report

↗
↗

Document Text

Text extracted from the exhibit documents filed with the FCC. Open a document above to read the original.

Cover Letter(s)

2025/06/10 Eurofins Electrical and Electronic Testing NA, Inc. 914 West Patapsco Avenue Baltimore, MD 21230 RE: LETTER OF AGENT AUTHORIZATION To Whom It May Concern: We, the undersigned, hereby authorize SIQ Ljubljana, Mašera-Spasićeva ulica 10, 1000 Ljubljana, Slovenia, to act on our behalf in all matters relating to application for equipment authorization, including the signing of all documents relating to these matters. We also hereby certify that no party to the application authorized hereunder is subject to the denial of benefits, including FCC benefits, pursuant to Section 5301 of the Anti-Drug Abuse Act of 1988, 21 U.S.C.853(a). This agreement expires one year from the current date. Sincerely, Kasper Lundberg Lykkegaard, CEO SENS Innovation ApS

RF Exposure Info

The report shall not be reproduced except in full. SIQ Ljubljana, Mašera-Spasićeva ulica 10, SI-1000 Ljubljana, Slovenia T +386 1 4778 100, F +386 1 4778 444, [email protected], www.siq.si TN405E-04 Test report Number: T251-0303/25 Project file: C20240241 Date: 2025-11-24 Pages: 5 Product: Activity sensor Type reference: SENS motion PLUS, model A016 Ratings: 3 V d.c. Class ||| Trademark: Applicant: SENS Innovation ApS Nannasgade 28, Copenhagen, Denmark Manufacturer: SENS Innovation ApS Nannasgade 28, Copenhagen, Denmark Place of manufacture: SENS Innovation ApS Nannasgade 28, Copenhagen, Denmark Summary of testing Testing method: KDB 447498 D01 General RF Exposure Guidance v06 Testing location: SIQ Ljubljana Mašera-Spasićeva ulica 10, SI-1000 Ljubljana, Slovenia Remarks: Date of receipt of test items: 2025-03-13 Number of items tested: 1 Date of performance of tests: 2025-04-09 The test results presented in this report relate only to the items tested. The test items were tested in the condition as received. The product complies with the requirements of the testing methods. Tested by: Nik Vončina Approved by: Marjan Mak T251-0303/25 Page: 2 (5) CONTENTS page 1 GENERAL 3 1.1 EQUIPMENT UNDER TEST 3 2 ASSESSMENT PROCEDURE 4 3 MEASUREMENTS / CALCULATIONS 5 T251-0303/25 Page: 3 (5) 1 GENERAL History sheet Date Report No. Change Revision 2025-11-24 T251-0303/25 Initial Test Report issued. -- 1.1 Equipment under test Activity sensor Type: SENS motion PLUS, model A016 FCC ID: 2BFOF-A016 Environment: Uncontrolled / General Public Assessment distance: 5 mm Reviewed test report T251-0273/25 from SIQ Ljubljana. T251-0303/25 Page: 4 (5) 2 ASSESSMENT PROCEDURE KDB 447498 D01 General RF Exposure Guidance v06 Clause 4.3.1. Standalone SAR test exclusion considerations SAR evaluation for general population exposure conditions, by measurement or numerical simulation, is not required when the corresponding SAR Test Exclusion Threshold condition(s), listed below, is (are) satisfied. These test exclusion conditions are based on source-based time-averaged maximum conducted output power of the RF channel requiring evaluation, adjusted for tune-up tolerance, and the minimum test separation distance required for the exposure conditions. For frequencies between 100 MHz and 6 GHz, the following may be considered for SAR test exclusion: a) For 100 MHz to 6 GHz and test separation distances ≤ 50 mm, the 1-g and 10-g SAR test exclusion thresholds are determined by the following: [(max. power of channel, including tune-up tolerance, mW) / (min. test separation distance, mm)] · [√f(GHz)] ≤ 3.0 for 1-g SAR, and ≤ 7.5 for 10-g extremity SAR, 30 where • f(GHz) is the RF channel transmit frequency in GHz • Power and distance are rounded to the nearest mW and mm before calculation31 • The result is rounded to one decimal place for comparison • The values 3.0 and 7.5 are referred to as numeric thresholds in step b) below The test exclusions are applicable only when the minimum test separation distance is ≤ 50 mm, and for transmission frequencies between 100 MHz and 6 GHz. When the minimum test separation distance is < 5 mm, a distance of 5 mm according to 4.1 f) is applied to determine SAR test exclusion. b) For 100 MHz to 6 GHz and test separation distances > 50 mm, the 1-g and 10-g SAR test exclusion thresholds are determined by the following (also illustrated in Appendix B): 1) {[Power allowed at numeric threshold for 50 mm in step a)] + [(test separation distance – 50 mm)·(f(MHz)/150)]} mW, for 100 MHz to 1500 MHz 2) {[Power allowed at numeric threshold for 50 mm in step a)] + [(test separation distance – 50 mm)·10]} mW, for > 1500 MHz and ≤ 6 GHz c) For frequencies below 100 MHz, the following may be considered for SAR test exclusion (also illustrated in Appendix C): 1) For test separation distances > 50 mm and < 200 mm, the power threshold at the corresponding test separation distance at 100 MHz in step b) is multiplied by [1 + log(100/f(MHz))] 2) For test separation distances ≤ 50 mm, the power threshold determined by the equation in c) 1) for 50 mm and 100 MHz is multiplied by ½ 3) SAR measurement procedures are not established below 100 MHz. T251-0303/25 Page: 5 (5) 3 MEASUREMENTS / CALCULATIONS Antenna type and gain: Integral antenna, 3.68 dBi KDB 447498 D01 General RF Exposure Guidance v06 Clause 4.3.1: Frequency (MHz) Maximum* power with tune-up (dBm) Maximum* power with tune-up (mW) SAR Test Exclusion Threshold (mW) 2402-2480 0.48 1.12 10 * Gated power with Duty Cycle calculated in ** maximum tolerance provided from manufacturer is ±2dB. Conclusion: PASS; SAR Evaluation is not required due to SAR Test Exclusion Thresholds are met. There is no simultaneous transmission between any other transmitter.

Test Report

The report shall not be reproduced except in full. SIQ Ljubljana, Mašera-Spasićeva ulica 10, SI-1000 Ljubljana, Slovenia T +386 1 4778 100, F +386 1 4778 444, [email protected], www.siq.si TN405E-04 Test report Number T251-0273/25 Project file: C20240241 Date: 2025-11-24 Pages: 60 Product: Activity sensor Type reference: SENS motion PLUS, model A016 Ratings: 3 V d.c. Class ||| Trademark: Applicant: SENS Innovation ApS Nannasgade 28, Copenhagen, Denmark Manufacturer: SENS Innovation ApS Nannasgade 28, Copenhagen, Denmark Place of manufacture: SENS Innovation ApS Nannasgade 28, Copenhagen, Denmark Summary of testing Testing method: 47 CFR Part 15, Subpart C (Clause 15.247) last amended 2024-08-08 in conjunction with ANSI C63.10-2020+Cor. 1- 2023+C63.10a-2024 RSS-Gen (Issue 5), RSS-247 (Issue 3) in conjunction with ANSI C63.10-2020+Cor. 1-2023+C63.10a-2024 Testing location: SIQ Ljubljana Mašera-Spasićeva ulica 10, SI-1000 Ljubljana, Slovenia Remarks: Date of receipt of test items: 2025-03-13 Number of items tested: 6 Date of performance of tests: 2025-03-19 – 2025-04-02 The test results presented in this report relate only to the items tested. The test items were tested in the condition as received. The product complies with the requirements of the testing methods. Tested by: Luka Cvajnar Approved by: Marjan Mak T251-0273/25 Page: 2 (60) CONTENTS Page 1 GENERAL 3 1.1 DESCRIPTION OF EQUIPMENT UNDER TEST 3 1.2 DESCRIPTION OF THE TEST MODES 5 1.3 TEST EQUIPMENT USED FOR TESTING 6 1.4 TEST SETUP CONFIGURATIONS 7 2 TEST SUMMARY 9 2.1 APPLICATION OF DECISION RULE 9 3 TESTS RESULTS 10 3.1 47 CFR § 15.203 – ANTENNA REQUIREMENTS 10 3.2 47 CFR § 15.207 – AC LINE CONDUCTED EMISSION 11 3.3 47 CFR § 15.247(D) - SPURIOUS EMISSION - CONDUCTED 12 3.4 47 CFR § 15.205, § 15.209, § 15.247 (D) – SPURIOUS EMISSION - RADIATED 21 3.5 47 CFR § 15.247 (A) (2) – 6 DB EMISSION BANDWIDTH 38 3.6 47 CFR § 15.247 (B) (3) – MAXIMUM PEAK OUTPUT POWER 43 3.7 47 CFR § 15.247 (D) –100 KHZ BANDWIDTH OF FREQUENCY BAND EDGE 45 3.8 47 CFR § 15.247 (E) – POWER SPECTRAL DENSITY 52 4 ANNEX I – ADDITIONAL DATA 56 4.1 OCCUPIED BANDWIDTH (99% EMISSION BANDWIDTH) 57 T251-0273/25 Page: 3 (60) 1 GENERAL EUT passed the performed tests. History sheet Date Report No. Change Revision 2025-11-24 T251-0273/25 Initial Test Report issued. -- 1.1 Description of equipment under test Activity sensor Type: SENS motion PLUS, model A016 FCC ID: 2BFOF-A016 IC: 32412-A016 Adaptive / non-adaptive equipment non-adaptive equipment Modulation type Other than FHSS Operating mode Single antenna SIQ sample number: S202501526, S202501527, S202501528, S202501530, S202501531, S202501532 Maximum RF Output power 0 dBm Operating frequency 2402 MHz – 2480 MHz Number of channels 40 Antenna type and gain Integral antenna, 3.68 dBi Antenna Beamforming / Nominal channel bandwidth 1 MHz Hardware version (HVIN): A016 Firmware version (FVIN) 2.0.19-L Radio firmware version: 2.2.1-radio T251-0273/25 Page: 4 (60) Copy of marking plate The artwork below may be only a draft. The use of certification marks on a product must be authorized by the respective NCBs that own these marks. T251-0273/25 Page: 5 (60) 1.2 Description of the test modes The equipment uses only one antenna at any moment. On all devices run test firmware Bluetooth Direct Test Mode. For testing purposes sample with integral antenna and modified sample with added SMA connector were provided. For each test channel, radiated and conducted samples were provided by the manufacturer. 1.2.1 Tested Channels Channels Data rate Frequency [MHz] TX power Settings [dBm] Packet type Packet length (bytes) 0 (Lowest) 1 Mbps 2402 0 PRBS9 37 19 (Middle) 1 Mbps 2440 0 PRBS9 37 39 (Highest) 1 Mbps 2480 0 PRBS9 37 Normal test condition: Ambient temperature: 15 C to 35 C Relative humidity: 30 % to 60 % Atmospheric pressure: 860 mbar to 1060 mbar T251-0273/25 Page: 6 (60) 1.3 Test Equipment used for testing Manufacturer Description Model No. Used Calibrated Calibrated until Comtest engineering, SAC2 (together with controlling equipment) SAC 3m X 2022-04-14 2025-04-14 Maturo, Turn table (2 m diameter) TT 2.0 SI X / / Maturo, Bore-sight antenna mast BAM-4.0-P X / / Maturo, positioning equipment NCD X / / Rohde & Schwarz, RFI receiver ESW 44 X 2024-09-26 2026-03-26 R&S, Ultra Broadband Antenna HL562E X 2023-09-26 2026-09-26 R&S, Horn Antenna HF907 X 2023-08-22 2026-08-22 R&S, Spectrum Analyzer FSV 40 X 2024-11-08 2026-05-08 R&S, Vector signal generator SMBV100B / / / R&S, Signal generator SMB100A X 2024-11-11 2026-11-11 R&S High resolution power meter OSP-B157W8 X 2024-11-12 2026-11-12 R&S Switch unit OSP-B157WX X 2024-11-13 2026-11-13 Wainwright Instruments, High pass filter High pass filter WHNX6-2555- 3500-26500 X 2024-05-30 2025-05-30 Wainwright Instruments, High pass filter WHNX6-5925- 7500-26500 / / / Wainwright Instruments, High pass filter WHW2-16340- 21500-40000 / / / Hp, Manual step attenuator 8494B 11 DB X / / Hp, Manual step attenuator 8496B 110 DB X / / PMI Low noise amplifier PEC-42-1G40G X 2025-03-12 2026-09-12 KEYSIGHT, attenuator 8491B 10 DB / / / Kambič, Temperature chamber I-190 CK / / / 1.3.1 Measurement uncertainty The following measurement uncertainty levels have been estimated for tests performed on the product, as specified in ETSI TR 100 028-2 and C63.23. This represents an expanded uncertainty expressed at 95% confidence level using a coverage factor k=2. Measurements ULAB U ETSI TR 100 028-2 U C63.23 AC Line Conducted Emission 3.2 dB / ±4,13 Spurious emission 30 – 300 MHz 4.2 dB ±6 / Spurious emission 300 – 1000 MHz 4.4 dB ±6 / Spurious emission 1 GHz – 18 GHz 5.1 dB ±6 / Spurious emission 18 GHz – 26GHz 5.6 dB ±6 / Tx spurious emission - conducted < 1.8 dB ±4 / 6 dB Emission Bandwidth < 2% ±5% / Maximum peak output power < 1 dB ±0,75 dB / 100 kHz Bandwidth of Frequency Band Edge < 0.8 dB / / Power Spectral Density < 1.3 dB ±3 dB / Occupied bandwidth (99% emission bandwidth) < 2% ±5% / Note: Measurement uncertainty…

Text truncated - open the document above for the full version.

Test Report

The report shall not be reproduced except in full. SIQ Ljubljana, Mašera-Spasićeva ulica 10, SI-1000 Ljubljana, Slovenia T +386 1 4778 100, F +386 1 4778 444, [email protected], www.siq.si TN405E-04 Test report Number T251-0301/25 Project file: C20240241 Date: 2025-11-24 Pages: 9 Product: Activity sensor Type reference: SENS motion PLUS, model A016 Ratings: 3 V d.c. Class: ||| Trademark: Applicant: SENS Innovation ApS Nannasgade 28, Copenhagen, Denmark Manufacturer: SENS Innovation ApS Nannasgade 28, Copenhagen, Denmark Place of manufacture: SENS Innovation ApS Nannasgade 28, Copenhagen, Denmark Summary of testing Testing method: Antenna pattern measurements Testing location: SIQ Ljubljana Mašera-Spasićeva ulica 10, SI-1000 Ljubljana, Slovenia Remarks: Date of receipt of test items: 2025-03-13 Number of items tested: 3 Date of performance of tests: 2025-03-19 The test results presented in this report relate only to the items tested. The test items were tested in the condition as received. The product complies with the requirements of the testing methods. Tested by: Luka Cvajnar Approved by: Marjan Mak T251-0301/25 Page: 2 (9) CONTENTS page 1 GENERAL 3 1.1 EQUIPMENT UNDER TEST 3 1.2 ANTENNA PATTERN MEASUREMENT 4 1.3 TEST RESULTS BLE ANTENNA 5 1.4 MAXIMUM BLE ANTENNA GAIN 8 2 USED TEST EQUIPMENT 9 T251-0301/25 Page: 3 (9) 1 GENERAL History sheet Date Report No. Change Revision 2025-11-24 T251-0301/25 Initial Test Report issued. -- Environmental conditions: Ambient temperature: 15 C to 35 C Relative humidity: 30 % to 60 % Atmospheric pressure: 860 mbar to 1060 mbar 1.1 Equipment under test Activity sensor Type: SENS motion PLUS, model A016 Tested was antenna pattern of the sample below. Picture of test sample T251-0301/25 Page: 4 (9) Copy of marking plate The artwork below may be only a draft. The use of certification marks on a product must be authorized by the respective NCBs that own these marks. 1.2 Antenna pattern measurement 1.2.1 Test procedure The radiation pattern for BLE antenna implemented to PCB reference design has been measured in an anechoic chamber with 3 meters test distance. Test results show radiation patterns for two planes, measured with vertical and horizontal polarization of measuring antenna. All measurements were performed at 2402, 2440 and 2480 MHz frequency. T251-0301/25 Page: 5 (9) 1.3 Test results BLE Antenna T251-0301/25 Page: 6 (9) T251-0301/25 Page: 7 (9) T251-0301/25 Page: 8 (9) 1.4 Maximum BLE antenna gain DUT Frequency (MHz) Maximum antenna gain (dBi) 2402.000000 3.68 2440.000000 -6.85 2480.000000 -11.01 T251-0301/25 Page: 9 (9) 2 USED TEST EQUIPMENT Antenna pattern measurement Manufacturer Description Model No. Used Calibrated Calibrated until Comtest engineering, SAC2 (together with controlling equipment) SAC 3m X 2022-04-14 2025-04-14 Maturo, Turn table (2 m diameter) TT 2.0 SI X / / Maturo, Bore-sight antenna mast BAM-4.0-P X / / Maturo, positioning equipment NCD X / / Rohde & Schwarz, RFI receiver ESW 44 X 2024-09-26 2026-03-26 R&S, Ultra Broadband Antenna HL562E 2023-09-26 2026-09-26 R&S, Horn Antenna HF907 X 2023-08-22 2026-08-22 ----------------------------------------------------------END OF TEST REPORT------------------------------------------------------------

Contact Information

Applicant

Kasper L Lykkegaard(CEO)
[email protected]Fax: +4526238234

Test Firm

SIQ LjubljanaMarjan Mak
[email protected]Fax: 0038614778100

Technical Specifications

#Rule PartsFrequency RangePower Output
115C2.40 GHz - 2.48 GHz300.00 µW
Confidentiality
Long Term
Grant Notes
Output power listed is conducted.