Text extracted from the exhibit documents filed with the FCC. Open a document above to read the original.
+45 70 20 31 24 www.auditdata.com [email protected] Classification: Commercial Confidential Date: Oct 24, 2025 Auditdata A/S Wildersgade 10B DK-1408 Copenhagen Denmark Company ID: DK-29526532 Referred in this document as: AUDITDATA QAI Laboratories Ltd. 3980 North Fraser Way, Burnaby B.C. V5J 5K5 Canada Referred in this document as: QAI Product Name: “2000-1 Fitting Unit” Product Reference: PN00887 Product FCC ID: 2BK6Q-PN00887 To whom it may concern, Re: Product Certification Representative Authorization Letter We authorize QAI to act on our behalf on all matters concerning the above named equipment. This shall include the following: • Testing • Approval applications and procedures We declare that QAI is allowed to forward all information related to the approval project to the applied certification body and to the notification entities and to discuss any issues concerning the approval application. Any and all acts carried out by QAI on our behalf shall have the same effect as acts of our own. https://www.fcc.gov/supplychain/coveredlist By: Jonas Svendsen Sign and date on behalf of AUDITDATA: Title: Director of Hardware Engineering _______________________ Contact: Phone +45 20461214 - E-mail: [email protected] Jonas Svendsen Jonas Svendsen Copenhagen, 24 Oct 2025
Classification: Sensitive Auditdata A/S Wildersgade 10B DK-1408 Copenhagen Denmark Company ID: DK-29526532 Referred in this document as: AUDITDATA Equipment Authorization Division Federal Communications Commission 7435 Oakland Mills Road Columbia, MD 21046 USA Referred in this document as: FCC Product Name: “2000-1 Fitting Unit” Product Reference: PN00887 Product FCC ID: 2BK6Q-PN00887 AUDITDATA - Request for Confidentiality Pursuant to sections 0.457 and 0.459 of CFR 47, the applicant requests permanent confidential treatment of the following information accompanying this application: Exhibits Long-term Short-term ID Label/Location No No Attestation Statements No No External Photos No Yes Block Diagram Yes Yes Schematics Yes Yes Test Report No No Test Setup Photos No Yes User Manual (IFU) No No Internal Photos No Yes Parts List (BOM) Yes Yes Tune Up Yes Yes RF Exposure Info No Yes Operational Description Yes Yes Cover Letter No No SDR Software Info No No These materials contain trade secrets and proprietary information and are not customarily released to the public. The public disclosure of this information might be harmful to the company and provide unjustified benefits to our competitors. We further certify that the applicant nor any party to the application is subject to a denial of Federal benefits, that includes FCC benefits, pursuant to section 5301 of the Anti-Drug abuse Act of 1988, 21 U.S.C. Section 862. Sign and date on behalf of AUDITDATA: Date: 6 Jan 2026 Title: Director of Hardware Engineering _______________________ Contact: Phone +45 20461214 - E-mail: [email protected] Jonas Svendsen Jonas Svendsen
Classification: Sensitive Auditdata A/S Wildersgade 10B DK-1408 Copenhagen Denmark Company ID: DK-29526532 Referred in this document as Applicant: AUDITDATA QAI Laboratories Ltd. 3980 North Fraser Way Burnaby, BC V5J 5K5 Canada Referred in this document as: QAI Product Name: “2000-1 Fitting Unit” Product Reference: PN00887 Product FCC ID: 2BK6Q-PN00887 Federal Communications Commission Authorization and Evaluation Division 7435 Oakland Mills Road Columbia, MD 21046 USA Ref. #1: Attestation Statements Part 2.911(d)(5)(i) Filing AUDITDATA certifies that the equipment for which authorization is sought is not “covered” equipment prohibited from receiving an equipment authorization pursuant to section 2.903 of the FCC rules. Ref. #2: Attestation Statements Part 2.911(d)(5)(ii) Filing AUDITDATA certifies that, as of the date of the filing of the application, the applicant is not identified on the Covered List (as a specifically named entity or any of its subsidiaries or affiliates)as an entity producing “covered” equipment. Sign and date on behalf of AUDITDATA: Date: 5 Nov 2025 Title: Director of Hardware Engineering _______________________ Contact: Phone +45 20461214 - E-mail: [email protected] Jonas Svendsen Jonas Svendsen
Federal Communica ons Commission Authoriza on and Evalua on Division 7435 Oakland Mills Road Columbia, MD 21046 USA Date: 16/01/2026 Ref: A esta on Statement P art 2.911(d)(7) Filing FCC ID: 2BK6Q-PN00887 Jonas Svendsen cer fies that, as of the dates of this applica on, Americas Compliance Consul ng LLC dba iCer fi is our designated U.S. agent for service of process for the above referenced FCC ID. Manufacturer accepts to maintain an agent for no less than two year a er the grantee has terminated all marke ng and importa on or the conclusion of any commission-related proceeding involving the equipment. iCer fi accepts, as of the date of the filing of the applica on, the obliga on of the designated U .S. ag ent f or service process for the above r eferenced FCC ID . Designated U.S. Agent Informa on: Name: Americas Compliance Consul ng LLC dba iCer fi FCC FRN: 0033399411 Address: 5809 US-2, Suite B , Albeni Plaza Priest Riv er, Idaho 83856, United St ates Contact P erson: Paul Pr eston Tel: +1 (866) 885-4575 Email: fccagent@icer fi.com Sincerely, Audit DataAmericas Compliance Consul ng LLC dba iCer fi FRN: FRN: 0033399411 Jonas SvendsenPaul Pr eston [email protected]@icer fi.com +4520461214+1 (866) 885-4575 Signature: Signature: Jonas Svendsen
Classification: Sensitive Auditdata A/S Wildersgade 10B DK-1408 Copenhagen Denmark Company ID: DK-29526532 Referred in this document as Applicant: AUDITDATA QAI Laboratories Ltd. 3980 North Fraser Way Burnaby, BC V5J 5K5 Canada Referred in this document as: QAI Product Name: “2000-1 Fitting Unit” Product Reference: PN00887 Product FCC ID: 2BK6Q-PN00887 TO: Federal Communications Commission Authorization and Evaluation Division 7435 Oakland Mills Road Columbia, MD 21046 USA Ref.: Kaspersky Attestation Statement AUDITDATA confirms: The equipment identified above does not contain cybersecurity or antivirus software produced or provided by Kaspersky Lab, Inc., its successors, assignees, or affiliates. The equipment cannot store, run, or deliver any software identified on the FCC Covered List, including Kaspersky software. The device’s software, hardware and firmware architecture have been reviewed, and no components from Kaspersky or other prohibited entities are present. Sign and date on behalf of AUDITDATA: Date: 06 Jan 2026 Title: Hardware Engineering Manager _______________________ Contact: Phone +45 20461214 - E-mail: [email protected] Jonas Svendsen Jonas Svendsen
Classification: Sensitive Auditdata A/S Wildersgade 10B DK-1408 Copenhagen Denmark Company ID: DK-29526532 Referred in this document as: AUDITDATA QAI Laboratories Ltd. 3980 North Fraser Way Burnaby, BC V5J 5K5 Canada Referred in this document as: QAI Cover Letter To Whom It May Concern, We hereby submit this cover letter in support of our application for FCC equipment authorization for the following product: Product Name: Product Reference: Product FCC ID: “2000-1 Fitting Unit” PN00887 2BK6Q-PN00887 Grantee Code: S52-7-12-12-00-1 (QAI) Nature of Filing Request: New Certification This filing includes all required documentation as per QAI’s checklist, including: User Manual with FCC statements System Block Diagram Schematics Parts List Operational Description Internal and External Photos Label Sample and Location Authorization Letter Confidentiality Request Letter We confirm that the information provided is accurate and complete to the best of our knowledge. We understand that any missing or incomplete documentation may delay the certification process. Should you require any further information, please do not hesitate to contact us. Sign and date on behalf of AUDITDATA: Date: 24 Oct 2025 Title: Director of Hardware Engineering _______________________ Contact: Phone +45 20461214 - E-mail: [email protected] Jonas Svendsen Jonas Svendsen Copemhagen 24 Oct 2025
+45 70 20 31 24 www.auditdata.com [email protected] Classification: Commercial Confidential PN00887 Type Labels
DOC ID: DN03042/01 RF Exposure Compliance Issued under ISED Canada RSS-102 Issue 6 SAR Exemption Declaration 1. Manufacturer Information Manufacturer: Auditdata A/S Address: Wildersgade 10B, 1408 Copenhagen , Denmark Contact person: Jonas Svendsen, Hardware Engineering Manager 2. Product Identification Product name: Audiometer with wireless interface for REM measurements (Bluetooth based) Model(s): “2000-1 W-REM Probe” Ref.: PN00988 Description: The 2000-1 W-REM Probe is a wireless Probe used for measuring hearing loss. The Probe has 4 microphones for Real Ear Measurement. The sound captured from the 4 microphones are transmitted wirelessly to the “Main Unit”. The Probe is designed to work/connect with the “Main Unit” only and no other devices. The Probe has build-in battery and are charged by USB-C power supply. The Probe is not active while charging and can not be connected to other devices. 3. Applicable Directive and Standards This declaration is issued to demonstrate compliance with the RF exposure requirements of RSS-102 Issue 6, published by Innovation, Science and Economic Development (ISED) Canada. The device qualifies for SAR exemption under Section 6.3 of RSS-102 Issue 6, based on its low output power. The following standards have been applied: Standard Title Remark RSS-102 Issue 6 Radio Frequency (RF) Exposure Compliance of Radiocommunication Apparatus (All Frequency Bands) Used for RF exposure assessment RSS-GEN General Requirements for Compliance of Radio Apparatus Used for electrical and mechanical safety DOC ID: DN03042/01 4. Assessment Summary • The Bluetooth radio operates in the 2.4 GHz ISM band with a maximum EIRP of approximately +2.5 dBm (≈ 1.8 mW). • According to RSS-102 Issue 6 Section 6.3, devices operating at such low power levels are exempt from SAR testing. • The device meets the exemption limits defined for uncontrolled environments. Therefore, the equipment is exempt from SAR testing in accordance with RSS-102 Issue 6. 5. Conclusion Based on the evaluation above, the equipment complies with the RF exposure requirements of RSS- 102 Issue 6. No Specific Absorption Rate (SAR) testing is required for this product. 6. Signatures Place and Date of Issue: Copenhagen, Denmark – 3-Feb 2026 Name and Title: Jonas Svendsen, Hardware Engineering Manager Signature: ______________________________ Jonas Svendsen
DOC ID: DN03049/01 RF Exposure Compliance Issued under ISED Canada RSS-102 Issue 6 SAR Exemption Declaration 1. Manufacturer Information Manufacturer: Auditdata A/S Address: Wildersgade 10B, 1408 Copenhagen , Denmark Contact person: Jonas Svendsen, Hardware Engineering Manager 2. Product Identification Product name: Audiometer with wireless interface for REM measurements (Bluetooth based) Model(s): “2000-1 Fitting Unit” Ref.: PN00887 Description: The 2000-1 Fitting Unit is an Audiometer / Fitting System for measuring hearing loss. The system supports Real Ear Measurement – a way to test hearing loss on a client with hearing aids. The REM uses a Probe to measure the sound, for left and right ear, both outside the ear (reference) and on the inside of the hearing aid using a tube. The system can use Probe with wired and wireless connection. For wireless connection, the system can use “Wireless Probe” The system has two active wireless devices for the wireless REM connection and three inactive devices. The system is considered stationary and does not come in close human contact during intended use. 3. Applicable Directive and Standards This declaration is issued to demonstrate compliance with the RF exposure requirements of RSS-102 Issue 6, published by Innovation, Science and Economic Development (ISED) Canada. The device qualifies for SAR exemption under Section 6.3 of RSS-102 Issue 6, based on its low output power. The following standards have been applied: Standard Title Remark RSS-102 Issue 6 Radio Frequency (RF) Exposure Compliance of Radiocommunication Apparatus (All Frequency Bands) Used for RF exposure assessment RSS-GEN General Requirements for Compliance of Radio Apparatus Used for electrical and mechanical safety DOC ID: DN03049/01 4. Assessment Summary • The Bluetooth radio operates in the 2.4 GHz ISM band with a maximum EIRP of approximately +2.5 dBm (≈ 1.8 mW). • According to RSS-102 Issue 6 Section 6.3, devices operating at such low power levels are exempt from SAR testing. • The device meets the exemption limits defined for uncontrolled environments. Therefore, the equipment is exempt from SAR testing in accordance with RSS-102 Issue 6. 5. Conclusion Based on the evaluation above, the equipment complies with the RF exposure requirements of RSS- 102 Issue 6. No Specific Absorption Rate (SAR) testing is required for this product. 6. Signatures Place and Date of Issue: Copenhagen, Denmark – 16-Feb 2026 Name and Title: Jonas Svendsen, Hardware Engineering Manager Signature: ______________________________ Jonas Svendsen
REPORT Testing & Reliability Services Accredited by DANAK under registration number 563 to testing. TEST: Testing laboratory (DS/EN ISO/IEC 17025) FCC reg. no.: DK0002 ISED: CABID: DK0001, Company Number: 10247A Hammerholmen 45A 2650 Hvidovre Denmark +45 7070 1499 [email protected] Date Reference Page 2025-10-20 P24-0097-6 1 (26) DANAK is the national accreditation body in Denmark in compliance with EU regulation No. 765/2008. DANAK participates in the multilateral agreements for testing and calibration under European co-operation for Accreditation (EA) and under International Laboratory Accreditation Cooperation (ILAC) based on peer-evaluation. Accredited test reports issued by laboratories accredited by DANAK are recognized cross border by members of EA and ILAC equal to test reports issued by these members’ accredited laboratories. The use of the accreditation mark on test reports, documents that the service is provided as an accredited service under the company's DANAK accreditation. EKTOS Testing & Reliability Services A/S FCC and ISED Test Report of 2000-1 Fitting Unit 2000-1 W-REM Probe (Wireless-REM Probe) according to FCC 47 CFR, Part 15 Subpart B, Class B ICES‑003, Issue 7:2020, Class B Report prepared by Approved by Bjørn L. Larsen Test Engineer, M. Sc. EE Johan Palmqvist Senior Test Engineer. REPORT Date Reference Page 2025-10-20 P24-0097-6 2 (26) EKTOS Testing & Reliability Services A/S Report no.: P24-0097-6 Report date: 2025-10-20 Test started: 2024-06-19 Test ended: 2024-06-27 Test laboratory: EKTOS TRS A/S Hammerholmen 45A 2650 Hvidovre Denmark Client: Auditdata A/S Wildergade 10B DK-1408 Copenhagen Denmark Contact person: Henrik Brosbøl Contact person: Jonas Svendsen Facility reg. no. FCC registration number: DK0002 ISED: CABID: DK0001, Company Number: 10247A Test specimens: Model: 2000-1 Fitting Unit P/N: 00887 Model: Wireless REM Probe P/N: 00988 Test specifications: FCC 47 CFR Part 15 Subpart B ICES‑003, Issue 7:2020 The tests relevant for the test specimens are listed in section 1.1. Documentation: This test report shall not be reproduced except in full, without written approval of the laboratory. The complete test documentation is archived for 10 years at the testing laboratory. Test results: The test specimen complies with relevant parts of the test specifications. The test results relate only to the specimen tested. Test personnel: Johan Palmqvist Bjørn L. Larsen REPORT Date Reference Page 2025-10-20 P24-0097-6 3 (26) EKTOS Testing & Reliability Services A/S CONTENTS 1 SUMMARY ......................................................................................................... 4 1.1 Test plan, 2000-1 Fitting Unit and 2000-1 W-REM Probe (Wireless REM Probe) ................... 4 1.2 Test Specimens ......................................................................................................................... 5 1.3 Auxiliary equipment ................................................................................................................... 8 1.4 I/O ports / cables to test specimens, ....................................................................................... 14 1.5 Test setup ................................................................................................................................ 14 2 TESTS .............................................................................................................. 15 2.1 Radiated emission ................................................................................................................... 15 2.2 Conducted emission ................................................................................................................ 23 3 MEASURING UNCERTAINTIES ...................................................................... 26 3.1 EMC......................................................................................................................................... 26 Appendices None REPORT Date Reference Page 2025-10-20 P24-0097-6 4 (26) EKTOS Testing & Reliability Services A/S 1 SUMMARY 1.1 Test plan, 2000-1 Fitting Unit and 2000-1 W-REM Probe (Wireless REM Probe) Test method Name of the test Results FCC 47 CFR Part 15 Subpart B, Class B ICES‑003, Issue 7:2020, Class B Radiated emission PASSED FCC 47 CFR Part 15 Subpart B, Class B ICES‑003, Issue 7:2020, Class B Conducted emission PASSED PASSED The test was performed and the test specimen complies with the essential requirements in the standard. FAILED The test was performed and the test specimen does not comply with the essential requirements in the standard. REF The test is covered by a test in another report and/or on a similar test specimen. NR The test is not relevant for the test specimen or has been waived by the manufacturer. REPORT Date Reference Page 2025-10-20 P24-0097-6 5 (26) EKTOS Testing & Reliability Services A/S 1.2 Test Specimens 1.2.1 2001-1 Fitting Unit Manufacturer Auditdata Model 2000-1 Fitting Unit Serial no. /UID 33100006, 33100008 Part no. 00887 Firmware V4.1.0.0 0308 Revision (HW) 3.00 Supply voltage 100 - 240 VAC 50-60 Hz 600-300 mA from AC/DC adapter supply 15 VDC/ 2A.. Operational mode Standby/Ready-For-Use and Active Photo 1. 2000-1 Fitting Unit. REPORT Date Reference Page 2025-10-20 P24-0097-6 6 (26) EKTOS Testing & Reliability Services A/S 1.2.2 2000-1 Fitting Unit – Charger (AC/DC Adapter) Manufacturer FRIWO Gerätebau GmbH Model FW8030M/15 Serial no. - Part no. 100.111 Details Switching Power Supply, Medical Adapter. Normally only used if free field speakers or higher output levels are required. Supply voltage 100 - 240 VAC 50-60 Hz 600-300 mA 120 VAC 50 Hz is used in tests unless otherwise specified Output Voltage 15.0 VDC / 2.0 A Operational mode AC/DC charger for Fitting Unit Photo 2. 2000-1 Fitting Unit – AC/DC Charger. REPORT Date Reference Page 2025-10-20 P24-0097-6 7 (26) EKTOS Testing & Reliability Services A/S 1.2.3 2000-1 W-REM Probe Manufacturer Auditdata Model Wire…
Text truncated - open the document above for the full version.
REPORT Testing & Reliability Services Accredited by DANAK under registration number 563 to testing. TEST: Testing laboratory (DS/EN ISO/IEC 17025) FCC reg. no.: DK0002 ISED: CABID: DK0001, Company Number: 10247A Hammerholmen 45A 2650 Hvidovre Denmark +45 7070 1499 [email protected] Date Reference Page 2025-10-20 P24-0097-6 1 (26) DANAK is the national accreditation body in Denmark in compliance with EU regulation No. 765/2008. DANAK participates in the multilateral agreements for testing and calibration under European co-operation for Accreditation (EA) and under International Laboratory Accreditation Cooperation (ILAC) based on peer-evaluation. Accredited test reports issued by laboratories accredited by DANAK are recognized cross border by members of EA and ILAC equal to test reports issued by these members’ accredited laboratories. The use of the accreditation mark on test reports, documents that the service is provided as an accredited service under the company's DANAK accreditation. EKTOS Testing & Reliability Services A/S FCC and ISED Test Report of 2000-1 Fitting Unit 2000-1 W-REM Probe (Wireless-REM Probe) according to FCC 47 CFR, Part 15 Subpart B, Class B ICES‑003, Issue 7:2020, Class B Report prepared by Approved by Bjørn L. Larsen Test Engineer, M. Sc. EE Johan Palmqvist Senior Test Engineer. REPORT Date Reference Page 2025-10-20 P24-0097-6 2 (26) EKTOS Testing & Reliability Services A/S Report no.: P24-0097-6 Report date: 2025-10-20 Test started: 2024-06-19 Test ended: 2024-06-27 Test laboratory: EKTOS TRS A/S Hammerholmen 45A 2650 Hvidovre Denmark Client: Auditdata A/S Wildergade 10B DK-1408 Copenhagen Denmark Contact person: Henrik Brosbøl Contact person: Jonas Svendsen Facility reg. no. FCC registration number: DK0002 ISED: CABID: DK0001, Company Number: 10247A Test specimens: Model: 2000-1 Fitting Unit P/N: 00887 Model: Wireless REM Probe P/N: 00988 Test specifications: FCC 47 CFR Part 15 Subpart B ICES‑003, Issue 7:2020 The tests relevant for the test specimens are listed in section 1.1. Documentation: This test report shall not be reproduced except in full, without written approval of the laboratory. The complete test documentation is archived for 10 years at the testing laboratory. Test results: The test specimen complies with relevant parts of the test specifications. The test results relate only to the specimen tested. Test personnel: Johan Palmqvist Bjørn L. Larsen REPORT Date Reference Page 2025-10-20 P24-0097-6 3 (26) EKTOS Testing & Reliability Services A/S CONTENTS 1 SUMMARY ......................................................................................................... 4 1.1 Test plan, 2000-1 Fitting Unit and 2000-1 W-REM Probe (Wireless REM Probe) ................... 4 1.2 Test Specimens ......................................................................................................................... 5 1.3 Auxiliary equipment ................................................................................................................... 8 1.4 I/O ports / cables to test specimens, ....................................................................................... 14 1.5 Test setup ................................................................................................................................ 14 2 TESTS .............................................................................................................. 15 2.1 Radiated emission ................................................................................................................... 15 2.2 Conducted emission ................................................................................................................ 23 3 MEASURING UNCERTAINTIES ...................................................................... 26 3.1 EMC......................................................................................................................................... 26 Appendices None REPORT Date Reference Page 2025-10-20 P24-0097-6 4 (26) EKTOS Testing & Reliability Services A/S 1 SUMMARY 1.1 Test plan, 2000-1 Fitting Unit and 2000-1 W-REM Probe (Wireless REM Probe) Test method Name of the test Results FCC 47 CFR Part 15 Subpart B, Class B ICES‑003, Issue 7:2020, Class B Radiated emission PASSED FCC 47 CFR Part 15 Subpart B, Class B ICES‑003, Issue 7:2020, Class B Conducted emission PASSED PASSED The test was performed and the test specimen complies with the essential requirements in the standard. FAILED The test was performed and the test specimen does not comply with the essential requirements in the standard. REF The test is covered by a test in another report and/or on a similar test specimen. NR The test is not relevant for the test specimen or has been waived by the manufacturer. REPORT Date Reference Page 2025-10-20 P24-0097-6 5 (26) EKTOS Testing & Reliability Services A/S 1.2 Test Specimens 1.2.1 2001-1 Fitting Unit Manufacturer Auditdata Model 2000-1 Fitting Unit Serial no. /UID 33100006, 33100008 Part no. 00887 Firmware V4.1.0.0 0308 Revision (HW) 3.00 Supply voltage 100 - 240 VAC 50-60 Hz 600-300 mA from AC/DC adapter supply 15 VDC/ 2A.. Operational mode Standby/Ready-For-Use and Active Photo 1. 2000-1 Fitting Unit. REPORT Date Reference Page 2025-10-20 P24-0097-6 6 (26) EKTOS Testing & Reliability Services A/S 1.2.2 2000-1 Fitting Unit – Charger (AC/DC Adapter) Manufacturer FRIWO Gerätebau GmbH Model FW8030M/15 Serial no. - Part no. 100.111 Details Switching Power Supply, Medical Adapter. Normally only used if free field speakers or higher output levels are required. Supply voltage 100 - 240 VAC 50-60 Hz 600-300 mA 120 VAC 50 Hz is used in tests unless otherwise specified Output Voltage 15.0 VDC / 2.0 A Operational mode AC/DC charger for Fitting Unit Photo 2. 2000-1 Fitting Unit – AC/DC Charger. REPORT Date Reference Page 2025-10-20 P24-0097-6 7 (26) EKTOS Testing & Reliability Services A/S 1.2.3 2000-1 W-REM Probe Manufacturer Auditdata Model Wire…
Text truncated - open the document above for the full version.
REPORT Testing & Reliability Services Accredited by DANAK under registration number 563 to testing. TEST: Testing laboratory (DS/EN ISO/IEC 17025) FCC reg. no.: DK0002 ISED CAB identifier: DK0001 Hammerholmen 45A 2650 Hvidovre Denmark +45 7070 1499 [email protected] Date Reference Page 2025-03-20 P24-0097-2 1 (115) DANAK is the national accreditation body in Denmark in compliance with EU regulation No. 765/2008. DANAK participates in the multilateral agreements for testing and calibration under European co-operation for Accreditation (EA) and under International Laboratory Accreditation Cooperation (ILAC) based on peer-evaluation. Accredited test reports issued by laboratories accredited by DANAK are recognized cross border by members of EA and ILAC equal to test reports issued by these members’ accredited laboratories. The use of the accreditation mark on test reports, documents that the service is provided as an accredited service under the company's DANAK accreditation. EKTOS Testing & Reliability Services A/S Test Report of Model: 2000-1 Fitting Unit. Part no: PN00887 FCC ID: 2BK6Q-PN00887 ISED ID: 33071-PN00887 according to FCC 47 CFR, Part 15 Subpart C 15.247 Operation within the band 2400 to 2483.5 MHz and ISED RSS-247 Issue 3 Aug. 2023 Performed by Examined by Søren Søltoft Senior EMC Engineer Volodymyr Hraivoronskyi Test Engineer, M. Sc. EE. REPORT Date Reference Page FCC ID: 2BK6Q-PN00887 2025-03-20 P24-0097-2 2 (115) ISED ID: 33071-PN00887 EKTOS Testing & Reliability Services A/S Report no.: P24-0097-2 Report date: 2025-03-20 Test started: 2024-10-01 Test ended: 2025-03-14 Test laboratory: EKTOS TRS A/S Hammerholmen 45A 2650 Hvidovre Denmark Client: Auditdata A/S Wildersgade 10B 1408 Copenhagen Denmark Contact person: Jens Denborg Contact person: Jonas Svendsen Facility reg. no. FCC Designation number: DK0002 ISED CAB identifier: DK0001 Test specimen: Model: 2000-1 Fitting Unit. Part no: PN00887. Test specification: FCC 47 CFR Part 15 Subpart C 15.247 Operation within the band 2400 to 2483.5 MHz. ISED RSS-247 Issue 3 Aug. 2023. The tests relevant for the test specimens are listed in section 1.1. Documentation: This test report must always be reproduced in full; reproduction of an excerpt only is subject to written approval of the testing laboratory. The complete test documentation is archived for 10 years at the testing laboratory. Test results: The test specimen complies with relevant parts of the test specifications. The test results relate only to the specimen tested. Test personnel: Søren Søltoft REPORT Date Reference Page FCC ID: 2BK6Q-PN00887 2025-03-20 P24-0097-2 3 (115) ISED ID: 33071-PN00887 EKTOS Testing & Reliability Services A/S CONTENTS 1 SUMMARY ......................................................................................................... 4 1.1 Test plan .................................................................................................................................... 4 1.2 Test specimens ......................................................................................................................... 5 1.3 Auxiliary Equipment................................................................................................................... 6 2 TESTS ................................................................................................................ 7 2.1 Carrier frequency separation ..................................................................................................... 7 2.2 Number of hopping frequencies .............................................................................................. 10 2.3 Time of occupation .................................................................................................................. 13 2.4 Output power ........................................................................................................................... 18 2.5 Occupied bandwidth ................................................................................................................ 23 2.6 Conducted spurious emission ................................................................................................. 37 2.7 Conducted band edge ............................................................................................................. 66 2.8 Restricted-band band-edge. .................................................................................................... 73 2.9 Emission in Restricted bands .................................................................................................. 77 2.10 AC power-line conducted emission ....................................................................................... 109 2.11 Frequency stability versus supply voltage and temperature variation .................................. 113 3 MEASURING UNCERTAINTIES ......…
Text truncated - open the document above for the full version.
| # | Rule Parts | Frequency Range | Power Output | Emission | Tolerance |
|---|---|---|---|---|---|
| 1 | 15C | 2.40 GHz - 2.48 GHz | 1.74 mW | 1M00F1D | 50.0000000000Hz |
The 2000-1 W-REM Probe is a wireless Probe used for measuring hearing loss
Equipment Class
DSS - Part 15 Spread Spectrum Transmitter