Text extracted from the exhibit documents filed with the FCC. Open a document above to read the original.
Registered SC575629, GLUCORX TECHNOLOGIES LTD, DINGWALL BUSINESS CENTRE, STRATHPEFFER ROAD, DINGWALL, IV15 9QF, SCOTLAND GlucoRx Technologies Ltd Dingwall Business Centre, Strathpeffer Road Dingwall, Scotland IV15 9QF Phone: +44 (0)1349204002 05/14/2026 Eurofins Electrical and Electronic Testing NA, Inc. 914 West Patapsco Avenue Baltimore, MD 21230 RE: Attestation Statements Part 2.911(d)(5)(i) request for FCC ID: 2BVYL-MODUS-CGM GlucoRx Technologies Ltd certifies that the equipment for which authorization is sought is not “covered” equipment prohibited from receiving an equipment authorization pursuant to section 2.903 of the FCC rules. We also certify that, as of the date of the filing of the application, the equipment does not utilize cyber-security or anti-virus software from Kaspersky Lab, Inc, or any of its successors and assignees, which is also on the “covered list”. Sincerely, Leanne McIntosh Group QA/RA Lead and Business Unit Manager Registered SC575629, GLUCORX TECHNOLOGIES LTD, DINGWALL BUSINESS CENTRE, STRATHPEFFER ROAD, DINGWALL, IV15 9QF, SCOTLAND GlucoRx Technologies Ltd Dingwall Business Centre, Strathpeffer Road Dingwall, Scotland IV15 9QF Phone: +44 (0)1349204002 05/14/2026 Eurofins Electrical and Electronic Testing NA, Inc. 914 West Patapsco Avenue Baltimore, MD 21230 RE: Attestation Statements Part 2.911(d)(5)(ii) request for FCC ID: 2BVYL-MODUS-CGM GlucoRx Technologies Ltd certifies that, as of the date of the filing of the application, the applicant is not identified on the Covered List as an entity producing “covered” equipment. Sincerely, Leanne McIntosh Group QA/RA Lead and Business Unit Manager
Registered SC575629, GLUCORX TECHNOLOGIES LTD, DINGWALL BUSINESS CENTRE, STRATHPEFFER ROAD, DINGWALL, IV15 9QF, SCOTLAND GlucoRx Technologies Ltd Dingwall Business Centre, Strathpeffer Road Dingwall, Scotland IV15 9QF Phone: +44 (0)1349204002 Attestation Statements 06/07/2026 TO: Federal Communications Commission Authorization and Evaluation Division 7435 Oakland Mills Road Columbia, MD 21046, USA RE: Attestation Statements Part 2.911(d)(7) request for FCC ID: 2BVYL-MODUS-CGM GlucoRx Technologies Ltd, the undersigned, hereby authorize FCC US Agent, LLC to act as our designated U.S. agent for service of process, GlucoRx Technologies Ltd accepts to maintain an agent for no less than one year after the grantee has terminated all marketing and importation or the conclusion of any Commission-related proceeding involving the equipment. FCC US Agent, LLC accepts the responsibility to act as designated U.S. agent for service of process. Applicant Information: Company Name: GlucoRx Technologies Ltd Contact Name: Christopher Cregan Address: Dingwall Business Centre, Strathpeffer Road, Dingwall, IV15 9QF, United Kingdom FRN: 0038386041 Telephone No: 01349 204002 Email: [email protected] Signature Christopher Cregan Registered SC575629, GLUCORX TECHNOLOGIES LTD, DINGWALL BUSINESS CENTRE, STRATHPEFFER ROAD, DINGWALL, IV15 9QF, SCOTLAND GlucoRx Technologies Ltd Dingwall Business Centre, Strathpeffer Road Dingwall, Scotland IV15 9QF Phone: +44 (0)1349204002 U.S. Agent Information: Company Name: FCC US Agent, LLC Address: 2213 S. 3722 Illinois Avenue, Saint Charles, IL, 60174, USA FRN: 0033402884 Contact Name: Tim Payne Telephone No: 708-571-3148 Email: [email protected] Signature Tim Payne Tim Payne (Jul 7, 2026 10:21:04 CDT) Tim Payne
Registered SC575629, GLUCORX TECHNOLOGIES LTD, DINGWALL BUSINESS CENTRE, STRATHPEFFER ROAD, DINGWALL, IV15 9QF, SCOTLAND GlucoRx Technologies Ltd Dingwall Business Centre, Strathpeffer Road Dingwall, Scotland IV15 9QF Phone: +44 (0)1349204002 05/14/2026 Eurofins Electrical and Electronic Testing NA, Inc. 914 West Patapsco Avenue Baltimore, MD 21230 RE: LETTER OF AGENT AUTHORIZATION To Whom It May Concern: We, the undersigned, hereby authorize James Beevers at Eurofins Electrical & Electronic UK Ltd ) to act on our behalf in all matters relating to application for equipment authorization, including the signing of all documents relating to these matters. We also hereby certify that no party to the application authorized hereunder is subject to the denial of benefits, including FCC benefits, pursuant to Section 5301 of the Anti-Drug Abuse Act of 1988, 21 U.S.C.853(a). This agreement expires one year from the current date. Sincerely, Leanne McIntosh Group QA/RA Lead and Business Unit Manager +44 (0)7769 325 150 [email protected]
Registered SC575629, GLUCORX TECHNOLOGIES LTD, DINGWALL BUSINESS CENTRE, STRATHPEFFER ROAD, DINGWALL, IV15 9QF, SCOTLAND GlucoRx Technologies Ltd Dingwall Business Centre, Strathpeffer Road Dingwall, Scotland IV15 9QF Phone: +44 (0)1349204002 07/24/2026 Eurofins Electrical and Electronic Testing NA, Inc. 914 West Patapsco Avenue Baltimore, MD 21230 RE: CONFIDENTIALITY REQUEST FOR MODUS-CGM AND FCC ID: 2BVYL-MODUS-CGM To Whom It May Concern: This letter serves as an official request for confidentiality under sections 0.457 and 0.459 of CFR 47. We have requested that the items listed below are required to be submitted with this application be permanently withheld from public review. • Detailed block diagrams • Schematic diagrams • Detailed operational descriptions • Parts lists/ tune up information The above documents contain proprietary technical information that would not routinely be made available to the public. Disclosure of this information could give competitors an unfair commercial advantage; therefore, confidential treatment is requested. Please contact me if there is any information you may need. Sincerely, Leanne McIntosh Group QA/RA Lead and Business Unit Manager +44 (0)7769 325 150 [email protected]
Registered SC575629, GLUCORX TECHNOLOGIES LTD, DINGWALL BUSINESS CENTRE, STRATHPEFFER ROAD, DINGWALL, IV15 9QF, SCOTLAND GlucoRx Technologies Ltd Dingwall Business Centre, Strathpeffer Road Dingwall, Scotland IV15 9QF Phone: +44 (0)1349204002 05/14/2026 Eurofins Electrical and Electronic Testing NA, Inc. 914 West Patapsco Avenue Baltimore, MD 21230 RE: Short Term Confidentiality Request For MODUS-CGM and FCC ID: 2BVYL-MODUS-CGM To Whom It May Concern: In accordance with sections 0.457 and 0.459 of CFR 47, GlucoRx Technologies Ltd, requests short-term Confidentiality of following sections of this application for a period of 180 days after grant is issued. Test setup photos External photos Internal photos User’s manual These documents contain detailed system and equipment descriptions and related information about the product, which GlucoRx Technologies Ltd considers to be proprietary, confidential, and a custom design and otherwise, would not release to the general public. Since this design is a basis from which future technological products will evolve, GlucoRx Technologies Ltd feels that this information would be of benefit to its competitors, and that the disclosure of the information in these documents would give our competitors an unfair advantage in the market. Please contact me if there is any information you may need. Sincerely, Leanne McIntosh Group QA/RA Lead and Business Unit Manager +44 (0)7769 325 150 [email protected]
GlucoRx Technologies Ltd Strathpeffer Road, Dingwall, IV15 9QF, UK GlucoRx Europe Unit 1, Airport Hub, Furry Park Estate, Swords Road, Dublin 9, D09 W5YV www.glucorx.co.uk 2797 Alerts IPX7 No Calibration Reading every minute 15 DAYS No fingersticks Search and download the CGMx™ 1 App from the App Store or Google Play. GlucoRx CGMx™ 1 device required. Scan the QR Code for the GlucoRx CGMx™ 1 Instructions. Contents: 1 Continuous Glucose Monitoring Sensor 1 Quick Guide 15 Day Real-Time Continuous Glucose Monitoring Sensor Medical Device 15 DAYS OF CONFIDENCE © 2025 GlucoRx Technologies Ltd. All rights reserved. CGMx™ 1 is a registered trademark of GlucoRx Technologies Ltd. AW-0374 V1 07/2026 Initial MODUS Carton design - To be updated with additional info at a later date FCC ID: 2BVYL-MODUS-CGM Note: FCC ID is also located on page 67 of the user manual.
SAR Exclusion Calculation Project: G5426v2 Report number: G7454TR1 Date: 4 th June 2026 Customer Details Company name Gluco RX Technologies Ltd Address Dingwall Business Centre Strathpeffer Road Dingwall IV15 9QF United Kingdom Contact Lauren Valente Email [email protected] Equipment Under Test: MODUS-H Operating frequency bands: BLE 2400MHz to 2483.5MHz Report No: G77454TR1 Date: 4 th June 2026 Page 2 of 3 SAR Exclusion Calculation - FCC The calculation is according to the FCC Knowledge Database (KDB) document KDB 447498 D01 General RF Exposure Guidance V06. Section 4.3.1 General SAR Test Exclusion Bands was applied. Section 4.3.1 a) for 100MHz to 6GHz and test separation distances ≤50mm was specifically applied. According to this section the SAR test exclusion thresholds are determined by the following: Threshold = [ 푃( 푚푊) 푑( 푚푚) ] . √ 푓(퐺퐻푧) Where P is the maximum power of the channel including tune-up tolerance in mW d is the minimum test separation distance in mm f is the RF channel frequency in GHz The minimum test separation distance was considered to be 5 mm for the Bluetooth Low Energy, in consultation with Section 4.2.2 Body-worn accessory exposure conditions. This evaluation of the threshold is compared to the threshold limit. a. Limit for 1-g head or body worn devices: threshold limit ≤3 b. Limit for 10-g extremity worn devices: threshold limit ≤7.5 Since the device is body worn the a limit applies. Bluetooth Low Energy For this apparatus, the calculated threshold was evaluated using a distance of 5 mm, and the results are tabulated in the table below. The power level values are taken from the G7453TR1 report Frequency (GHz) Power (mW) EIRP (mW) Dist (mm) Calculated Threshold Threshold Limit SAR Test required 2.402 0.8054 0.9683 5 0.30 3.0 No 2.440 0.8279 0.9954 5 0.31 3.0 No 2.480 0.7499 0.9016 5 0.28 3.0 No Table 1 Calculation of threshold BLE Notes 1: EIRP calculated using an antenna gain of 0.8 dBi. Note 2: A minimum distance of 5mm is applied per FCC guidelines for distances < 5mm. Note 3: Calculated threshold results rounded to two decimal places Report No: G77454TR1 Date: 4 th June 2026 Page 3 of 3 Conclusion For the required stated distance, the apparatus met the exclusion requirements for SAR testing (FCC). Prepared by: Dmytro Kutsko, Senior Test Engineer -----------------------------------------------END OF REPORT----------------------------------------------------
Test Report for the FCC and ISED Testing of an MODUS-CGM to FCC Rule 47CFR 15.247 and ISED RSS-247 Issue 4 for Gluco RX Technologies Ltd Report number: G5426v3 Project number: G7453TR1 Author: ................................................................................................. Dmytro Kutsko Senior Test Engineer Radio/EMC Testing Team Leader (Grangemouth) Checked: ............................................................................................. Dr. James Beevers Radio/EMC Testing Team Leader (Castleford) Approved: ............................................................................................. Didier Bozec, Laboratories Director Date: 4 th June 2026 Issue: 1 This report shall not be reproduced, except in full, without the prior written approval of Eurofins E&E UK. This test report relates only to the unit(s) tested. Commercial in Confidence Date: 4 th June 2026 Report Number: G7453TR1 Page 2 of 61 FCC 15_247_DTS Report Template Issue 10 CONTENTS Test Report Change History .......................................................................................................................6 Section 1 Test Location ......................................................................................................................7 1.1 UKAS Accreditation .........................................................................................................................7 1.2 Laboratory Designation ...................................................................................................................7 Section 2 Customer Information .......................................................................................................8 Section 3 Equipment Details ..............................................................................................................9 3.1 Equipment Under Test (EUT) ..........................................................................................................9 3.2 EUT Photographs ......................................................................................................................... 10 3.3 Configuration of EUT .................................................................................................................... 10 3.4 EUT Monitoring/Auxiliary Equipment ........................................................................................... 11 3.5 Monitoring Software ..................................................................................................................... 11 Section 4 Test Specifications ......................................................................................................... 12 4.1 Knowledge Database References ................................................................................................ 14 4.1.1 Radiated Emissions (30MHz to 1000MHz) ............................................................................ 14 4.1.2 Radiated Emissions (1GHz to 40GHz) .................................................................................. 14 4.2 Compliance Statement ................................................................................................................. 14 4.3 Decision Rule Statement .............................................................................................................. 15 Section 5 Spurious Emission Results – Radiated and Conducted ............................................. 16 5.1 Test Specification ......................................................................................................................... 16 5.2 Procedure and Test Software Version ......................................................................................... 16 5.3 Radiated Emissions (30MHz to 1GHz) ........................................................................................ 17 5.3.1 Limits at 3m ............................................................................................................................ 17 5.3.2 Date of Test ............................................................................................................................ 17 5.3.3 Test Area ................................................................................................................................ 17 5.3.4 Tested by ................................................................................................................................ 17 5.3.5 Test Setup .............................................................................................................................. 18 5.3.6 Electric field emissions, 30MHz to 1GHz ............................................................................... 19 5.3.7 Quasi Peak correction factors ................................................................................................ 22 5.3.8 Sample Data........................................................................................................................... 22 5.4 Radiated Emissions (1GHz to 18GHz)......................................................................................... 23 5.4.1 Limits ...................................................................................................................................... 23 5.4.2 Receiver Settings ................................................................................................................... 23 5.4.3 Date of Test ............................................................................................................................ 23 5.4.4 Test Area ................................................................................................................................ 23 5.4.5 Tested by ................................................................................................................................ 23 5.4.6 Test Setup ............................................................................…
Text truncated - open the document above for the full version.
Registered SC575629, GLUCORX TECHNOLOGIES LTD, DINGWALL BUSINESS CENTRE, STRATHPEFFER ROAD, DINGWALL, IV15 9QF, SCOTLAND GlucoRx Technologies Ltd Dingwall Business Centre, Strathpeffer Road Dingwall, Scotland IV15 9QF Phone: +44 (0)1349204002 Declaration concerning Antenna Specification It is hereby declared that the product Model No.: MODUS-CGM FCC ID: 2BVYL-MODUS-CGM fulfils the requirement in FCC test relating to the antenna type. The device specified above confirms to the FCC recommendations for an internal antenna type described below: Model No. of antenna: 2450AT07A0100 (Johanson Technology) Type of antenna: 2.4 GHz ceramic chip antenna Peak Gain of the antenna: 0.8 dBi Frequency range: 2400-2500 MHz The antenna is permanently mounted to the transmitter PCB, and no other antenna should be used. The red outline indicated the antenna area; the orange outline indicated the associated tuning / matching components. Company name: GlucoRx Technologies Ltd Address: Dingwall Business Centre, Strathpeffer Road City: Dingwall ZIP / Postal code: IV15 9QF Country: United Kingdom Place: Dingwall Office Date: 2026/07/27 Signature:
Ultra-Miniature 2.4GHz Chip antenna 0.37mm max Thickness P/N 2450AT07A0100 Detail Specification:Page 1 of 5 Evaluation Board No. L±± 1 W±± 2 T 3 a 4 b c Johanson Technology, Inc. reserves the right to make design changes without notice. All sales are subject to Johanson Technology, Inc. terms and conditions. https://www.johansontechnology.com 4001 Calle Tecate • Camarillo, CA 93012 • TEL 805.389.1166 FAX 805.389.1821 2020 Johanson Technology, Inc. All Rights Reserved P/N Suffix Packing Style "High Frequency Ceramic Solutions" 2W max. (CW) 50 Ω 10/12/2020 Average Gain (dBi typ.) Peak Gain (dBi typ.) Part Number2450AT07A0100 2400 - 2500 1.0 (XZ-Total) Frequency (MHz) 6.5 dB min. 0.014 max -1.5 (XZ-Total) Function Feed Return Loss -40 to +125°C Terminal Configuration Recommended Storage Conditions of unused product on T&R +5 ~ +35 °C Humidity 45~75%RH 18 months max. 0.004 0.004 mmIn 0.50 Mechanical Dimensions Input Power GND GND +0.1/-0.05 Feed+0.1/-0.05 +0.1/-0.05 0.10 0.37 max 0.25 0.08 0.15 Ver 3.1 Operating Temperature General Specifications 1.000.100.039 Impedance 0.003 0.010 +.004/-.006 +.004/-.006 +.004/-.0060.006 0.020 Part Number Explanation 2450AT07A0100-EB1SMA Bulk (loose) Suffix = Se.g. 2450AT07A0100S TraySuffix = Te.g. 2450AT07A0100T 1 2 34 L W a b c T Ultra-Miniature 2.4GHz Chip antenna 0.37mm max Thickness P/N 2450AT07A0100 Detail Specification:Page 2 of 5 Johanson Technology, Inc. reserves the right to make design changes without notice. All sales are subject to Johanson Technology, Inc. terms and conditions. https://www.johansontechnology.com 4001 Calle Tecate • Camarillo, CA 93012 • TEL 805.389.1166 FAX 805.389.1821 2020 Johanson Technology, Inc. All Rights Reserved Ver 3.1 "High Frequency Ceramic Solutions" 10/12/2020 Mounting Considerations Test Board-Top View (Units in mm) Land Solder 0.9 4.0 0.4 0.3 0.3 0.7 1.2 0.25 1.5 5.0 3 .0 Feed line Matching circuit No Ground (Yellow area) GND Plane * 0.1 1 23 4 It is recommended that the designer leave available slots for a "pi" (shunt-series-shunt) network. The antenna matching network values above are used when antenna is mounted on Johanson's evaluation board. Matching components will vary depending on PCB design. Let us help you tune our antenna on your new layout, contact our RF Applications Eng. Team at: https://www.johansontechnology.com/ask-a-question *Line width should be designed to maintain 50Ω impedance matching characteristics. Units: mm Ultra-Miniature 2.4GHz Chip antenna 0.37mm max Thickness P/N 2450AT07A0100 Detail Specification:Page 3 of 5 Johanson Technology, Inc. reserves the right to make design changes without notice. All sales are subject to Johanson Technology, Inc. terms and conditions. https://www.johansontechnology.com 4001 Calle Tecate • Camarillo, CA 93012 • TEL 805.389.1166 FAX 805.389.1821 2020 Johanson Technology, Inc. All Rights Reserved "High Frequency Ceramic Solutions" 10/12/2020 Ver 3.1 Layout Guidelines and EVB Specifications No ground (Yellow area) Ground Pi matching network recommended Feed 3 5 20 40 Test Board-Top View (Units in mm) 40 20 Ground clearance (Yellow area) 3 5 Test Board-Bottom View (Units in mm) To order a pre-tuned evaluation board with SMA connector click here: https://www.johansontechnology.com/request-a-sample Reference p/n: 2450AT07A0100-EB1SMA To request the layout files go to: https://www.johansontechnology.com/ask-a-question Ground Ultra-Miniature 2.4GHz Chip antenna 0.37mm max Thickness P/N 2450AT07A0100 Detail Specification:Page 4 of 5 Johanson Technology, Inc. reserves the right to make design changes without notice. All sales are subject to Johanson Technology, Inc. terms and conditions. https://www.johansontechnology.com 4001 Calle Tecate • Camarillo, CA 93012 • TEL 805.389.1166 FAX 805.389.1821 2020 Johanson Technology, Inc. All Rights Reserved "High Frequency Ceramic Solutions" 10/12/2020 Ver 3.1 Typical Radiation Patterns @25C YZ‐plane XZ‐plane XY‐plane θ +Z (θ= 0°) +Y (θ= 90°) -Y -Z θ +Z (θ= 0°) +X (θ= 90°) -X -Z φ +X (φ= 0°) -X +Y (φ= 90°) -Y +Z +Z +Y +Y +X +X -Y -X -X -Z -Z -Y Ultra-Miniature 2.4GHz Chip antenna 0.37mm max Thickness P/N 2450AT07A0100 Detail Specification:Page 5 of 5 Johanson Technology, Inc. reserves the right to make design changes without notice. All sales are subject to Johanson Technology, Inc. terms and conditions. https://www.johansontechnology.com 4001 Calle Tecate • Camarillo, CA 93012 • TEL 805.389.1166 FAX 805.389.1821 2020 Johanson Technology, Inc. All Rights Reserved https://www.johansontechnology.com/tape-reel-packaging https://www.johansontechnology.com/ipcsoldering-profile https://www.johansontechnology.com/rohs-compliance Ver 3.1 "High Frequency Ceramic Solutions" 10/12/2020 Typical Electrical Performance (T=25°C) Packaging information Soldering Information Antenna layout and tuning techniques (How to obtain the new antenna matching values) https://www.johansontechnology.com/tuning Antenna layout review, tuning, and characterization services https://www.johansontechnology.com/ipc-antenna-services RoHS Compliance https://www.johansontechnology.com/msl-rating MSL Info Return Loss / Without Matching Circuits 2.12.22.32.42.52.62.72.82.92.03.0 -20 -15 -10 -5 -25 0 freq, GHz dB(S(1,1)) m1 m2 m3 m1 freq= dB(S(1,1))=-5.100 2.370GHz m2 freq= dB(S(1,1))=-20.709 2.450GHz m3 freq= dB(S(1,1))=-5.437 2.530GHz To download antenna s-parameters, please go to: https://www.johansontechnology.com/antennas
3722 Illinois Avenue · Saint Charles, Illinois, 60174 · United States
| # | Rule Parts | Frequency Range | Power Output |
|---|---|---|---|
| 1 | 15C | 2.40 GHz - 2.48 GHz | 1.00 mW |