User Manual Professional Temperature and Humidity Data Logger V1.4 3 2 4 1 DEC B F GH A Temperature(Digital)Humidity(Digital) 0~100%RH(Non-condesing) 0.1%RH ±3%RH(10~90%RH,25°C) ±4%RH(Others) Internal: 15 mins, Seperated: 5 mins -30~65°C (Without battery) 1.5V AAA (Alkaline battery) About 1 year (Condiction: when the environment temperature is 25°C, backlight and buzzer are off, and Bluetooth is not connected.) Micro USB, DC 5V/2A 96*108*20mm About 120g (Not including battery) 25920 or 500000 groups of readings (Optional from software) -40~125°C ±0.3°C(-10~65°C) ±0.5°C(Others) Temperature(Digital)Humidity(Digital) 0~100%RH(Non-condesing) 0.1°C 0.1%RH P2(Seperated) -40~125°C Temperature(Digital) 0.1°C Temperature(PT1000) 0.1°C P4(Seperated) USB2.0, Bluetooth 4.2 (Only for models with Bluetooth) 1. Product Safety Instructions Before using the instrument, please read the following safety instructions carefully: (1)The temperature and humidity range of the normal working environment of the instrument is -30~65°C, 0~100%RH (non-condensing). If it is over the limited range while using, the instrument may be damaged. (2)The safe working temperature range of ordinary AAA alkaline batteries is -15~50°C. If the instrument needs to be used in an environment beyond this temperature range, it must be used with a suitable battery; if you have any questions, please contact the after-sales service department. (3)For precision instruments, please do not disassemble or repair without permission. (4)Please remove the battery if the instrument will not be used for a long time. ①LCD display ②Turn on/off , Record, Browse time ③Turn the screen on/off , View data , Mark ④External probe port ⑤External probe port ⑥External probe port ⑦External probe port ⑧Reset button ① ② ③ ④ ⑤ ⑥ ⑦ ⑧ Measurement Parameter Measuring Range Resolution Accuracy Response Time(t₉₀) Sample Interval Operation Temperature Storage Temperature Battery Capacity Battery Life Power Connector Size Case Material Protection Class -30~65°C 0.1°C 5 seconds -30~65°C ABS IP20 Memory Capacity Probe Model P1(Internal) -30~65°C Connection 2. Product Selection Guide ±0.3°C(-10~65°C) ±0.5°C(Others) 0.1°C P5(Internal) Temperature(Digital) Weight 3. View 4. Specification 5 5. LCD 6 : Display when the current record point exceeds the high limit. : Display when the current record point exceeds the low limit. : Display when the current or previous record point exceeds limits. 29 17 18 19 20 21 22 23 24 25 26 27 28 32 30 31 Dew point temperature symbol Measurement data display area 2 Temperature and humidity unit Waiting reminder symbol Average kinetic temperature Time unit Measurement data display area 1 Main mode indication Query mode indication Setting mode indication Maximum and average symbol Warning symbols Buzzer alarm activation indication Backlight activation indicator Key lock symbol Battery level indicator Emoji Point mark MK indication Time display area Percent sign Clock File symbol 11 12 13 14 15 16 ①Channel number ②High limit exceeded indication ③Sign for alarm trace ④Low limit exceeded indication ⑤Default sign ⑥External probe ⑦Record status indication ⑧USB communication ⑨USB power supply ⑩Wireless signal indication ● means that the function is supported, ○ means that the function is not supported. C DE P3, P3 P4, P4 F G H P2 P1 P3, P3 P4, P4 ●● ○○ ●● B Model Optional A Probe Type P2 Bluetooth P1 ○○ 29 17 18 19 20 21 22 23 24 25 26 27 28 32 30 31 11 1213 14 15 16 ⑩ ⑨ ⑧ ⑦ ⑥ ⑤ ③ ② ④ ① See the datasheet for details See the datasheet for detailsSee the datasheet for detailsSee the datasheet for detailsSee the datasheet for details 8 7 6.7 Stop Recording (1) Long press the button for 3 seconds to stop recording (function of Stop by Button needs to be set). In addition, for models with Bluetooth, users can also choose to stop in the mobile APP. (2) When the number of recording points reaches the maximum limit, the recording will stop automatically. (3) If the battery is too low, it will automatically stop recording, save data and shut down. (4) The instrument can be stop recording while connected with PC or other equipment (non-adapter) during recording (NODATA function has to be turned off). 6.8 Query Max, Min, Average, High and Low Limits (1) When valid data is recorded (both probes record at least one reading or more), click the button in turn to query the maximum, minimum, average value and the high, low limit. (2) When there is no valid data recorded, click the button in turn to query the high and low limits. (3) After entering the query state, the icon disappears and the icon will be displayed. If there is no any operation for 10 seconds, or pressing any other buttons, it will return to the main interface. The icon disappears and the icon will be displayed. 6.9 Turn Off Buzzer Alarm When the instrument is over-limits, the over-limit indication or of the relevant channel will be displayed. If the time of the buzzer alarm has been set to a valid value, the buzzer will sound. Click any button to turn off the alarm bell. 6.10 Point Mark Operation When the instrument is in the recording state, enter the point marking interface, press and hold the button for 3 seconds to do point marking once, and it can mark 3 points maximum. 6. Quick Guide If there is no operation on the touch button for more than 15 seconds, the instrument will automatically lock and display the icon . If user needs to operate the instrument, must press and hold the button for 3 seconds to unlock it. When the buzzer of the instrument button sounds ”beep...” after the button is short-pressed or long-pressed by user, which means that the operation is valid. 6.1 Turn On When the power supply is restarted, the instrument will automatically turn on. 6.2 Turn Off When the battery power is too low, the instrument will automatically shut down (you must reconnect the power to turn it on). 6.3 Turn On Display When the display is off, press button 3 seconds to turn on the display. 6.4 Turn Off…
Text truncated - open the document above for the full version.
SHENZHEN YOWEXA MEASUREMENT TECHNOLOGY CO.,LTD 16th Floor, Building 2, Guangming Phoenix Plaza, Guangming District, Shenzhen, China Federal Communications Commission Authorization and Evaluation Division Equipment Authorization Branch 7435 Oakland Mills Road Columbia, MD 21046 2026/07/09 U.S. Agent Designation SHENZHEN YOWEXA MEASUREMENT TECHNOLOGY CO.,LTD acknowledges their consent for the following contact located in the United States to act as their agent for service of process for the equipment for which authorization is sought (FCC ID: 2BQEA-YDP20EBE): Name and company of U.S. agent: GLOCERTS TECHNOLOGY INC Physical U.S. address of agent: 19784 SUNKISSED RIDGE DR RIVERSIDE, CA 92507 United States FRN of U.S. agent: 0036339737 Phone of U.S. agent: (626) 588-8039 Email address of U.S. agent: [email protected] GLOCERTS TECHNOLOGY INC acknowledges their obligation to accept service of process on behalf of SHENZHEN YOWEXA MEASUREMENT TECHNOLOGY CO.,LTD with the FRN [0037181252] accepts its obligation to maintain an agent for service of process in the United States for no less than one year after either the grantee has permanently terminated all marketing and importation of the applicable equipment within the U.S., or the conclusion of any Commission-related administrative or judicial proceeding involving the equipment, whichever is later. By signing this form, we confirm the above and that we are aware of the application requirements listed under § 2.911(d)(7). The information provided in this letter is based on the referenced rule parts of Title 47 of the CFR as well as KDB 986446 D01. An expiration date for this letter does not exist. In case the U.S. agent changes, the grantee is obligated to notify the FCC within 30 days via the granting TCB. An expiration date for this letter does not exist. Sincerely, Signature Feng Luo Manager SHENZHEN YOWEXA MEASUREMENT TECHNOLOGY CO.,LTD Signature Huiying Fan Director GLOCERTS TECHNOLOGY INC 48_[mo del] US_Agent_2.911(d)(7) - U, V0 5
FCC ID:2BQEA-YDP20EBE External Photos of the EUT FCC ID:2BQEA-YDP20EBE FCC ID:2BQEA-YDP20EBE
Label and Location Size: L: 2.0cm*W: 1.0cm Label and Location FCC ID : 2BQEA-YDP20EBE
FCC ID:2BQEA-YDP20EBE Internal Photos of the EUT FCC ID:2BQEA-YDP20EBE ANT FCC ID:2BQEA-YDP20EBE
Shenzhen CTA Testing Technology Co., Ltd. Room 106, Building 1, Yibaolai Industrial Park, Qiaotou Community, Fuhai Street, Bao’an District, Shenzhen, China Shenzhen CTA Testing Technology Co., Ltd. Room 106, Building 1, Yibaolai Industrial Park, Qiaotou Community, Fuhai Street, Bao’an District, Shenzhen, China Tel:+86-755 2322 5875 E-mail:[email protected] Web:http://www.cta-test.cn RF Exposure MPE Report Reference No. ..................... : FCC ID............................................. : CTA26062600702 2BQEA-YDP20EBE Compiled by ( position+printed name+signature) . : File administrators Zoey Cao Supervised by ( position+printed name+signature) . : Project Engineer Ace Chai Approved by ( position+printed name+signature) . : RF Manager Eric Wang Date of issue .................................... : Jul. 07, 2026 Testing Laboratory Name .............. : Shenzhen CTA Testing Technology Co., Ltd. Address............................................ : Room 106, Building 1, Yibaolai Industrial Park, Qiaotou Community, Fuhai Street, Bao’an District, Shenzhen, China Applicant’s name ............................ : SHENZHEN YOWEXA MEASUREMENT TECHNOLOGY CO.,LTD Address............................................ : 16th Floor, Building 2, Guangming Phoenix Plaza, Guangming District, Shenzhen, China Standard ........................................... : 47CFR §1.1310 47CFR §2.1091 KDB447498 D01 General RF Exposure Guidance v06 Shenzhen CTA Testing Technology Co., Ltd. All rights reserved. This publication may be reproduced in whole or in part for non-commercial purpses as long as the Shenzhen CTA Testing Technology Co., Ltd. is acknowledged as copyright owner and source of the material. Shenzhen CTA Testing Technology Co., Ltd. takes no responsibility for and will not assume liability for damages resulting from the reader's interpretation of the reproduced material due to its placement and context. Test item description ..................... : Bluetooth Temperature Humidity Data Logger Trade Mark ....................................... : N/A Manufacturer .................................... : SHENZHEN YOWEXA MEASUREMENT TECHNOLOGY CO.,LTD Model/Type reference ...................... : YDP-20EB-E Listed Models ................................... : YDP-20EB, YDP-20B, YDP-10EB, YDP-700EB Rating ............................................... : Input: 5V2A or 4.5V(3x1.5V AAA dry batteries) Result ............................................... : PASS Report No.: CTA26062600702 Page 2 of 8 Shenzhen CTA Testing Technology Co., Ltd. Room 106, Building 1, Yibaolai Industrial Park, Qiaotou Community, Fuhai Street, Bao’an District, Shenzhen, China Tel:+86-755 2322 5875 E-mail:[email protected] Web:http://www.cta-test.cn T E S T R E P O R T Equipment under Test : Bluetooth Temperature Humidity Data Logger Model /Type : YDP-20EB-E Listed Models : YDP-20EB, YDP-20B, YDP-10EB, YDP-700EB Model difference : The PCB board, circuit, structure and internal of these models are the same, only model number and colour are different for these model. So the tests were performed on model YDP-20EB-E and the test results for the main test model can cover the listed models. Applicant : SHENZHEN YOWEXA MEASUREMENT TECHNOLOGY CO.,LTD Address : 16th Floor, Building 2, Guangming Phoenix Plaza, Guangming District, Shenzhen, China Manufacturer : SHENZHEN YOWEXA MEASUREMENT TECHNOLOGY CO.,LTD Address : 16th Floor, Building 2, Guangming Phoenix Plaza, Guangming District, Shenzhen, China Test Result: PASS The test report merely corresponds to the test sample. It is not permitted to copy extracts of these test result without the written permission of the test laboratory. Report No.: CTA26062600702 Page 3 of 8 Shenzhen CTA Testing Technology Co., Ltd. Room 106, Building 1, Yibaolai Industrial Park, Qiaotou Community, Fuhai Street, Bao’an District, Shenzhen, China Tel:+86-755 2322 5875 E-mail:[email protected] Web:http://www.cta-test.cn Contents 1 TEST STANDARDS ............................................................................................................ 4 2 SUMMARY .......................................................................................................................... 5 2.1 General Remarks ..................................................................................................... 5 2.2 Product Description ................................................................................................. 5 2.3 Special Accessories ................................................................................................. 5 2.4 Modifications ............................................................................................................ 5 3 TEST ENVIRONMENT ....................................................................................................... 6 3.1 Address of the test laboratory .................................................................................. 6 3.2 Test Facility .............................................................................................................. 6 3.3 Statement of the measurement uncertainty ............................................................. 6 4 Test limit .............................................................................................................................. 7 4.1 Requirement ............................................................................................................ 7 4.2 MPE Calculation Method ......................................................................................... 7 4.3 Conducted Power Results ....................................................................................... 7 4.4 Manufacturing tolerance .......................................................................................... 7 4.5 Standalone MPE Result ........................................................................................... 8 4.6 Simultaneous Transmission for MPE R…
Text truncated - open the document above for the full version.
1. Antenna Description: Antenna type PCB antenna Model YDP-20EB-E Frequency range 2400-2500MHz Manufacturer SHENZHEN YOWEXA MEASUREMENT TECHNOLOGY CO.,LTD Address 16th Floor, Building 2, Guangming Phoenix Plaza, Guangming District, Shenzhen, China ANT Peak Gain 0.37 dBi 2. Antenna Photo: 3. Test Equipment & Conditions 3.1 Test instrument VSWR is tested using Agilent E5071B connected to a hand mechanism, and the status is free space. 3.2 Test environment and methods 4.Antenna Radiation Pattern 2400MHz 2450MHz 2500MHz
Shenzhen CTA Testing Technology Co., Ltd. Room 106, Building 1, Yibaolai Industrial Park, Qiaotou Community, Fuhai Street, Bao’an District, Shenzhen, China Shenzhen CTA Testing Technology Co., Ltd. Room 106, Building 1, Yibaolai Industrial Park, Qiaotou Community, Fuhai Street, Bao’an District, Shenzhen, China Tel:+86-755 2322 5875 E-mail:[email protected] Web:http://www.cta-test.cn FCC PART 15 SUBPART C TEST REPORT FCC PART 15.247 Report Reference No. ...................: FCC ID. ......................................... : CTA26062600701 2BQEA-YDP20EBE Compiled by ( position+printed name+signature) : File administrators Zoey Cao Supervised by ( position+printed name+signature) : Project Engineer Ace Chai Approved by ( position+printed name+signature) : RF Manager Eric Wang Date of issue ...................................: Jul. 07, 2026 Testing Laboratory Name .............: Shenzhen CTA Testing Technology Co., Ltd. Address ..........................................: Room 106, Building 1, Yibaolai Industrial Park, Qiaotou Community, Fuhai Street, Bao’an District, Shenzhen, China Applicant’s name...........................: SHENZHEN YOWEXA MEASUREMENT TECHNOLOGY CO.,LTD Address ...........................................: 16th Floor, Building 2, Guangming Phoenix Plaza, Guangming District, Shenzhen, China Test specification ......................... : Standard ..........................................: FCC Part 15.247 Shenzhen CTA Testing Technology Co., Ltd. All rights reserved. This publication may be reproduced in whole or in part for non-commercial purposes as long as the Shenzhen CTA Testing Technology Co., Ltd. is acknowledged as copyright owner and source of the material. Shenzhen CTA Testing Technology Co., Ltd. takes no responsibility for and will not assume liability for damages resulting from the reader's interpretation of the reproduced material due to its placement and context. Equipment description ................ : Bluetooth Temperature Humidity Data Logger Trade Mark ......................................: N/A Manufacturer ...................................: SHENZHEN YOWEXA MEASUREMENT TECHNOLOGY CO.,LTD Model/Type reference .....................: YDP-20EB-E Listed Models ..................................: YDP-20EB, YDP-20B, YDP-10EB, YDP-700EB Modulation ......................................: GFSK Frequency .......................................: From 2402MHz to 2480MHz Ratings ............................................: Input: 5V2A or 4.5V(3x1.5V AAA dry batteries) Result ..............................................: PASS Report No.: CTA26062600701 Page 2 of 35 Shenzhen CTA Testing Technology Co., Ltd. Room 106, Building 1, Yibaolai Industrial Park, Qiaotou Community, Fuhai Street, Bao’an District, Shenzhen, China Tel:+86-755 2322 5875 E-mail:[email protected] Web:http://www.cta-test.cn T E S T R E P O R T Equipment under Test : Bluetooth Temperature Humidity Data Logger Model /Type : YDP-20EB-E Listed Models : YDP-20EB, YDP-20B, YDP-10EB, YDP-700EB Model difference : The PCB board, circuit, structure and internal of these models are the same, only model number and colour are different for these model. So the tests were performed on model YDP-20EB-E and the test results for the main test model can cover the listed models. Applicant : SHENZHEN YOWEXA MEASUREMENT TECHNOLOGY CO.,LTD Address : 16th Floor, Building 2, Guangming Phoenix Plaza, Guangming District, Shenzhen, China Manufacturer : SHENZHEN YOWEXA MEASUREMENT TECHNOLOGY CO.,LTD Address : 16th Floor, Building 2, Guangming Phoenix Plaza, Guangming District, Shenzhen, China Test Result: PASS The test report merely corresponds to the test sample. It is not permitted to copy extracts of these test result without the written permission of the test laboratory. Report No.: CTA26062600701 Page 3 of 35 Shenzhen CTA Testing Technology Co., Ltd. Room 106, Building 1, Yibaolai Industrial Park, Qiaotou Community, Fuhai Street, Bao’an District, Shenzhen, China Tel:+86-755 2322 5875 E-mail:[email protected] Web:http://www.cta-test.cn Contents 1 TEST STANDARDS ...................................................................................................................................... 4 2 SUMMARY .................................................................................................................................................... 5 2.1 General Remarks .............................................................................................................................. 5 2.2 Product Description* .......................................................................................................................... 5 2.3 Equipment Under Test ....................................................................................................................... 5 2.4 Short description of the Equipment under Test (EUT) ...................................................................... 5 2.5 EUT configuration .............................................................................................................................. 5 2.6 EUT operation mode ......................................................................................................................... 5 2.7 Block Diagram of Test Setup ............................................................................................................. 6 2.8 Related Submittal(s) / Grant (s) ......................................................................................................... 6 2.9 Modifications ...................................................................................................................................... 6 3 TEST ENVIRONMENT ................................................................................................................................. 7 3.1 Address of the test laboratory ......................................................................................................…
Text truncated - open the document above for the full version.
1 FCC Appendix RF Test Data for BLE FCC ID : 2BQEA-YDP20EBE Company : SHENZHEN YOWEXA MEASUREMENT TECHNOLOGY CO.,LTD EUT : Bluetooth Temperature Humidity Data Logger Model Number : YDP-20EB-E Report No. : CTA26062600701 Reviewed By : Eric Wang 2 Appendix A: DTS Bandwidth Test Result TestMode Antenna Freq(MHz) DTS BW [MHz] FL[MHz] FH[MHz] Limit[MHz] Verdict BLE_1M Ant1 2402 0.644 2401.688 2402.332 0.5 PASS 2440 0.664 2439.676 2440.340 0.5 PASS 2480 0.644 2479.692 2480.336 0.5 PASS 3 Test Graphs BLE_1M_Ant1_2402 BLE_1M_Ant1_2440 BLE_1M_Ant1_2480 4 5 Appendix B: Occupied Channel Bandwidth Test Result TestMode Antenna Freq(MHz) OCB [MHz] FL[MHz] FH[MHz] Limit[MHz] Verdict BLE_1M Ant1 2402 1.0123 2401.5152 2402.5275 --- --- 2440 1.0047 2439.5204 2440.5251 --- --- 2480 1.0039 2479.5178 2480.5217 --- --- 6 Test Graphs BLE_1M_Ant1_2402 BLE_1M_Ant1_2440 BLE_1M_Ant1_2480 7 8 Appendix C: Maximum conducted output power Test Result Peak TestMode Antenna Freq(MHz) Conducted Peak Powert[dBm] Conducted Limit[dBm] Verdict BLE_1M Ant1 2402 -3.21 ≤30 PASS 2440 -3.25 ≤30 PASS 2480 -3.52 ≤30 PASS 9 Appendix D: Maximum power spectral density Test Result TestMode Antenna Freq(MHz) Result[dBm/3kHz] Limit[dBm/3kHz] Verdict BLE_1M Ant1 2402 -19.02 ≤8.00 PASS 2440 -19.02 ≤8.00 PASS 2480 -19.38 ≤8.00 PASS 10 Test Graphs BLE_1M_Ant1_2402 BLE_1M_Ant1_2440 BLE_1M_Ant1_2480 11 12 Appendix E: Band edge measurements Test Result TestMode Antenna ChName Freq(MHz) RefLevel[dBm] Result[dBm] Limit[dBm] Verdict BLE_1M Ant1 Low 2402 -3.96 -57.46 ≤-23.96 PASS High 2480 -4.10 -53.88 ≤-24.1 PASS 13 Test Graphs BLE_1M_Ant1_Low_2402 BLE_1M_Ant1_High_2480 14 Appendix F: Conducted Spurious Emission Test Result TestMode Antenna Freq(MHz) FreqRange [MHz] RefLevel [dBm] Result[dBm] Limit[dBm] Verdict BLE_1M Ant1 2402 Reference -4.75 -4.75 --- PASS 30~1000 -4.75 -68.8 ≤-24.75 PASS 1000~26500 -4.75 -45.55 ≤-24.75 PASS 2440 Reference -4.95 -4.95 --- PASS 30~1000 -4.95 -64.72 ≤-24.95 PASS 1000~26500 -4.95 -44.91 ≤-24.95 PASS 2480 Reference -5.58 -5.58 --- PASS 30~1000 -5.58 -68.29 ≤-25.58 PASS 1000~26500 -5.58 -46.55 ≤-25.58 PASS 15 Test Graphs BLE_1M_Ant1_2402_0~Reference BLE_1M_Ant1_2402_30~1000 BLE_1M_Ant1_2402_1000~26500 16 BLE_1M_Ant1_2440_0~Reference BLE_1M_Ant1_2440_30~1000 17 BLE_1M_Ant1_2440_1000~26500 BLE_1M_Ant1_2480_0~Reference BLE_1M_Ant1_2480_30~1000 18 BLE_1M_Ant1_2480_1000~26500 19 Appendix G: Duty Cycle Test Result TestMode Antenna Freq(MHz) ON Time [ms] Period [ms] X DC [%] xFactor Limit Verdict BLE_1M Ant1 2402 0.41 0.62 0.6613 66.13 1.80 --- --- 2440 0.41 0.62 0.6613 66.13 1.80 --- --- 2480 0.41 0.62 0.6613 66.13 1.80 --- --- 20 Test Graphs BLE_1M_Ant1_2402 BLE_1M_Ant1_2440 BLE_1M_Ant1_2480 21 22 Appendix H: Emissions in Restricted Bands Test Result TestMod e Antenn a ChNam e Freq(MH z) Detector Freq [MHz] Result [dBm] Gain [dB] DC Factor [dB] Result [dBuV/m] Limit [dBuV/m] Verdict BLE_1M Ant1 Low 2402 AV 2310.000 -60.25 2.00 1.80 38.75 ≤54 PASS AV 2367.515 -58.85 2.00 1.80 40.15 ≤54 PASS AV 2390.000 -60.03 2.00 1.80 38.97 ≤54 PASS Peak 2310.000 -52.99 2.00 0.00 44.21 ≤74 PASS Peak 2375.915 -48.04 2.00 0.00 49.16 ≤74 PASS Peak 2390.000 -49.92 2.00 0.00 47.28 ≤74 PASS High 2480 AV 2483.500 -58.39 2.00 1.80 40.61 ≤54 PASS AV 2484.800 -58.08 2.00 1.80 40.92 ≤54 PASS AV 2500.000 -59.36 2.00 1.80 39.64 ≤54 PASS Peak 2483.500 -51.33 2.00 0.00 45.87 ≤74 PASS Peak 2485.200 -47.47 2.00 0.00 49.73 ≤74 PASS Peak 2500.000 -51.16 2.00 0.00 46.04 ≤74 PASS Note: 1. The limit in dBm for average detector is conversion from 54dBuV/m, according to 15.209(a). The limit in dBm for peak detector is 20dB above the limit of average detector in dBm. 23 Test Graphs BLE_1M_Ant1_Low_2402_AV BLE_1M_Ant1_Low_2402_Peak BLE_1M_Ant1_High_2480_AV 24 BLE_1M_Ant1_High_2480_Peak
FCC ID:2BQEA-YDP20EBE Test Setup Photos of the EUT FCC ID:2BQEA-YDP20EBE
| # | Rule Parts | Frequency Range | Power Output |
|---|---|---|---|
| 1 | 15C | 2.40 GHz - 2.48 GHz | 500.00 µW |