Intelligent Locations Hand-Hygiene / Location Extender (ILAT-HW350) User Manual Introduction: Intelligent Locations’ hand-hygiene / location extender tracks hand hygiene protocols in real-time to reduce the risk of infection, protect healthcare staff & patients while aiding in the delivery of quality care. Multipurpose device with the initial intent to help with handwashing compliance. The device scans nearby BLE device from Intelligent Location's product catalog. It broadcasts identifier and RSSI of scanned devices, to get an idea of its proximity. The device includes a voice chip and speaker to allow instructions, warnings, and other applications to be sent to people present. The specifications for this device are as follows: CONNECTIVITY Bluetooth LE 5.2(direction finding) CHIPSET: Silicon Labs BG24 • 32-bit 78.0 MHz ARM Cortex®-M33 FEATURES Voice chip and speaker to allow instructions, warnings, and other applications to be sent to people present BATTERY CAPACITY: Rechargeable 7.4V Lithium Ion Battery PHYSICAL DIMENSIONS: • L ⎯ 3.3 in (84.20 mm) • W ⎯ 3.3 in (84.20 mm) H ⎯ 1.3 in (34.67 mm) These devices are shipped in bulk to a customer site location. The Intelligent Locations Account team is responsible for installing these devices. Additional Information This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. Class B: NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radi-ate radio frequency energy and, if not in- stalled and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: —Reorient or relocate the receiving antenna. —Increase the separation between the equipment and receiver. —Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. —Consult the dealer or an experienced radio/ TV technician for help.
2025-12- 08 Eurofins Electrical and Electronic Testing NA, Inc. 914 West Patapsco Avenue Baltimore, MD 21230 RE: Attestation Statements Part 2.911(d)(5)(i) request for FCC ID : 2BTDPILATHW350 [ Intelligent Locations ] certifies that the equipment for which authorization is sought is not “covered” equipment prohibited from receiving an equipment authorization pursuant to section 2.903 of the FCC rules. We also certify that, as of the date of the filing of the application, the equipment does not utilize cyber-security or anti-virus software from Kaspersky Lab, Inc, or any of its successors and assignees, which is also on the “covered list”. Sincerely, Signature: Name: Daniel Mogin 2025-12-08 Eurofins Electrical and Electronic Testing NA, Inc. 914 West Patapsco Avenue Baltimore, MD 21230 RE: Attestation Statements Part 2.911(d)(5)(ii) request for FCC ID : 2BTDPILATHW350 Intelligent Locations certifies that, as of the date of the filing of the application, the applicant is not identified on the Covered List as an entity producing “covered” equipment. Sincerely, Signature: Name: Daniel Mogin
U.S. Agent Designation for Service of Process (Applicant and U.S. Agent are Separate Parties) ATTENTION: FCC Certification – Section 2.911(d)(7) Information Designated U.S. Agent Company Name: Intelligent Locations Address: 8601 Six Forks Rd., Suite 205, Raleig, North Carolina, 27615-5236, United States FRN: 0037688892 Contact Name: Bogdan Nedelcu Telephone No: (847) 809-7172 Email: [email protected] This letter is to confirm that we (“the Designated U.S. Agent) have accepted the responsibility to act as the Designated U.S. Agent for Service of Process as required by section 47 CFR 2.911(d)(7) on behalf of the Applicant noted below. The Applicant acknowledges that they must maintain an agent for no less than one year after terminating all marketing and importation OR the conclusion of any Commission-related proceeding involving the equipment. The applicant further acknowledges their responsibility to inform the FCC whenever the Designated U.S. Agent information above changes. Applicant Company name: Intelligent Locations FRN: 0037688892 Grantee Code: 2BTDP FCC ID: 2BTDPILATHW350 Contact Name: Bogdan Nedelcu Address: 8601 Six Forks Rd., Suite 205, Raleig, North Carolina, 27615-5236, United States Telephone No: (847) 809-7172 Email: [email protected] Signature: Date: 2025-12-08 Signed by (Printed Name/Title): Daniel Mogin
Date: 2025-12-08 Eurofins Electrical and Electronic Testing NA, Inc. 914 West Patapsco Avenue Baltimore, MD 21230 RE: CONFIDENTIALITY REQUEST FOR ( Hand-Hygiene Location Extender / ILAT-HW350 / FCC ID: 2BTDPILATHW350) To Whom It May Concern: This letter serves as an official request for confidentiality under sections 0.457 and 0.459 of CFR 47. We have requested that the ( Block Diagrams, Schematics, Operational Descriptions) required to be submitted with this application be permanently withheld from public review. The above documents are company proprietary information that could never get into the possession of your competition. Please contact me if there is any information you may need. Sincerely, Signature: Name: Daniel Mogin Applicant: Intelligent Locations Address: 8601 Six Forks Rd., Suite 205, Raleig, North Carolina, 27615-5236, United States
Date: 2026-04-20 Eurofins Electrical and Electronic Testing NA, Inc. 914 West Patapsco Avenue Baltimore, MD 21230 RE: LETTER OF AGENT AUTHORIZATION To Whom It May Concern: We, the undersigned, hereby authorize Daniel Mogin to act on our behalf in all matters relating to application for equipment authorization, including the signing of all documents relating to these matters. We also hereby certify that no party to the application authorized hereunder is subject to the denial of benefits, including FCC benefits, pursuant to Section 5301 of the Anti-Drug Abuse Act of 1988, 21 U.S.C.853(a). This agreement expires one year from the current date. Sincerely, Signature: Name: Bogdan Nedelcu Applicant: Intelligent Locations Address: 8601 Six Forks Rd., Suite 205, Raleig, North Carolina, 27615-5236, United States
Radio frequency radiation exposure evaluation: portable devices RESULT : Pass Test Specification Test item : Hand-Hygiene Location Extender Identification / Type No. : ILAT-HW350 FCC ID : 2BTDPILATHW350 Test standard : CFR47 FCC Part 2: Section 2.1093 FCC KDB Publication 447498 D04 Measurement Record: Since maximum peak output power of the transmitter is -5.68dBm ≈ 0.27mW < 3 mW. Hence the EUT is excluded from SAR evaluation according to FCC KDB Publication 447498 D04
MORLAB ShenzhenMorlabCommunicationsTechnologyCo.,Ltd.,FL.1-3,BuildingA, FeiYangSciencePark,No.8LongChangRoad,Block67,BaoAnDistrict, ShenZhen,GuangDongProvince,P.R.China Tel:86-755-36698555Fax:86-755-36698525 Http://www.morlab.cnE-mail:[email protected] Page1of 3 REPORTNo.:SZ26020117E01 AnnexDEUTPhotos 1.Testenvironment MORLAB ShenzhenMorlabCommunicationsTechnologyCo.,Ltd.FL.1-3,BuildingA, FeiYangSciencePark,No.8LongChangRoad,Block67,BaoAnDistrict, ShenZhen,GuangDongProvince,P.R.China Tel:86-755-36698555Fax:86-755-36698525 Http://www.morlab.cnE-mail:[email protected] Page2of3 REPORTNo.:SZ26020117E01 2.EUT MORLAB ShenzhenMorlabCommunicationsTechnologyCo.,Ltd.FL.1-3,BuildingA, FeiYangSciencePark,No.8LongChangRoad,Block67,BaoAnDistrict, ShenZhen,GuangDongProvince,P.R.China Tel:86-755-36698555Fax:86-755-36698525 Http://www.morlab.cnE-mail:[email protected] Page3of3 REPORTNo.:SZ26020117E01
Appendix A: DTS Bandwidth Test Result TestMode Antenna Frequency[MHz] DTS BW [MHz] FL[MHz] FH[MHz] Limit[MHz] Verdict BLE_1M Ant1 2402 0.660 2401.660 2402.320 0.5 PASS 2440 0.664 2439.660 2440.324 0.5 PASS 2480 0.648 2479.664 2480.312 0.5 PASS Test Graphs BLE_1M_Ant1_2402 BLE_1M_Ant1_2440 BLE_1M_Ant1_2480 Appendix B: Occupied Channel Bandwidth Test Result TestMode Antenna Frequency[MHz] OCB [MHz] FL[MHz] FH[MHz] Limit[MHz] Verdict BLE_1M Ant1 2402 1.0811 2401.4660 2402.5471 --- --- 2440 1.2271 2439.3899 2440.6170 --- --- 2480 1.3133 2479.3568 2480.6701 --- --- Test Graphs BLE_1M_Ant1_2402 BLE_1M_Ant1_2440 BLE_1M_Ant1_2480 Appendix C: Maximum conducted output power Test Result Peak TestMode Antenna Frequency[MHz] Conducted Peak Powert[dBm] Conducted Limit[dBm] BLE_1M Ant1 2402 -5.68 ≤30 2440 -7.68 ≤30 2480 -9.64 ≤30 Test Graphs Peak BLE_1M_Ant1_2402 BLE_1M_Ant1_2440 BLE_1M_Ant1_2480 Appendix D: Maximum power spectral density Test Result TestMode Antenna Frequency[MHz] Result[dBm/3kHz] Limit[dBm/3kHz] Verdict BLE_1M Ant1 2402 -7.58 ≤8.00 PASS 2440 -10.18 ≤8.00 PASS 2480 -12.25 ≤8.00 PASS Test Graphs BLE_1M_Ant1_2402 BLE_1M_Ant1_2440 BLE_1M_Ant1_2480 Appendix E: Band edge measurements Test Result TestMode Antenna ChName Frequency [MHz] RefLevel[dBm] Result[dBm] Limit[dBm] Verdict BLE_1M Ant1 Low 2402 -6.68 -46.89 ≤-26.68 PASS High 2480 -11.26 -46.82 ≤-31.26 PASS Test Graphs BLE_1M_Ant1_Low_2402 BLE_1M_Ant1_High_2480 Appendix F: Conducted Spurious Emission Test Result TestMode Antenna Frequency [MHz] FreqRange [MHz] RefLevel [dBm] Result[dBm] Limit[dBm] Verdict BLE_1M Ant1 2402 Reference -6.94 -6.94 --- PASS 30~1000 -6.94 -57.29 ≤-26.94 PASS 1000~26500 -6.94 -44.6 ≤-26.94 PASS 2440 Reference -9.48 -9.48 --- PASS 30~1000 -9.48 -58.43 ≤-29.48 PASS 1000~26500 -9.48 -49.7 ≤-29.48 PASS 2480 Reference -11.55 -11.55 --- PASS 30~1000 -11.55 -57.37 ≤-31.55 PASS 1000~26500 -11.55 -47.45 ≤-31.55 PASS Test Graphs BLE_1M_Ant1_2402_0~Reference BLE_1M_Ant1_2402_30~1000 BLE_1M_Ant1_2402_1000~26500 BLE_1M_Ant1_2440_0~Reference BLE_1M_Ant1_2440_30~1000 BLE_1M_Ant1_2440_1000~26500 BLE_1M_Ant1_2480_0~Reference BLE_1M_Ant1_2480_30~1000 BLE_1M_Ant1_2480_1000~26500 Appendix G: Duty Cycle Test Result TestMode Antenna Frequency[MHz] ON Time [ms] Period [ms] Duty Cycle [%] BLE_1M Ant1 2402 0.39 0.63 61.90 2440 0.39 0.63 61.90 2480 0.39 0.63 61.90 Test Graphs BLE_1M_Ant1_2402 BLE_1M_Ant1_2440 BLE_1M_Ant1_2480 Appendix H: Restricted Band Mode: 1M Channel: 2402MHz Test Graph Suspected Data List NO. Frequency [MHz] Reading [dBμV/m] Level [dBμV/m] Factor [dB] Limit [dBμV/m] Margin [dB] Height [cm] Angle [°] Det Polarity Verdict 1 2310.0000 39.83 46.79 6.96 74.00 27.21 150 159 PK Horizontal PASS 2 2320.6400 41.42 48.59 7.17 74.00 25.41 150 4 PK Horizontal PASS 3 2336.0000 40.92 48.02 7.10 74.00 25.98 150 131 PK Horizontal PASS 4 2350.1600 40.53 47.93 7.40 74.00 26.07 150 109 PK Horizontal PASS 5 2364.3200 40.93 48.45 7.52 74.00 25.55 150 347 PK Horizontal PASS 6 2390.0000 40.36 48.64 8.28 74.00 25.36 150 343 PK Horizontal PASS Note:(1)Level=Reading+Factor (2)Margin=Limit-Level Mode: 1M Channel: 2402MHz Test Graph Suspected Data List NO. Frequency [MHz] Reading [dBμV/m] Level [dBμV/m] Factor [dB] Limit [dBμV/m] Margin [dB] Height [cm] Angle [°] Det Polarity Verdict 1 2310.0000 40.04 47.00 6.96 74.00 27.00 150 9 PK Vertical PASS 2 2319.9200 41.59 48.77 7.18 74.00 25.23 150 263 PK Vertical PASS 3 2332.6400 41.64 48.75 7.11 74.00 25.25 150 338 PK Vertical PASS 4 2342.7200 41.67 48.88 7.21 74.00 25.12 150 245 PK Vertical PASS 5 2363.0400 42.08 49.53 7.45 74.00 24.47 150 185 PK Vertical PASS 6 2390.0000 39.70 47.98 8.28 74.00 26.02 150 9 PK Vertical PASS Note:(1)Level=Reading+Factor (2)Margin=Limit-Level Mode: 1M Channel: 2480MHz Test Graph Final Data List NO. Frequency [MHz] Factor [dB] PK Reading [dBμV/ m] PK Value [dBμV/m] PK Limit [dBμV/m] PK Margin [dB] AV Reading [dBμV/ m] AV Value [dBμV/m] AV Limit [dBμV/m] AV Margin [dB] Height [cm] Angle [°] Polarity Verdict 1 2483.5000 7.53 49.55 57.08 74.00 16.92 42.17 49.70 54.00 4.30 150 330 Horizontal PASS 2 2484.8695 7.60 52.47 60.07 74.00 13.93 41.15 48.75 54.00 5.25 150 341 Horizontal PASS 3 2487.2125 7.60 49.1 56.70 74.00 17.30 40.55 48.15 54.00 5.85 150 270 Horizontal PASS 4 2490.0340 7.65 48.64 56.29 74.00 17.71 36.02 43.67 54.00 10.33 150 323 Horizontal PASS 5 2492.9050 7.66 48.76 56.42 74.00 17.58 35.68 43.34 54.00 10.66 150 68 Horizontal PASS 6 2500.0000 7.60 47.32 54.92 74.00 19.08 35.10 42.70 54.00 11.30 150 309 Horizontal PASS Note:(1)Level=Reading+Factor (2)Margin=Limit-Level Mode: 1M Channel: 2480MHz Test Graph Final Data List NO. Frequency [MHz] Factor [dB] PK Reading [dBμV/ m] PK Value [dBμV/m] PK Limit [dBμV/m] PK Margin [dB] AV Reading [dBμV/ m] AV Value [dBμV/m] AV Limit [dBμV/m] AV Margin [dB] Height [cm] Angle [°] Polarity Verdict 1 2483.5000 7.53 47.53 55.06 74.00 18.94 40.63 48.16 54.00 5.84 150 217 Vertical PASS 2 2485.2985 7.59 48.85 56.44 74.00 17.56 38.05 45.64 54.00 8.36 150 328 Vertical PASS 3 2487.4600 7.58 48.72 56.30 74.00 17.70 35.37 42.95 54.00 11.05 150 135 Vertical PASS 4 2490.1660 7.65 48.71 56.36 74.00 17.64 35.31 42.96 54.00 11.04 150 342 Vertical PASS 5 2492.9710 7.66 48.41 56.07 74.00 17.93 34.36 42.02 54.00 11.98 150 121 Vertical PASS 6 2500.0000 7.60 47.19 54.79 74.00 19.21 35.25 42.85 54.00 11.15 150 338 Vertical PASS Note:(1)Level=Reading+Factor (2)Margin=Limit-Level Appendix I: Spurious emissions Final Data List NO. Freq. [MHz] Factor [dB/m] QP Value [dBμV/m] QP Limit [dBμV/m] QP Margin [dB] Height [cm] Angle [°] Polarity Verdic t 1 43.7012 -14.12 20.74 40.00 19.26 100 181 Vertical PASS 2 51.0975 -14.47 26.99 40.00 13.01 200 34 Vertical PASS 3 585.6888 -7.12 25.87 46.00 20.13 100 248 Vertical PASS 4 713.4862 -4.88 24.72 46.00 21.28 200 308 Vertical PASS 5 837.8888 -2.65 27.61 46.00 18.39 100 70 Vertical PASS 6 946.6500 0.23 30.12 46.00 15.88 200 312 Vertical PASS Final Data List NO. Freq. [MHz] Factor [dB/m] QP Value [dBμV/m] QP Limit […
Text truncated - open the document above for the full version.
EUROFINS ELECTRICAL TESTING SERVICE (SHENZHEN) CO., LTD. RADIO TEST - REPORT FCC Compliance Test Report for Product name: Hand-Hygiene Location Extender Model name: ILAT-HW350 FCC ID: 2BTDPILATHW350 Test Report Number: EFGX25090520-IE-08-E01 Eurofins Electrical Testing Service (Shenzhen) Co., Ltd. Phone: +86-0755-829118671 Room 20 of 2/F., 1/F., Building 2, Spring Block & Room 102, Building 8 Fax: +86-0755-82910749 Autumn Block, Meishenghuigu Innovation Park, No.83, Dabao Road, Bao’an District, Shenzhen, Guangdong, China Page 1 of 26 1 General Information 3 1.1 Notes 3 1.2 Testing laboratory 4 1.3 Details of applicant 4 1.4 Details of manufacturer 4 1.5 Application details 5 1.6 Test item 5 1.7 Test standards 5 2 Technical test 6 2.1 Summary of test results 6 2.2 Test environment 6 2.3 Measurement uncertainty 6 2.4 Test mode 7 2.5 Test equipment utilized 8 2.6 Auxiliary equipment used during test 9 2.7 Test software information 9 2.8 Test setup 10 2.9 Test results 12 3 Technical Requirement 13 3.1 Conducted emission AC power port 13 3.2 Duty cycle 14 3.3 6dB bandwidth 15 3.4 99% Occupied Bandwidth 16 3.6 Conducted Peak output power 17 3.7 Power spectral density 18 3.8 Conducted Spurious Emissions 19 3.9 Band edge 20 3.10 Spurious radiated emissions for transmitter 23 4 Test Setup Photos 25 5 External Photo 25 6 Internal Photos 25 7 Appendix 25 Test Report No.: EFGX25090520-IE-08-E01 Eurofins Electrical Testing Service (Shenzhen) Co., Ltd. Room 20 of 2/F., 1/F., Building 2, Spring Block & Room 102, Building 8, Autumn Block, Meishenghuigu Innovation Park, No.83, Dabao Road, Bao’an District, Shenzhen, Guangdong, China Page 3 of 25 1 General Information 1.1 Notes The results of this test report relate exclusively to the item tested as specified in chapter “Description of test item” and are not transferable to any other test items. Eurofins Electrical Testing Service (Shenzhen) Co., Ltd. is not responsible for any generalisations and conclusions drawn from this report. Any modification of the test item can lead to invalidity of test results and this test report may therefore be not applicable to the modified test item. The test report may only be reproduced or published in full. Reproducing or publishing extracts of the report requires the prior written approval of the Eurofins Electrical Testing Service (Shenzhen) Co., Ltd. This document is subject to the General Terms and Conditions and the Testing and Certification System of Eurofins Electrical Testing Service (Shenzhen) Co., Ltd., available on request or accessible at www.eurofins.com. Operator: 2025-12-08 Bruce Zheng / Project Engineer Date Eurofins-Lab. Name / Title Signature Technical responsibility for area of testing: 2025-12-08 Albert Xu / Lab Manager Date Eurofins-Lab. Name / Title Signature Test Report No.: EFGX25090520-IE-08-E01 Eurofins Electrical Testing Service (Shenzhen) Co., Ltd. Room 20 of 2/F., 1/F., Building 2, Spring Block & Room 102, Building 8, Autumn Block, Meishenghuigu Innovation Park, No.83, Dabao Road, Bao’an District, Shenzhen, Guangdong, China Page 4 of 25 1.2 Testing laboratory Eurofins Electrical Testing Service (Shenzhen) Co., Ltd. Room 20 of 2/F., 1/F., Building 2, Spring Block & Room 102, Building 8, Autumn Block, Meishenghuigu Innovation Park, No.83, Dabao Road, Bao’an District, Shenzhen, Guangdong, China Telephone : +86-755-82911867 Fax : +86-755-82910749 The Laboratory has passed the Accreditation by the American Association for Laboratory Accrediation (A2LA). The Accreditation number is 5376.01 The Laboratory has been listed by industry Canada to perform electromagnetic emission measurements, The CAB identifier is CN0088 1.3 Details of applicant Name : Intelligent Locations Address : 8601 Six Forks Rd., Suite 205, Raleig, North Carolina, 27615-5236, United States Telephone : ./. Fax : ./. 1.4 Details of manufacturer Name : Kaertech Limited Address : Suite 1601-02. Seaview Commercial Building, 21-24 Connaught Road West, Hong Kong Telephone : ./. Fax : ./. Test Report No.: EFGX25090520-IE-08-E01 Eurofins Electrical Testing Service (Shenzhen) Co., Ltd. Room 20 of 2/F., 1/F., Building 2, Spring Block & Room 102, Building 8, Autumn Block, Meishenghuigu Innovation Park, No.83, Dabao Road, Bao’an District, Shenzhen, Guangdong, China Page 5 of 25 1.5 Application details Date of receipt of application : 2025-09-27 Date of receipt of test item : 2025-09-27 Date of test : 2025-09-27 to 2025-12-08 Date of issue : 2025-12-08 1.6 Test item Product type : Hand-Hygiene Location Extender Test Model : ILAT-HW350 Sample ID : 250927-11-003 Brand : ./. Serial number : ./. Hardware Version : 3.5 Software / Firmware Version : 1.9 Ratings : Rechargeable 7.4V Lithium Ion Battery 120V~ 60Hz for Adapter Test voltage : DC 7.4V RadioTechnical data Frequency range : 2402MHz – 2480MHz Radio Tech. : Bluetooth Low Energy Frequency channel : 40 Channels Modulation : GFSK Antenna type : PCB antenna Maximum antenna gain : -6.99dBi Additional information : ./. The above sample(s) and sample information was/were submitted and identified on behalf of the applicant. Eurofins assures objectivity and impartiality of the test, and fulfills the obligation of confidentiality for applicant’s commercial infor- mation and technical documents. 1.7 Test standards Test Standards FCC Part 15 Subpart C PART 15 - RADIO FREQUENCY DEVICES Subpart C - Intentional Radiators Test Method 1: ANSI C63.4-2014, American National Standard for Methods of Measurement of Radio-Noise Emis- sions from Low-Voltage Electrical and Electronic Equipment in the Range of 9 kHz to 40 GHz. 2: ANSI C63.10-2020, American National Standard for Testing Unlicensed Wireless Devices. 3: KDB558074 D01 15.247 Meas Guidance v05r02 Test Report No.: EFGX25090520-IE-08-E01 Eurofins Electrical Testing Service (Shenzhen) Co., Ltd. Room 20 of 2/F., 1/F., Building 2, Spring Block & Room 102, Building 8, Autumn Block, Meishenghuigu Innovation Park, No.83, Dabao Road, Bao’an District, Shenzhen, Guan…
Text truncated - open the document above for the full version.
NOTE:ThisdocumentisissuedbyShenzhenMorlabCommunicationsTechnologyCo.,Ltd.,thetestreportshallnotbereproducedexceptin fullwithoutpriorwrittenpermissionofthecompany.Thetestresultsapplyonlytotheparticularsample(s)testedandtothespecifictestscarried outwhichisavailableonrequestforvalidationandinformationconfirmedatourwebsite. MORLAB ShenzhenMorlabCommunicationsTechnologyCo.,Ltd.,FL.1-3,BuildingA, FeiYangSciencePark,No.8LongChangRoad,Block67,BaoAnDistrict, ShenZhen,GuangDongProvince,P.R.China Tel:86-755-36698555Fax:86-755-36698525 Http://www.morlab.cnE-mail:[email protected] Page1of 11 REPORTNo.:SZ26020117E01 TESTREPORT APPLICANT:IntelligentLocations PRODUCTNAME:PCBAntenna MODELNAME:ILAT-HW350 TRADENAME:N/A BRANDNAME:N/A STANDARD(S):IEEEStd149-2021 RECEIPTDATE:2026-02-09 TESTDATE:2026-03-02 ISSUEDATE:2026-04-20 Editedby: MaoRujie(Rapporteur) Approvedby: ChiShide(Supervisor) MORLAB ShenzhenMorlabCommunicationsTechnologyCo.,Ltd.FL.1-3,BuildingA, FeiYangSciencePark,No.8LongChangRoad,Block67,BaoAnDistrict, ShenZhen,GuangDongProvince,P.R.China Tel:86-755-36698555Fax:86-755-36698525 Http://www.morlab.cnE-mail:[email protected] Page2of11 REPORTNo.:SZ26020117E01 DIRECTORY 1.TechnicalInformation ........................................................................................................3 1.1.ApplicantandManufacturerInformation ....................................................................3 1.2.EquipmentUnderTest(EUT)Description...................................................................3 2.TestResults........................................................................................................................4 2.1.AppliedReferenceDocuments.....................................................................................4 2.2.TestConditions...............................................................................................................4 2.3.MeasurementUncertainty.............................................................................................4 2.4.TestResultslists.............................................................................................................5 AnnexATestSetupPhotos..................................................................................................6 AnnexBFigures .....................................................................................................................7 1.2DRadiationPattern .........................................................................................................7 2.3DRadiationPattern .........................................................................................................8 AnnexCTestingLaboratoryInformation .........................................................................10 1.IdentificationoftheResponsibleTestingLaboratory ................................................10 2.IdentificationoftheResponsibleTestingLocation....................................................10 3.TestEquipmentUtilized..................................................................................................10 3.1TestEquipments............................................................................................................10 3.2TestSoftware..................................................................................................................10 Notes......................................................................................................................................11 AnnexDEUTPhotos ChangeHistory VersionDateReasonforchange 1.02026-04-20Firstedition Note:thelatestrevisionofthereportsupersedesallpreviousversion. MORLAB ShenzhenMorlabCommunicationsTechnologyCo.,Ltd.FL.1-3,BuildingA, FeiYangSciencePark,No.8LongChangRoad,Block67,BaoAnDistrict, ShenZhen,GuangDongProvince,P.R.China Tel:86-755-36698555Fax:86-755-36698525 Http://www.morlab.cnE-mail:[email protected] Page3of11 REPORTNo.:SZ26020117E01 1.TechnicalInformation Note:Providebyapplicant. 1.1.ApplicantandManufacturerInformation Applicant: IntelligentLocations ApplicantAddress: 8601SixForksRd.,Suite205,Raleig,NorthCarolina,27615-5236, UnitedStates Manufacturer: N/A ManufacturerAddress: N/A 1.2.EquipmentUnderTest(EUT)Description WirelessTypeN/A Frequency2400MHz-2500MHz AntennaTypeN/A ProductHWVersionN/A ProductSWVersionN/A SerialNumberN/A SampleNo.1# MORLAB ShenzhenMorlabCommunicationsTechnologyCo.,Ltd.FL.1-3,BuildingA, FeiYangSciencePark,No.8LongChangRoad,Block67,BaoAnDistrict, ShenZhen,GuangDongProvince,P.R.China Tel:86-755-36698555Fax:86-755-36698525 Http://www.morlab.cnE-mail:[email protected] Page4of11 REPORTNo.:SZ26020117E01 2.TestResults 2.1.AppliedReferenceDocuments Leadingreferencedocumentsfortesting: No.IdentityDocumentTitle 1IEEEStd149-2021 IEEERecommendedPracticeforAntenna Measurements Notes:Anyadditions,deviation,orexclusionsfromthemethodshallbenotedinthe"Remark". 2.2.TestConditions TestEnvironmentConditions : RelativeHumidity(%):25-75 Temperature(°C):10-30 2.3.MeasurementUncertainty Theuncertaintyiscalculatedusingthemethodssuggestedinthe“GuidetotheExpressionof UncertaintyinMeasurement”(GUM)publishedbyISO. MORLAB ShenzhenMorlabCommunicationsTechnologyCo.,Ltd.FL.1-3,BuildingA, FeiYangSciencePark,No.8LongChangRoad,Block67,BaoAnDistrict, ShenZhen,GuangDongProvince,P.R.China Tel:86-755-36698555Fax:86-755-36698525 Http://www.morlab.cnE-mail:[email protected] Page5of11 REPORTNo.:SZ26020117E01 2.4.TestResultslists 2.4.1.Gain Frequency (MHz) Gain(dBi) 2400-8.59 2410-8.55 2420-8.77 2430-8.94 2440-8.72 2450-8.33 2460-7.89 2470-7.65 2480-7.45 2490-7.22 2500-6.99 MORLAB ShenzhenMorlabCommunicationsTechnologyCo.,Ltd.FL.1-3,BuildingA, FeiYangSciencePark,No.8LongChangRoad,Block67,BaoAnDistrict, ShenZhen,GuangDongProvince,P.R.China Tel:86-755-36698555Fax:86-755-36698525 Http://www.morlab.cnE-mail:[email protected] Page6of11 REPORTNo.:SZ26020117E01 AnnexATestSetupPhotos MORLAB ShenzhenMorlabCommunicationsTechnologyCo.,Ltd.FL.1-3,BuildingA, FeiYangSciencePark,No.8LongCha…
Text truncated - open the document above for the full version.
8601 Six Forks Rd., Suite 205 · Raleig, 27615-5236 · United States
| # | Rule Parts | Frequency Range | Power Output |
|---|---|---|---|
| 1 | 15C | 2.40 GHz - 2.48 GHz | 270.30 µW |