FCCID.CO- FCC ID Database
HomeCompaniesEquipment ClassesSearch

© 2026 FCC ID Database. All rights reserved.

AboutContactData sourced from FCC public records
  1. Home/
  2. Shenzhen Finicare Co., Ltd./
  3. 2A3QXFCBP133

2A3QXFCBP133Upper Arm Electronic Blood Pressure Monitor

Shenzhen Finicare Co., Ltd.

Application Details

Equipment Class
DTS - Digital Transmission System
Date of Grant
Jun 09, 2026
Application Purpose
Original Equipment
Equipment Note
Upper Arm Electronic Blood Pressure Monitor
Frequency Range
2402.00000000 - 2480.00000000
Company
Shenzhen Finicare Co., Ltd.
Country
China

Documents & Files

Select a file to view

Attestation Statements

↗
↗

Cover Letter(s)

↗
↗
↗

ID Label/Location Info

↗

RF Exposure Info

↗

Test Report

↗
↗

Document Text

Text extracted from the exhibit documents filed with the FCC. Open a document above to read the original.

Attestation Statements

201 and 301 of Building A22, Building A22 and A23 Tantou Community, Songgang Street, Bao'an District Shenzhen China

Attestation Statements

201 and 301 of Building A22, Building A22 and A23 Tantou Community, Songgang Street, Bao'an District Shenzhen China

Cover Letter(s)

201 and 301 of Building A22, Building A22 and A23 Tantou Community, Songgang Street, Bao'an District Shenzhen China

ID Label/Location Info

Label statement This device complies with part 15 of the FCC Rules. Operationis subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. 2026-02 IP22 Upper Arm Electronic Blood Pressure Monitor Model : FC-BP133 Rating: 6VDC, 4 X AAA batteries Type-C USB: 5VDC1A 201 and 301 of Building A22, Building A22 and A23, No.4 Industrial Park, Tantou Community, Songgang Street, Bao'an District, Shenzhen 518105 China Shenzhen Finicare Co., Ltd SN MD CHN FCC ID: 2A3QXFCBP133

RF Exposure Info

SGS-CSTC Standards Technical Services Co., Ltd. Shenzhen Branch SZEMC-TRF-01 Rev. A/2 Report No.: SZCR260300125103 Page: 1 of 10 Keny Xu EMC Laboratory Manager RF EXPOSURE EVALUATION REPORT Application No.: SZCR2603001251ME Applicant: Shenzhen Finicare Co., Ltd. Address of Applicant: 201 and 301 of Building A22, Building A22 and Building A23 Tantou Community, Songgang Street, Bao’an District, Shenzhen 518105 China Manufacturer: Shenzhen Finicare Co., Ltd. Address of Manufacturer: 201 and 301 of Building A22, Building A22 and Building A23, No.4 Industral Park, Tantou Community, Songgang Street, Bao’an District, Shenzhen 518105 China Factory: Shenzhen Finicare Co., Ltd. Address of Factory: 201 and 301 of Building A22, Building A22 and Building A23, No.4 Industral Park, Tantou Community, Songgang Street, Bao’an District, Shenzhen 518105 China Equipment Under Test (EUT): EUT Name: Upper Arm Electronic Blood Pressure Monitor Model No.: FC-BP133 FCC ID: 2A3QXFCBP133 Standard(s) : FCC Rules 47 CFR §2.1091 KDB 447498 D04 interim General RF Exposure Guidance v01 Date of Receipt: 2026-03-30 Date of Evaluation: 2026-04-04 to 2026-05-15 Date of Issue: 2026-05-17 Evaluation Result: Pass* * In the configuration evaluated, the EUT complied with the standards specified above. SGS-CSTC Standards Technical Services Co., Ltd. Shenzhen Branch SZEMC-TRF-01 Rev. A/2 Report No.: SZCR260300125103 Page: 2 of 10 Revision Record Version Chapter Date Modifier Remark 01 2026-05-17 Original Authorized for issue by: Edison Li/Project Engineer Eric Fu/Reviewer SGS-CSTC Standards Technical Services Co., Ltd. Shenzhen Branch SZEMC-TRF-01 Rev. A/2 Report No.: SZCR260300125103 Page: 3 of 10 2 Contents Page 1 Cover Page ................................................................................................................................................ 1 2 Contents .................................................................................................................................................... 3 3 General Information .................................................................................................................................. 4 3.1 General Description of E.U.T. ............................................................................................................ 4 3.2 Details of E.U.T. ................................................................................................................................. 4 3.3 Separation Distance ........................................................................................................................... 4 3.4 Test Location ...................................................................................................................................... 5 3.5 Test Facility ........................................................................................................................................ 5 3.6 Deviation from Standards ................................................................................................................... 5 3.7 Abnormalities from Standard Conditions ........................................................................................... 5 4 FCC Radiofrequency radiation exposure limits .................................................................................... 6 4.1 Blanket 1 mW Blanket Exemption ...................................................................................................... 6 4.2 MPE-based Exemption ...................................................................................................................... 6 4.3 SAR-based Exemption ....................................................................................................................... 7 4.4 RF Exposure Test Exemptions for Simultaneous Transmission ........................................................ 9 5 Measurement and Calculation ............................................................................................................... 10 5.1 Maximum transmit power ................................................................................................................. 10 5.2 RF Exposure Calculation ................................................................................................................. 10 6 EUT Constructional Details (EUT Photos) ........................................................................................... 10 SGS-CSTC Standards Technical Services Co., Ltd. Shenzhen Branch SZEMC-TRF-01 Rev. A/2 Report No.: SZCR260300125103 Page: 4 of 10 3 General Information 3.1 General Description of E.U.T. Product Type: Portable device Mobile device Fixed device 3.2 Details of E.U.T. Power supply: Medical Adapter Model: LXCP12X-050200BG Output: DC 5.0V, 2.0A DC 6V(4 AAA alkaline batteries) Cable(s): USB Type-C cable:140cm unshielded Bluetooth Version: V4.2 Operation Frequency: 2402MHz to 2480MHz Modulation Type: GFSK Channel Spacing: 2MHz Number of Channels: 40 Data rate: 1Mbps Antenna Type: PCB Antenna Antenna Gain: 0.2 dBi Remark:The information in this section is provided by the applicant or manufacturer, SGS is not liable to the accuracy, suitability, reliability or/and integrity of the information. 3.3 Separation Distance Minimum test separation distance: 20cm Remark: This minimum test separation distance is determined by the smallest distance from the antenna and radiating structures or outer surface of the device, according to the host form factor, exposure conditions and platform requirements, to any part of the body or extremity of a user or bystander. SGS-CSTC Standards Technical Services Co., Ltd. Shenzhen Branch SZEMC-TRF-01 Rev. A/2 Report No.: SZCR260300125103 Page: 5 of 10 3.4 Test Location All tests were performed at: SGS-CSTC Standards Technical Services Co., Ltd., Shenzhen Branch No. 1 Workshop, M-10, Middle Section, Science & Technology Park, Nanshan District, Shenzhen, Guangdong, China. 518057. Te…

Text truncated - open the document above for the full version.

Test Report

SGS-CSTC Standards Technical Services Co., Ltd. Shenzhen Branch SZEMC-TRF-01 Rev. A/2 Report No.: SZCR260300125102 Page: 1 of 59 Keny Xu EMC Laboratory Manager TEST REPORT Application No.: SZCR2603001251ME Applicant: Shenzhen Finicare Co., Ltd. Address of Applicant: 201 and 301 of Building A22, Building A22 and Building A23 Tantou Community, Songgang Street, Bao’an District, Shenzhen 518105 China Manufacturer: Shenzhen Finicare Co., Ltd. Address of Manufacturer: 201 and 301 of Building A22, Building A22 and Building A23, No.4 Industral Park, Tantou Community, Songgang Street, Bao’an District, Shenzhen 518105 China Factory: Shenzhen Finicare Co., Ltd. Address of Factory: 201 and 301 of Building A22, Building A22 and Building A23, No.4 Industral Park, Tantou Community, Songgang Street, Bao’an District, Shenzhen 518105 China Equipment Under Test (EUT): EUT Name: Upper Arm Electronic Blood Pressure Monitor Model No.: FC-BP133 FCC ID: 2A3QXFCBP133 Standard(s) : 47 CFR Part 15, Subpart C 15.247 Date of Receipt: 2026-03-30 Date of Test: 2026-04-04 to 2026-05-15 Date of Issue: 2026-05-17 Test Result: Pass* * In the configuration tested, the EUT complied with the standards specified above. SGS-CSTC Standards Technical Services Co., Ltd. Shenzhen Branch SZEMC-TRF-01 Rev. A/2 Report No.: SZCR260300125102 Page: 2 of 59 Revision Record Version Chapter Date Modifier Remark 01 2026-05-17 Original Authorized for issue by: Edison Li/Project Engineer Eric Fu/Reviewer SGS-CSTC Standards Technical Services Co., Ltd. Shenzhen Branch SZEMC-TRF-01 Rev. A/2 Report No.: SZCR260300125102 Page: 3 of 59 2 Test Summary Radio Spectrum Technical Requirement Item Standard Method Requirement Result Antenna Requirement 47 CFR Part 15, Subpart C 15.247 N/A 47 CFR Part 15, Subpart C 15.203 & 15.247(b)(4) Pass Radio Spectrum Matter Part Item Standard Method Requirement Result Conducted Emissions at AC Power Line (150kHz-30MHz) 47 CFR Part 15, Subpart C 15.247 ANSI C63.10 (2020) Section 6.2 47 CFR Part 15, Subpart C 15.207 Pass Radiated Emissions which fall in the restricted bands ANSI C63.10 (2020) Section 6.10.5 & 11.12 47 CFR Part 15, Subpart C 15.205 & 15.209 Pass Radiated Spurious Emissions Below 1GHz ANSI C63.10 (2020) Section 6.4,6.5 47 CFR Part 15, Subpart C 15.205 & 15.209 Pass Radiated Spurious Emissions Above 1GHz ANSI C63.10 (2020) Section 6.6 & 11.12 47 CFR Part 15, Subpart C 15.205 & 15.209 Pass Maximum Conducted Output Power ANSI C63.10 (2020) Section 11.9.1 47 CFR Part 15, Subpart C 15.247(b)(3) Pass Minimum 6dB Bandwidth ANSI C63.10 (2020) Section 11.8.1 47 CFR Part 15, Subpart C 15.247a(2) Pass Power Spectral Density ANSI C63.10 (2020) Section 11.10.3 47 CFR Part 15, Subpart C 15.247(e) Pass Conducted Band Edges Measurement ANSI C63.10 (2020) Section 6.10.4 47 CFR Part 15, Subpart C 15.247(d) Pass Conducted Spurious Emissions ANSI C63.10 (2020) Section 11.11 47 CFR Part 15, Subpart C 15.247(d) Pass SGS-CSTC Standards Technical Services Co., Ltd. Shenzhen Branch SZEMC-TRF-01 Rev. A/2 Report No.: SZCR260300125102 Page: 4 of 59 3 Contents Page 1 Cover Page ................................................................................................................................................ 1 2 Test Summary ........................................................................................................................................... 3 3 Contents .................................................................................................................................................... 4 4 General Information .................................................................................................................................. 6 4.1 Details of E.U.T. ................................................................................................................................. 6 4.2 Description of Support Units .............................................................................................................. 6 4.3 Measurement Uncertainty .................................................................................................................. 7 4.4 Test Location ...................................................................................................................................... 8 4.5 Test Facility ........................................................................................................................................ 8 4.6 Deviation from Standards ................................................................................................................... 8 4.7 Abnormalities from Standard Conditions ........................................................................................... 8 5 Equipment List .......................................................................................................................................... 9 6 Radio Spectrum Technical Requirement ............................................................................................. 11 6.1 Antenna Requirement ...................................................................................................................... 11 6.1.1 Test Requirement: ....................................................................................................................... 11 6.1.2 Conclusion ................................................................................................................................... 11 7 Radio Spectrum Matter Test Results .................................................................................................... 12 7.1 Conducted Emissions at AC Power Line (150kHz-30MHz) ............................................................. 12 7.1.1 E.U.T. Operation .......................................................................................................................... 12 7.1.2 Test Mode Description ......................................................................…

Text truncated - open the document above for the full version.

Test Report

YLB1Antennatestreport ProductName :YLB1 AntennaType :PCB antenna Writer : Fangbingzhong Maximum Gain:0.2dBi Efficiency:36.98% YLB1 Antennatest report Version1.0 .0Date:202 2-01 -11NO:0000000001 Antenna Test Report GlobalIntelligentPlatform Hangzhou Tuya Information Technology Co., Ltd Block A, Huace Centre, Wuchanggang Road, Xihu District, Hangzhou, Zhejiang Province YLB1Antennatestreport Content 1. GeneralDescription.................................................................................3 2. TechnicalInformation.............................................................................3 3. TestStandards..........................................................................................3 4. TestItem..................................................................................................4 5. TestPhoto................................................................................................4 AppendixA S11/VSWR..............................................................................5 AppendixB GainandEfficiency................................................................6 AppendixC RadiationPattern(1Dand3D)................................................7 AppendixD TestSetupPhoto.....................................................................9 YLB1Antennatestreport 1.GeneralDescription EUTNameYLB1Antenna WirelesstypeBluetooth AntennaTypePCBantenna DimensionsAppendixA 2.TechnicalInformation FrequencyRange2402MHz~2480MHz TestFrequency2400,2410......2500MHz 3.TestStandards IdentityDocumentTitle IEEE149-1979IEEEStandardTestProcedures forAntennas YLB1Antennatestreport 4.TestItem ReportSectionDescription AppendixBS11/VSWR AppendixCGainandEfficiency AppendixD RadiationPattern(1Dand3D) 5.TestPhoto ReportSectionDescription AppendixETestSetupPhoto YLB1Antennatestreport AppendixAAntennaDimension AppendixBS11/VSWR YLB1Antennatestreport Frequency(MHz)240024502500 ReturnLoss-7.37-11.76-7 VSWR2.91.62.9 AppendixCGainandEfficiency Frequency(MHz)Efficiency(%) Gain(dBi) 240035.08%-0.23 241036.56%-0.01 242035.56%-0.1 243036.56%0.02 244036.98%0.2 245035.56%-0.05 246035.16%0.04 247035.16%0.17 248035.81%0.13 YLB1Antennatestreport 249034.75%0.15 250032.28%-0.31 AppendixDRadiationPattern(1Dand3D) YLB1Antennatestreport YLB1Antennatestreport AppendixETestSetupPhoto

Contact Information

Applicant

Chao Li
[email protected]Fax: 0755-23574425

Test Firm

SGS-CSTC Standards Technical Services Co., Ltd. ShKeny Xu
[email protected]

Technical Specifications

#Rule PartsFrequency RangePower Output
115C2.40 GHz - 2.48 GHz3.90 mW
Confidentiality
Long Term
Grant Notes
Power Output listed is conducted. The antenna(s) used for this transmitter must be installed to provide a separation distance of at least 20 cm from all persons. Co-location of this module with other transmitters that operate simultaneously are required to be evaluated using the FCC multi-transmitter procedures. End-Users must be provided with transmitter operation conditions for satisfying RF exposure compliance.