Contents Safety Notice Owner's Manual Wrist-type Fully Automatic Blood Pressure Monitor Model DBP-8276H 1 Wrist Type 2 Document No.:JDBP-7604-054 Contact Information The lay operator or lay responsible or ganization should contact the manufacturer or the representative of manufacturer. -for assistance,if needed,in setting up,using or maintaining the product,or -to report unexpected operation or events. Manufactured by JOYTECH Healthcare Co.,Ltd. Email: [email protected] Telephone: +86-571-81957767 Fax: +86-571-81957750 No.365, Wuzhou Road,Yuhang Economic Development Zone,Hangzhou City,311100 Zhejiang,China Version: Z Date of Issue: 2020.06 Thank you for purchasing the DBP-8276H Blood Pressure Monitor. The unit has been constructed using reliable circuitry and durable materials. Used properly, this unit will provide yeas of satisfactory use. The device is intended for used by adults and adolescents age 12 through 21 years of age to measure the systolic and diastolic blood pressure and pulse rate. All functions can be used safely and values can be read out in one LCD DISPLAY. Measurement position is on adult upper wrist only. The PATIENT is an intended OPERATOR. Blood pressure measurement determined with this device are equivalent to those obtained by a trained observer using the cuff/ stethoscope auscultation method, within the limits prescribed by the Recognized Consensus Standard(IEC 81060-2-30) for electronic sphygmomanometers. Precautions to Ensure Safe, Reliable Operation 1. Do not drop the unit. Protect it from sudden jars or shocks. 2. Do not insert foreign objects into any openings. 3. Do not attempt to disassemble the unit. 4. Do not crush the pressure cuff. 5. If the unit has been stored at temperatures below 0 oC, leave it in a warm place for about 15 minutes before using it. Otherwise, the cuff may not inflate properly. 6. If the unit has been stored at temperatures above 40 oC,leave it in a cool place for about 15 minutes before using it. Otherwise, the cuff may not inflate properly. 7. Do not store the unit in direct sunlight, high humidity or dust. 8.To avoid any possibility of accidental strangulation, keep this unit away from children and do not drape tubing around your neck. 9.Ensure that children do not use the instrument unsupervised; some parts are small enough to be swallowed. 10.Some may get a skin irritation from the cuff taking frequent readings over the course of the day, but this irritation typically goes away on its own after the monitor is removed. Safety NoticeSafety Notice 34 Important Instructions Before Use 1. Do not confuse self-monitoring with self-diagnosis. Blood pressure measurements should only be interpreted by a health professional who is familiar with your medical history. 2. Contact your physician if test results regularly indicate abnormal readings. 3. If you are taking medication, consult with your physician to determine the most appropriate time to measure your blood pressure. NEVER change a prescribed medication without first consulting with your physician. 4. Individuals with serious circulation problems may experience discomfort. Consult your physician prior to use. 5. For persons with irregular or unstable circulation resulting from diabetes, liver disease, arteriosclerosis or other medical conditions, there may be variations in blood pressure values measured at the wrist versus at the upper arm. Monitoring the trends in your blood pressure taken at either the arm or the wrist is nevertheless useful and important. 6. People suffering from vascular constriction, liver disorders or diabetes, people with cardiac pacemakers or a weak pulse, and women who are pregnant should consult their physician before measuring their blood pressure themselves. Different values may be obtained due to their condition. 7. People suffering from arrhythmias such as atrial or ventricular premature beats or atrial fibrillation only use this blood pressure monitor in consultation with your doctor. In certain cases oscillometric measurement method can produce incorrect readings. 8.Too frequent measurements can cause injury to the patient due to blood flow interference. 9.The cuff should not be applied over a wound as this can cause further injury. 10.DO NOT attach the cuff to a limb being used for IV infusions or any other intravascular access, therapy or an arterio-venous (A- V) shunt. The cuff inflation can temporarily block blood flow, potentially causing harm to the patient. 11.The cuff should not be placed on the arm on the side of a mastectomy. In the case of a double mastectomy use the side of the least dominant arm. 12.Pressurization of the cuff can temporarily cause loss of function of simultaneously used monitoring equipment on the same limb. 13.A compressed or kinked connection hose may cause continuous cuff pressure resulting in blood flow interference and potentially harmful injury to the patient. 14.Check that operation of the unit does not result in prolonged impairment of the circulation of the patient.15. Product is designed for its intended use only. Do not misuse in anyway. 15. Product is designed for its intended use only. Do not misuse in any way. 16. Product is not intended for infants or individuals who cannot express their intentions. 17. Prolonged over-inflation of the bladder may cause ecchymoma of your arm. 18. Do not disassemble the unit or wrist cuff. Do not attempt to repair. 19. Use only the approved wrist cuff for this unit. Use of other wrist cuffs may result in incorrect measurement results. 20. The system might produce incorrect readings if stored or used out- side the manufacturer's specified temperature and humidity ranges. Make sure to store the blood pressure monitor, children, pets and pests are outside of accessible range. Safety Notice ............................................................... 02 Unit Illustration ......................................................... 05 Important Testing Guidelines ....................…
Text truncated - open the document above for the full version.
ReportNo.:EED32N00018101Page 33 of 44 PHOTOGRAPHSOFEUTConstructionalDetails TestmodelNo.:DBP-8276H ViewofProduct-1 ViewofProduct-2 ReportNo.:EED32N00018101Page 34 of 44 ViewofProduct-3 ViewofProduct-4 ReportNo.:EED32N00018101Page 35 of 44 ViewofProduct-5 ViewofProduct-6 ReportNo.:EED32N00018101Page 36 of 44 ViewofProduct-7 ViewofProduct-8 ReportNo.:EED32N00018101Page 37 of 44 ViewofProduct-7 ViewofProduct-8 ReportNo.:EED32N00018101Page 38 of 44 ViewofProduct-9 ViewofProduct-10
Size: 39*13mm Wrist-Type Fully Automatic Digital Blood Pressure Monitor Manufacturers by JOYTECH Healthcare Co., Ltd. No. 365, Wuzhou Road, Yuhang Economic Development Zone, Hangzhou City, 311100 Zhejiang, China Power: 5.0V DC 1A or Li-ion 3.7V 180mAh Series number: Manufacturing Date: Model: DBP-8276H IP22 FCC ID:2AQVU0022 Made in china Labellocation:
ReportNo.:EED32N00018101Page 38 of 44 ViewofProduct-9 ViewofProduct-10 ReportNo.:EED32N00018101Page 39 of 44 ViewofProduct-11 ViewofProduct-12 ReportNo.:EED32N00018101Page 40 of 44 ViewofProduct-13 ViewofProduct-14 ReportNo.:EED32N00018101Page 41 of 44 ViewofProduct-15 ViewofProduct-16 ReportNo.:EED32N00018101Page 42 of 44 ViewofProduct-17 ViewofProduct-18 ReportNo.:EED32N00018101Page 43 of 44 ViewofProduct-19 ViewofProduct-20 ReportNo.:EED32N00018101Page 44 of 44 ViewofProduct-21 ViewofProduct-22 Thetestreportiseffectiveonlywithbothsignatureandspecializedstamp,Theresult(s)showninthis reportreferonlytothesample(s)tested.WithoutwrittenapprovalofCTI,thisreportcan’tbereproduced exceptinfull. ***EndofReport***
ReportNo.:EED32N00018102Page1of8 1COVERPAGE RFExposureEvaluationReport Product :Wrist-TypeFullyAutomaticDigital BloodPressureMonitor Trademark : N/A Model/Typereference : DBP-8276H SerialNumber : N/A ReportNumber : EED32N00018102 FCCID : 2AQVU0022 DateofIssue : May27,2021 TestStandards : 47CFRPart1.1307 47CFRPart2.1093 KDB447498D01GeneralRF ExposureGuidancev06 Testresult:PASS Preparedfor: JOYTECHHealthcareCo.,Ltd. No.365,WuzhouRoad,YuhangEconomicDevelopmentZone, HangzhouCity,311100ZhejiangP.R.China Preparedby : CentreTestingInternationalGroupCo.,Ltd. HongweiIndustrialZone,Bao’an70District, Shenzhen,Guangdong,China TEL:+86-755-33683668 FAX:+86-755-33683385 Compiledby:Reviewedby: VitoheAaronMa Approvedby:Date:May27,2021 DavidWang CheckNo.:4538041742 ReportNo.:EED32N00018102Page2of8 2Version VersionNo.DateDescription 00May27,2021Original ReportNo.:EED32N00018102Page3of8 3Contents Page 1COVERPAGE......................................................................................................................................................................1 2VERSION..............................................................................................................................................................................2 3CONTENTS..........................................................................................................................................................................3 4GENERALINFORMATION...............................................................................................................................................4 4.1C LIENT I NFORMATION ....................................................................................................................................................4 4.2G ENERAL D ESCRIPTIONOF EUT..................................................................................................................................4 4.3P RODUCT S PECIFICATIONSUBJECTIVETOTHISSTANDARD ........................................................................................4 4.4T EST L OCATION .............................................................................................................................................................5 4.5D EVIATIONFROM S TANDARDS ......................................................................................................................................5 4.6A BNORMALITIESFROM S TANDARD C ONDITIONS ..........................................................................................................5 4.7O THER I NFORMATION R EQUESTEDBYTHE C USTOMER ..............................................................................................5 5SAREVALUATION.............................................................................................................................................................6 5.1RFE XPOSURE C OMPLIANCE R EQUIREMENT ...............................................................................................................6 5.1.1StandardRequirement........................................................................................................................................6 5.1.2EUTRFExposure...............................................................................................................................................7 PHOTOGRAPHSOFEUTCONSTRUCTIONALDETAILS............................................................................................8 ReportNo.:EED32N00018102Page4of8 4GeneralInformation 4.1ClientInformation Applicant: JOYTECHHealthcareCo.,Ltd. AddressofApplicant: No.365,WuzhouRoad,YuhangEconomicDevelopmentZone,HangzhouCity, 311100ZhejiangP.R.China Manufacturer: JOYTECHHealthcareCo.,Ltd. AddressofManufacturer: No.365,WuzhouRoad,YuhangEconomicDevelopmentZone,HangzhouCity, 311100ZhejiangP.R.China Factory: JOYTECHHealthcareCo.,Ltd. AddressofFactory: No.365,WuzhouRoad,YuhangEconomicDevelopmentZone,HangzhouCity, 311100ZhejiangP.R.China 4.2GeneralDescriptionofEUT ProductName:Wrist-TypeFullyAutomaticDigitalBloodPressureMonitor ModeNo.(EUT):DBP-8276H Testmodel:DBP-8276H TradeMark: N/A EUTSupportsRadios application: BT5.0Singlemodule2402MHzto2480MHz 4.3 ProductSpecificationsubjectivetothisstandard FrequencyRange: BT5.0Singlemodule2402MHzto2480MHz; ModulationType:GFSK TestPowerGrade: Default TestSoftwareofEUT: PhyPlusKit.exe AntennaType: PCBantenna AntennaGain: 0.5dBi PowerSupply: Lithiumbattery:3.7V AC/DC ADAPTOR MODEL:MPSUL0501000 INPUT:100-240V~50/60HZ0.25A OUTPUT:5V 1000mA MaxConductedPeak OutputPower: -0.04dBm TheMaxConductedPeakOutputPowerdatarefertothereport EED32N00018101 SampleReceivedDate: Mar.16,2021 SampletestedDate: Mar.16,2021toApr.23,2021 CompanyNameandAddressshownonReport,thesample(s)andsampleInformationwas/wereprovided bytheapplicantwhoshouldberesponsiblefortheauthenticitywhichCTIhasn’tverified. ModelNo.:DBP-8276H ReportNo.:EED32N00018102Page5of8 4.4TestLocation Alltestswereperformedat: CentreTestingInternationalGroupCo.,Ltd BuildingC,HongweiIndustrialParkBlock70,Bao'anDistrict,Shenzhen,China Telephone:+86(0)75533683668Fax:+86(0)75533683385 Notestsweresub-contracted. FCCDesignationNo.:CN1164 4.5DeviationfromStandards None. 4.6AbnormalitiesfromStandardConditions None. 4.7OtherInformationRequestedbytheCustomer None. ReportNo.:EED32N00018102Page6of8 5SAREvaluation 5.1RFExposureComplianceRequirement 5.1.1StandardRequirement AccordingtoKDB447498D01GeneralRFExposureGuidancev06 StandaloneSARtestexclusionconsiderations UnlessspecificallyrequiredbythepublishedRFexposureKDBprocedures,standalone1-gheadorbody and10-gextremitySARevaluationforgeneralpopulationexposureconditions,bymeasurementor numericalsimulation,isnotrequiredwhenthecorrespondingSARExclusionThresholdcondition,listed below,issatisfied. Limits The1-gand10-gSARtestexclusionthresholdsfor100MHzto6GHzattestseparationdistances ≤50mmaredeterminedby: [(max.powerofchannel,includingtune-uptolerance,mW)/(min.testseparationdistance,mm)…
Text truncated - open the document above for the full version.
ReportNo.EED32N00018101Page 1 of 25 Appendix:BluetoothLE ReportNo.EED32N00018101Page 2 of 25 Contents Contents ............................................................................................................................................. 2 AppendixA:DTSBandwidth ....................................................................................................... 3 TestResult ................................................................................................................................................ 3 TestGraphs .............................................................................................................................................. 4 AppendixB:OccupiedChannelBandwidth ................................................................................ 6 TestResult ................................................................................................................................................ 6 TestGraphs .............................................................................................................................................. 7 AppendixC:Maximumconductedoutputpower ...................................................................... 9 TestResult ................................................................................................................................................ 9 TestGraphs ............................................................................................................................................ 10 AppendixD:Maximumpowerspectraldensity ....................................................................... 12 TestResult .............................................................................................................................................. 12 TestGraphs ............................................................................................................................................ 13 AppendixE:Bandedgemeasurements .................................................................................... 15 TestResult .............................................................................................................................................. 15 TestGraphs ............................................................................................................................................ 16 AppendixF:ConductedSpuriousEmission .............................................................................. 17 TestResult .............................................................................................................................................. 17 TestGraphs ............................................................................................................................................ 18 AppendixG:DutyCycle ............................................................................................................. 23 TestResult .............................................................................................................................................. 23 TestGraphs ............................................................................................................................................ 24 ReportNo.EED32N00018101Page 3 of 25 AppendixA:DTSBandwidth TestResult Test Mode AntennaChannel DTSBW [MHz] FL[MHz]FH[MHz]Limit[MHz]Verdict BLE_1MAnt1 24020.6082401.7802402.3880.5PASS 24400.6682439.7442440.4120.5PASS 24800.6922479.7482480.4400.5PASS ReportNo.EED32N00018101Page 4 of 25 TestGraphs BLE_1M_Ant1_2402 BLE_1M_Ant1_2440 ReportNo.EED32N00018101Page 5 of 25 BLE_1M_Ant1_2480 ReportNo.EED32N00018101Page 6 of 25 AppendixB:OccupiedChannelBandwidth TestResult Test Mode AntennaChannel OCB [MHz] FL[MHz]FH[MHz]Limit[MHz]Verdict BLE_1MAnt1 24021.04932401.5662402.616---PASS 24401.05942439.5642440.624---PASS 24801.06482479.5622480.627---PASS ReportNo.EED32N00018101Page 7 of 25 TestGraphs BLE_1M_Ant1_2402 BLE_1M_Ant1_2440 ReportNo.EED32N00018101Page 8 of 25 BLE_1M_Ant1_2480 ReportNo.EED32N00018101Page 9 of 25 AppendixC:Maximumconductedoutputpower TestResult TestModeAntennaChannelResult[dBm]Limit[dBm]Verdict BLE_1MAnt1 2402-0.04<=30PASS 2440-0.25<=30PASS 2480-0.24<=30PASS ReportNo.EED32N00018101Page 10 of 25 TestGraphs BLE_1M_Ant1_2402 BLE_1M_Ant1_2440 ReportNo.EED32N00018101Page 11 of 25 BLE_1M_Ant1_2480 ReportNo.EED32N00018101Page 12 of 25 AppendixD:Maximumpowerspectraldensity TestResult TestModeAntennaChannelResult[dBm/3-100kHz]Limit[dBm/3kHz]Verdict BLE_1MAnt1 2402-1<=8PASS 2440-1.18<=8PASS 2480-1.2<=8PASS ReportNo.EED32N00018101Page 13 of 25 TestGraphs BLE_1M_Ant1_2402 BLE_1M_Ant1_2440 ReportNo.EED32N00018101Page 14 of 25 BLE_1M_Ant1_2480 ReportNo.EED32N00018101Page 15 of 25 AppendixE:Bandedgemeasurements TestResult Test Mode Antenna Ch Name Channel Ref Level[dBm] Result [dBm] Limit [dBm] Verdict BLE_1MAnt1 Low2402-0.51-46.33<=-20.51PASS High2480-0.79-45.2<=-20.79PASS ReportNo.EED32N00018101Page 16 of 25 TestGraphs BLE_1M_Ant1_Low_2402 BLE_1M_Ant1_High_2480 ReportNo.EED32N00018101Page 17 of 25 AppendixF:ConductedSpuriousEmission TestResult Test Mode AntennaChannel Freq Range [MHz] RefLevel [dBm] Result [dBm] Limit [dBm] Verdict BLE_1MAnt1 2402 Reference-0.67-0.67---PASS 30~100030~1000-57.504<=-20.674PASS 1000~265 00 1000~26500-42.224<=-20.674PASS 2440 Reference-1.00-1.00---PASS 30~100030~1000-57.407<=-21.005PASS 1000~265 00 1000~26500-39.388<=-21.005PASS 2480 Reference-0.99-0.99---PASS 30~100030~1000-57.463<=-20.993PASS 1000~265 00 1000~26500-36.587<=-20.993PASS ReportNo.EED32N00018101Page 18 of 25 TestGraphs BLE_1M_Ant1_2402_0~Reference BLE_1M_Ant1_2402_30~1000 ReportNo.EED32N00018101Page 19 of 25 BLE_1M_Ant1_2402_1000~26500 BLE_1M_Ant1_2440_0~Reference ReportNo.EED32N00018101Page 20 of 25 BLE_1M_Ant1_2440_30~1000 BLE_1M_Ant1_2440_1000~26500 ReportNo.EED32N00018101Page 21 of 25 BLE_1M_Ant1_2480_0~Reference BLE_1M_Ant1_2480_30~1000 ReportNo.EED32N00018101Page 22 of 25 …
Text truncated - open the document above for the full version.
ReportNo.:EED32N00018101Page1of44 1 COVERPAGE TESTREPORT Product:Wrist-TypeFullyAutomaticDigital BloodPressureMonitor Trademark: N/A Model/Typereference: DBP-8276H SerialNumber: N/A ReportNumber: EED32N00018101 FCCID: 2AQVU0022 DateofIssue: May27,2021 TestStandards: 47CFRPart15SubpartC Testresult: PASS Preparedfor: JOYTECHHealthcareCo.,Ltd. No.365,WuzhouRoad,YuhangEconomicDevelopmentZone, HangzhouCity,311100ZhejiangP.R.China Preparedby : CentreTestingInternationalGroupCo.,Ltd. HongweiIndustrialZone,Bao’an70District, Shenzhen,Guangdong,China TEL:+86-755-33683668 FAX:+86-755-33683385 Compiledby:Reviewedby: VitoheAaronMa Approvedby:Date:May27,2021 DavidWang CheckNo.:4538041742 ReportNo.:EED32N00018101Page 2 of 44 2Content 1COVERPAGE.....................................................................................................................................................................1 2CONTENT.............................................................................................................................................................................2 3VERSION..............................................................................................................................................................................3 4TESTSUMMARY................................................................................................................................................................4 5GENERALINFORMATION...............................................................................................................................................5 5.1C LIENT I NFORMATION ....................................................................................................................................................5 5.2G ENERAL D ESCRIPTIONOF EUT..................................................................................................................................5 5.3T EST C ONFIGURATION ...................................................................................................................................................7 5.4T EST E NVIRONMENT ......................................................................................................................................................7 5.5D ESCRIPTIONOF S UPPORT U NITS ...............................................................................................................................8 5.6T EST L OCATION .............................................................................................................................................................8 5.7M EASUREMENT U NCERTAINTY (95% CONFIDENCELEVELS , K =2)..............................................................................8 6EQUIPMENTLIST..............................................................................................................................................................9 7TESTRESULTSANDMEASUREMENTDATA..........................................................................................................11 7.1A NTENNA R EQUIREMENT .............................................................................................................................................11 7.2C ONDUCTED E MISSIONS .............................................................................................................................................12 7.3M AXIMUM C ONDUCTED O UTPUT P OWER ...................................................................................................................15 7.4DTSB ANDWIDTH .........................................................................................................................................................16 7.5M AXIMUM P OWER S PECTRAL D ENSITY ......................................................................................................................17 7.6B AND E DGEMEASUREMENTSAND C ONDUCTED S PURIOUS E MISSION ....................................................................18 7.7R ADIATED S PURIOUS E MISSION &R ESTRICTEDBANDS ...........................................................................................19 8APPENDIXA......................................................................................................................................................................30 PHOTOGRAPHSOFTESTSETUP..................................................................................................................................31 PHOTOGRAPHSOFEUTCONSTRUCTIONALDETAILS.........................................................................................33 ReportNo.:EED32N00018101Page 3 of 44 3Version VersionNo.DateDescription 00May27,2021Original ReportNo.:EED32N00018101Page 4 of 44 4TestSummary TestItemTestRequirementResult AntennaRequirement 47CFRPart15SubpartCSection 15.203/15.247(c) PASS ACPowerLineConducted Emission 47CFRPart15SubpartCSection 15.207 PASS DTSBandwidth 47CFRPart15SubpartCSection 15.247(a)(2) PASS MaximumConductedOutput Power 47CFRPart15SubpartCSection 15.247(b)(3) PASS MaximumPowerSpectral Density 47CFRPart15SubpartCSection15.247(e)PASS BandEdgeMeasurements 47CFRPart15SubpartCSection 15.247(d) PASS ConductedSpurious Emissions 47CFRPart15SubpartCSection 15.247(d) PASS RadiatedSpuriousEmission& Restrictedbands 47CFRPart15SubpartCSection 15.205/15.209 PASS Remark: CompanyNameandAddressshownonReport,thesample(s)andsampleInformationwereprovidedbythe applicantwhoshouldberesponsiblefortheauthenticitywhichCTIhasn’tverified. ReportNo.:EED32N00018101Page 5 of 44 5GeneralInformation 5.1ClientInformation Applicant: JOYTECHHealthcareCo.,Ltd. AddressofApplicant: No.365,WuzhouRoad,YuhangEconomicDevelopmentZone,Hangzhou City,311100ZhejiangP.R.China Manufacturer: JOYTECHHealthcareCo.,Ltd. AddressofManufacturer: No.365,WuzhouRoad,YuhangEconomicDevelopmentZone,Hangzhou City,311100ZhejiangP.R.China Factory: JOYTECHHealthcareCo.,Ltd. AddressofFactory: No.365,WuzhouRoad,YuhangEco…
Text truncated - open the document above for the full version.
ReportNo.:EED32N00018101Page 31 of 44 PHOTOGRAPHSOFTESTSETUP TestmodelNo.:DBP-8276H RadiatedspuriousemissionTestSetup-1 (Below1GHz) RadiatedspuriousemissionTestSetup-2( Above1GHz) ReportNo.:EED32N00018101Page 32 of 44 ConductedEmissionsTestSetup
| # | Rule Parts | Frequency Range | Power Output |
|---|---|---|---|
| 1 | 15C | 2.40 GHz - 2.48 GHz | 1.00 mW |
Digital Thermometer
Equipment Class
DTS - Digital Transmission SystemAir purifier
Equipment Class
DTS - Digital Transmission SystemArm-type Fully Automatic Digital Blood Pressure Monitor
Equipment Class
DTS - Digital Transmission SystemArm-type Fully Automatic Digital Blood Pressure Monitor
Equipment Class
DTS - Digital Transmission System
Infrared Ear Thermometer
Equipment Class
DTS - Digital Transmission System