Text extracted from the exhibit documents filed with the FCC. Open a document above to read the original.
TEST SERVICES EMC EVALUATION OF THE SIEMENS MEDICAL SOLUTIONS TWELVE LEAD BIOMEDICAL TELEMETRY 608-614MHZ TRANSMITTER/RECEIVER Date: AUGUST 9, 2002 Test Report Number: TR3402.US.02 IN ACCORDANCE WITH FCC PART 15 SUBPART B AND PART 15.242 Prepared For: SIEMENS MEDICAL SOLUTIONS 16 ELECTRONICS AVENUE DANVERS, MASSACHUSETTS 01923 ATTENTION: JIM GRECO Prepared By: RONALD H. CROOKER CHOMERICS TEST SERVICES 77 DRAGON COURT WOBURN, MASSACHUSETTS 01888 Test Technician or Engineer: ___________________________ CTS Approved Signatory: ___________________________ This report shall not be reproduced except in full without the written approval of Chomerics Test Services. TEST SERVICES Siemens Medical Solutions Twelve Lead Biomedical Telemetry 608-614MHz Transmitter/Receiver Document #: TR3402.US.02 Date: August 9, 2002 Page 2 of 21 TABLE OF CONTENTS 1.0 General 1.1 Introduction 1.1.1 Purpose 1.1.2 Requirements 1.2 Test Summary 1.2.1 Summary of Recommendations 1.3 Administrative Data 1.3.1 Test Facility 1.3.2 Equipment Calibration 1.3.3 Test Personnel 1.4 Test Set-up 1.4.1 Test Site Matrix 1.4.2 Test Site Descriptions 1.4.3 Equipment Under Test 1.4.4 Block Diagram 1.5 Pass/Fail Criteria 2.0 Emissions Tests Performed 2.1 FCC Part 15 Subpart B Radiated Electromagnetic Emissions and Part 15.242 Field Strength Measurement 2.1.1 Equipment Used 2.1.2 Test Conditions 2.1.3 Test Method 2.1.4 Results 2.1.5 Test data 2.1.6 Photographic Documentation 2.2 FCC Part 15 Subpart B Conducted Electromagnetic Emissions 2.2.1 Equipment Used 2.2.2 Test Conditions 2.2.3 Test Method 2.2.4 Results 2.2.5 Test Data 2.2.6 Photographic Documentation Appendix A: Test Log TEST SERVICES Siemens Medical Solutions Twelve Lead Biomedical Telemetry 608-614MHz Transmitter/Receiver Document #: TR3402.US.02 Date: August 9, 2002 Page 3 of 21 LIST OF DEFINITIONS/ABBREVIATIONS AC Alternating Current BB Broadband BW Bandwidth cm Centimeter CPU Calibrate Prior to Use dB Decibel DC Direct Current EMC Electromagnetic Compatibility EMI Electromagnetic Interference ER Electric Radiation EUT Equipment Under Test GHz Gigahertz Hz Hertz I-face Interface kHz Kilohertz m Meter MHz Megahertz mm Millimeter mS Millisecond mV Millivolt MR Magnetic Radiation NB Narrowband NCR No Calibration Required PLC Power Line Conduction PPS Pulses Per Second uF Microfarad uH Microhenry uS Microsecond uV Microvolt UWC Use With Calibrated Equipment TEST SERVICES Siemens Medical Solutions Twelve Lead Biomedical Telemetry 608-614MHz Transmitter/Receiver Document #: TR3402.US.02 Date: August 9, 2002 Page 4 of 21 1.0 GENERAL 1.1 Introduction 1.1.1 Purpose The purpose of this report is to document the performance of the Siemens Medical Solutions Twelve Lead Biomedical Telemetry 608-614MHz Transmitter/Receiver during an electromagnetic interference (EMI) test and record the test requirements and procedures used. At the request of Siemens Medical Solutions, the tests were performed by Chomerics Test Services (CTS) of Woburn, Massachusetts. The assessment will determine the compliance or non-compliance with the requirements set up by the Electromagnetic Interference (EMI) Standard FCC Part 15 Subpart B and Part 15.242. The Radiated and Conducted Emission Standard FCC Part 15 Subpart B is designated for ITE, Information Technology Equipment. FCC Part 15.242 is designated for Intentional Radiators. Jim Greco from Siemens Medical Solutions was present during testing. Testing was performed on July 22, 2002 under purchase order number 154884. 1.1.2 Requirements The requirements for the sequence of tests performed on the Twelve Lead Biomedical Telemetry 608- 614MHz Transmitter/Receiver are as follows: FCC Part 15 Subpart B Radiated and Conducted Electromagnetic Emissions FCC Part 15 radiated and conducted emissions, Subpart B for Information Technology Equipment. FCC Part 15.242 Field Strength Measurement FCC Part 15.242 Intentional Radiators-Medical 608-614MHz Band. TEST SERVICES Siemens Medical Solutions Twelve Lead Biomedical Telemetry 608-614MHz Transmitter/Receiver Document #: TR3402.US.02 Date: August 9, 2002 Page 5 of 21 1.2 TEST SUMMARY The terms "Passed" or "Failed" in this section are intended to guide the reader as to whether or not the EUT met the minimum requirements that can be interpreted from the FCC Part 15 Subpart B and Part 15.242 Standard as defined in Section 1.5. The "Results" paragraph in each test section to follow and the test data sheets will outline specifically how the EUT performed during each test. FCC Part 15 Subpart B Radiated and Conducted Emissions PASSED FCC Part 15.242 Field Strength Measurement PASSED 1.2.1 Summary of Recommendations The Siemens Medical Solutions Twelve Lead Biomedical Telemetry 608-614MHz Transmitter/Receiver will require not modifications in order to ensure compliance with the Electromagnetic Interference Standard FCC Part 15 Subpart B and Part 15.242. Please note that if any modifications and or fixes were implemented to the EUT to achieve compliance, other approaches to solving the problem may exist. In addition, any EMI/EMC shielding products listed in this report may be substituted with an equivalent. 1.3 Administrative Data 1.3.1 Test Facility Chomerics Test Services in Woburn, Massachusetts is an American Association for Laboratory Accreditation (A2LA) accredited facility as defined on Certification Number 1980-01. For Emissions and Immunity testing, the Scope of Accreditation is limited to the following tests: CFR 47, FCC Part 15 Subpart B, CISPR 11, EN 55011, CISPR 13, EN55013, CISPR 14, EN55014-1, CISPR 22, EN55022, AS/NZS 3548, CNS 13438, CNS 13783-1, VCCI, EN 61000-3-2, EN 61000-3-3, EN 50081-1, EN55081-2, EN61000-6-3, EN 61000-6-4, EN 61000-4-2, EN 61000-4-3, EN61000-4-4, EN 61000-4-5, EN 61000-4-6, EN61000-4-8, EN 61000-4-11, EN 50082-1, EN 50082-2, EN 61000-6-1, EN 61000-6-2, IEC/EN 60601-1-2, EN 300 386, EN 61362-1, CISPR 24, EN55024, CISPR 14, EN 55014-2, EN 50083-2, EN 55103-1, and EN 55103-2. Any tests in this reβ¦
Text truncated - open the document above for the full version.
Siemens Medical Solutions Twelve Lead Biomedical Telemetry 608-614MHz Transmitter/Receiver Photographic Documentation CUSTOMER: SIEMENS MEDICAL SOLUTIONS DATE: 22 JULY 2002 EQUIPMENT: TWELVE LEAD BIOMEDICAL TELEMETRY 608-614MHZ TRANSMITTER/RECEIVER T EST NUMBER: 1 T ESTED BY: RONALD H. CROOKER COUPLING DEVICE: EMCO O PERATING MODE: NORMAL TEST SPEC: FCC PART 15 SUBPART B AND P ART 15.242 Photograph Description: Radiated set-up FORM CTS-PHOTO Siemens Medical Solutions Twelve Lead Biomedical Telemetry 608-614MHz Transmitter/Receiver Photographic Documentation CUSTOMER: SIEMENS MEDICAL SOLUTIONS DATE: 22 JULY 2002 EQUIPMENT: TWELVE LEAD BIOMEDICAL TELEMETRY 608-614MHZ TRANSMITTER/RECEIVER T EST NUMBER: 1 T ESTED BY: RONALD H. CROOKER COUPLING DEVICE: EMCO O PERATING MODE: NORMAL TEST SPEC: FCC PART 15 SUBPART B AND P ART 15.242 Photograph Description: Radiated set-up FORM CTS-PHOTO Siemens Medical Solutions Twelve Lead Biomedical Telemetry 608-614MHz Transmitter/Receiver Photographic Documentation CUSTOMER: SIEMENS MEDICAL SOLUTIONS DATE: 22 JULY 2002 E QUIPMENT: TWELVE LEAD BIOMEDICAL TELEMETRY 608-614MHZ TRANSMITTER/RECEIVER TEST NUMBER: 2 T ESTED BY: RONALD H. CROOKER COUPLING DEVICE: LISN O PERATING MODE: NORMAL TEST SPEC: FCC PART 15 SUBPART B Photograph Description: Conducted set-up FORM CTS-PHOTO
77 Dragon Court Β· Woburn, Massachusetts Β· United States
| # | Rule Parts | Frequency Range | Power Output |
|---|---|---|---|
| 1 | 15C | 608 MHz - 614 MHz | - |

Bio-Telemetry Transmitter
Equipment Class
DXX - Part 15 Low Power Communication Device Transmitter
Biomedical Telemetry Transmitter.
Equipment Class
DXX - Part 15 Low Power Communication Device TransmitterMedical Telemetry Transmitter
Equipment Class
DXX - Part 15 Low Power Communication Device TransmitterCYY - Communications Receiver used w/Pt 15 Transmitter
Equipment Class
CYY - Communications Receiver used w/Pt 15 TransmitterTransmitter frequency has been limited because of bandwidth requirements of Section 15.176(b).
Equipment Class
EST - Unkown (From ALTOS)