Technical Specifications
| # | Rule Parts | Frequency Range | Power Output |
|---|
| 1 | 15C | 2.40 GHz - 2.48 GHz | 100.23 mW |
Grant NotesOutput power listed is conducted.
The Main Unit is approved for mobile and fixed use with respect to RF exposure compliance. The antenna of this transmitter must provide a separation distance of at least 20 cm from all persons.
The Foot Pedal meets the SAR Test Exclusion threshold specified in KDB 447498 and since then, it is authorized for portable operation.
For the Remote Control, the highest reported SAR for body and extremity exposure conditions are 0.301 W/kg and 0.136 W/kg, respectively.
Multi-transmitter configurations, supporting simultaneous transmission, have been evaluated as described in this filling. Other multi-transmitter configurations have not been evaluated and shall be evaluated according to KDB Publication 447498 and §2.947(f), §15.31(h) and §15.31(k) composite system and §2.1 terms and concepts.