Revision 1.2 Document No. 14480 Page 1 Aera CT 3G Mobile Cardiac Monitor Operator Manual Revision 1.2 Document No. 14480 Page 2 Contents Device Overview ........................................................................................................................................................3 Description .......................................................................................................................................................................... 3 Indications for Use .............................................................................................................................................................. 4 Contraindications ................................................................................................................................................................ 4 Recorder Components ........................................................................................................................................................ 5 Equipment Symbols ............................................................................................................................................................ 7 Precautions ......................................................................................................................................................................... 8 Operation ..................................................................................................................................................................9 Battery Installation Instructions.......................................................................................................................................... 9 Initial Device Setup ........................................................................................................................................................... 10 User Interface ................................................................................................................................................................... 11 Starting a Study ................................................................................................................................................................. 12 Monitoring a Study ........................................................................................................................................................... 17 Ending a Study ................................................................................................................................................................... 19 Electrode Application ........................................................................................................................................................ 22 Patient Use Precautions .................................................................................................................................................... 24 Maintenance ........................................................................................................................................................... 25 Charging ............................................................................................................................................................................ 25 Storage .............................................................................................................................................................................. 25 Cleaning ............................................................................................................................................................................. 26 Testing ............................................................................................................................................................................... 26 Troubleshooting ...................................................................................................................................................... 27 Error Messages.................................................................................................................................................................. 27 Server Errors ..................................................................................................................................................................... 28 Wireless Compliance ................................................................................................................................................ 29 Specifications ........................................................................................................................................................... 30 TZ Medical Warranty ................................................................................................................................................ 32 Limited Warranty .............................................................................................................................................................. 32 Obtaining Warranty Service .............................................................................................................................................. 32 Accessories .............................................................................................................................................................. 33 Revision 1.2 Document No. 14480 Page 3 Device Overview Description Caution: US Federal law restricts this device to sale by or on the order of a physician. The Aera CT is a battery operated ambulatory electrocardiograph recorder for use in mobile cardiovascular telemetry. It is capable of storing 3 channels of patient ECG data acquired from three or five surface electrodes adhered to the body for over 28 days of continuous recording. The Aera CT employs a rechargeable Lithium- Ion battery to achieve such an extende…
Text truncated - open the document above for the full version.
TZ Medical – TZMR3G FCC ID: ZIMTZMR3G IC ID: 9647A‐TZMR3G
A��� CT 3G FCC ID: ZIMTZMR3G IC: 9647A-TZMR3G Model: TZMR3G RX Only Distributed by Philips Manufactured by TZ Medical 17750 SW Upper Boones Ferry Rd. #150 Portland, OR 97224 Manufactured by TZ Medical 17750 SW Upper Boones Ferry Rd. #150 Portland, OR 97224 IMEI: 1234567891234567 FCC ID: 1234564 1234564 Barcode SN: 8000001 Front Label - .01“ Scratch Resistant Polycarbonate, - Speaker and mic holes, display window, LED‘s show through - ~.002“ Adhesive Back Label - .01“ Scratch Resistant Polycarbonate - Serial label window - ~.002“ Adhesive Display protector - .03“ Polycarbonate - ~.002“ Adhesive border, no graphics Cap touch spacer - .03“ Polycarbonate - No adhesive Note: Reduce outer die-line dimensions to t case Portland, OR 97224 Manufactured by: TZMedical FCC ID: ZIMTZMR3G IC: 9647A-TZMR3G Model: TZMR3G RX Only RX Only FCC ID: ZIMTZMR3G IC: 9647A-TZMR3G Model: TZMR3G Distributed by RCS Manufactured by TZ Medical Portland, OR 97224 e holter CAT RX Only Distributed by Philips Manufactured by TZ Medical 17750 SW Upper Boones Ferry Rd. #150 Portland, OR 97224 Manufactured by TZ Medical 17750 SW Upper Boones Ferry Rd. #150 Portland, OR 97224 FCC ID: ZIMTZMR3G IC: 9647A-TZMR3G Model: TZMR3G IMEI: 1234567891234567 FCC ID: 1234564 1234564 Barcode SN: 8000001 Front Label - .01“ Scratch Resistant Polycarbonate, - Speaker and mic holes, display window, LED‘s show through - ~.002“ Adhesive Back Label - .01“ Scratch Resistant Polycarbonate - Serial label window - ~.002“ Adhesive Display protector - .03“ Polycarbonate - ~.002“ Adhesive border, no graphics Cap touch spacer - .03“ Polycarbonate - No adhesive Note: Reduce outer die-line dimensions to t case 1.915 0 0 2.413 DO NOT SCALE DRAWING 7414-02 INTERPRET GEOMETRIC THE INFORMATION CONTAINED IN THIS AH PART OR AS A WHOLE WITHOUT THE 1 D C B A AB C D 1 2 3 4 5 6 7 8 8 7 3 Bottom Label 2 UNLESS OTHERWISE SPECIFIED: TZ MEDICAL INC. ANY REPRODUCTION IN IS PROHIBITED. 5 DRAWING IS THE SOLE PROPERTY OFWRITTEN PERMISSION OF TZ MEDICAL INC TOLERANCING PER: MATERIAL FINISH COMMENTS: DRAWN BY TZ Medical Inc. TITLE: SIZE B DWG. NO. REV WEIGHT: SCALE: 1:1 4 PROPRIETARY AND CONFIDENTIAL (503) 639-0282 SHEET 2 OF 2 4 6 DIMENSIONS ARE IN INCHESTOLERANCES:FRACTIONAL 1/64 ANGULAR: MACH 0.5 BEND 1 TWO PLACE DECIMAL 0.01 THREE PLACE DECIMAL 0.001 GRAPHICS:0.002" Adhesive: 3M 467MPNOTE: THIS PART MUST COMPLY WITH THERefer to file# 10133 "Aera Labels" for printing graphicsRuff Tex overprint DOCUMENT # 11660 COSMETIC STANDARD FOR FINISHED PARTSMATERIAL:0.10" clear Marnot XL Polycarbonate VENDOR: Craftsman VENDOR PART NUMBER: 7414-02 Revision 4 change ADHESIVE THIS SIDE 1.410 0.910 THIS SIDEGRAPHICS CLEAR WINDOW,NO ADHESIVEBOTH SIDES .010 R 0 . 8 6 4 0.420 0.252 .010 R 0 . 1 7 5 0.187 X 5 0 0.280 1.859 3.541 3.569 3.539 3.490 3.424 3.343 3.252 3.152 3.829 2.972 2.499 3.444 SCREEN PRINT THIS SIDE DISPLAY ACTIVE AREA. DO NOT SCREEN PRINT IN THIS AREA. 0.010 ±0.002 0 2.278 0.300 1.601 1.701 1.792 1.873 1.939 1.988 2.018 0.312 1.579 0.515 0.988 1.460 THIS SIDE OPTICALLY BONDED TO CAP TOUCH SPACER (7414-3X-02) FIVE LIGHT PERMISSIVE AREAS FOR CAPACATIVE TOUCH LED'S R0.090 X3 R0.780 A 0.020 ±0.002 X8 THRU DETAIL A SCALE 10 : 1 D C B A A B C D 1 2 3 4 5 6 7 8 8 7 6 5 4 3 2 1 THE INFORMATION CONTAINED IN THIS DRAWINGIS THE SOLE PROPERTY OF TZ MEDICAL INC. ANY REPRODUCTION IN PART OR AS A WHOLE WITHOUT THE WRITTEN PERMISSION OF TZ MEDICAL INC IS PROHIBITED. PROPRIETARY AND CONFIDENTIAL DRAWN BY DWG NO: REVISION: SCALE: 1:0.75 REV-05 LHOOVER DESCRIPTION: AERA CT FRONT LABEL SUBSTRATE SHEET 2 OF 6 7414-3X-00 DO NOT SCALE DRAWING UNLESS OTHERWISE SPECIFIED: DIMENSIONS ARE IN INCHES TOLERANCES: FRACTIONAL 1/64 ANGULAR: MACH 0.5 BEND1 ONE PLACE DECIMAL 0.1 TWO PLACE DECIMAL 0.01 THREE PLACE DECIMAL 0.005 FINISH: FREE FROM MARKS OR SCRATCHES 7414-3X PART NO: MATERIAL: CLEAR POLYCARBONATE OR MFG EQUIV. 7 16 9 9 13 0 0 0 2 6 24 2 8 3 20 20.4 ±0.5 1.50 TYP 9.0 1.0 TYP 2 TYP 9.0 EXAMPLE SERIAL LABEL WHITE TEXT ON ABLACK BACKGROUND ONE PLACE DECIMAL D C B A AB C D 1 2 3 4 5 6 7 8 8 7 65 4 3 2 1 0.1 TWO PLACE DECIMAL DIRECT THERMAL ADHESIVE IS PROHIBITED. PROPRIETARY AND CONFIDENTIAL WRITTEN PERMISSION OF TZ MEDICAL INC TOLERANCING PER: MATERIAL FINISH COMMENTS: DRAWN BY TZ Medical Inc. TITLE: SIZE B DWG. NO. REV WEIGHT: B RW Serial Label SHEET 23 OF 37 7414-03 DO NOT SCALE DRAWING INTERPRET GEOMETRIC PART OR AS A WHOLE WITHOUT THE TZ MEDICAL INC. ANY REPRODUCTION IN DRAWING IS THE SOLE PROPERTY OFTHE INFORMATION CONTAINED IN THIS 0.01 (503) 639-0282 UNLESS OTHERWISE SPECIFIED: DIMENSIONS ARE IN MMTOLERANCES:FRACTIONAL 1/64 ANGULAR: MACH 0.5 BEND 1 WHOLE NUMBER: 0.5 This 7 digit number is the device serial number IMEI: SN ICCID: This field is used to identify which wireless module has been installed This 15 digit number is a unique identifier printed on each wireless module. PN in the finished device. which matches the digital PCB serial number. in the finished device. 8501 ‐ 01 This 20 digit number is a unique identifier printed on each SIM card. This field is used to identify which SIM card has been installed This is the TZ Medical finished part number NOT the Aera CT Box Build part number 0.004 MAX SN;ICCID;IMEI;ModelScannable 2D Barcode created in Semacode with the following format (with no spaces): White background behind the 2D barcode. 25.40 38.10
TZ Medical – TZMR3G FCC ID: ZIMTZMR3G IC ID: 9647A‐TZMR3G
FCC Listed #: A2LA Accredited IC Recognized # 3462B-1 CETECOM Inc. 411 Dixon Landing Road Milpitas, CA 95035 U.S.A. Phone: + 1 (408) 586 6200 Fax: + 1 (408) 586 6299 E-mail: [email protected] http://www.cetecom.com CETECOM Inc. is a Delaware Corporation with Corporation number: 2905571 V5.2 2013-02-15 © Copyright by CETECOM SAR Test Report FOR: Manufacturer: TZ Medical, Inc. Model Number: TZMR3G FCC ID: ZIMTZMR3G IC Certification Number: 9647A-TZMR3G Test Report #: SAR_TZMED_007_14001_FCCIC_rev1 Date of Report: October 27, 2014 Test Report #: SAR_TZMED_007_14001_FCCIC_rev1 FCC ID: ZIMTZMR3G Date of Report: October 27, 2014 IC Cert. No.: 9647A-TZMR3G Page 2 of 25 V5.1 2013-01-09 This report shall not be reproduced except in full without the written approval of: CETECOM Inc. SAR 411 Dixon Landing Road Milpitas, CA 95035 U.S.A. TABLE OF CONTENTS 1. Assessment ................................................................................................................................. 4 2. Administrative Data .................................................................................................................. 5 2.1. Identification of the Testing Laboratory Issuing the SAR Test Report ..................... 5 2.2. Identification of the Client .............................................................................................. 5 2.3. Identification of the Manufacturer ................................................................................. 5 3. Equipment under Test (EUT) .................................................................................................. 6 3.1. General Specification of the Equipment under Test .................................................... 6 3.2. Antenna Information ....................................................................................................... 7 3.3. Technical Specification of Supported Radios ................................................................ 7 3.4. Identification of the Equipment Under Test (EUT) ...................................................... 8 3.5. Identification of Accessory equipment ........................................................................... 8 3.6. Maximum SAR values ..................................................................................................... 8 4. Subject of Investigation ............................................................................................................. 9 4.1. The IEEE Standard C95.1 , FCC Exposure Criteria, and IC Exposure Criteria ..... 9 4.2. SAR Limit ......................................................................................................................... 9 5. Measurement Procedure ......................................................................................................... 10 5.1. General Requirements ................................................................................................... 10 5.2. Body-worn and Other Configurations ......................................................................... 10 5.3. Procedure for assessing the peak spatial-average SAR .............................................. 11 5.4. Determination of the largest peak spatial-average SAR ............................................ 13 6. The Measurement System ....................................................................................................... 14 6.1. Robot system specification ............................................................................................ 14 6.2. Isotropic E-Field Probe for Dosimetric Measurements ............................................. 15 6.3. Data Acquisition Electronics ......................................................................................... 15 6.4. Phantoms ........................................................................................................................ 15 6.5. Interpolation and Extrapolation schemes .................................................................... 15 7. Uncertainty Assessment .......................................................................................................... 16 Test Report #: SAR_TZMED_007_14001_FCCIC_rev1 FCC ID: ZIMTZMR3G Date of Report: October 27, 2014 IC Cert. No.: 9647A-TZMR3G Page 3 of 25 V5.1 2013-01-09 This report shall not be reproduced except in full without the written approval of: CETECOM Inc. SAR 411 Dixon Landing Road Milpitas, CA 95035 U.S.A. 7.1. Measurement Uncertainty Budget According to IEEE 1528:2013 ........................... 16 7.1. Measurement Uncertainty Budget According to EN 62209 ....................................... 17 8. Test results summary .............................................................................................................. 18 8.1. Conducted Average Output Power .............................................................................. 18 8.2. Stand-Alone SAR Evaluation Exclusion ...................................................................... 21 8.3. Test Positions and Configurations ................................................................................ 21 8.4. SAR Results for Body .................................................................................................... 22 8.5. SAR Measurement Variability ..................................................................................... 23 8.6. Dipole verification .......................................................................................................... 23 9. References ................................................................................................................................ 24 10. Report History ......................................................................................................................... 25 Appendices: Appendix A – Plots Appendix B – Antenna location, Test Setup Photos Appendix C – Tissue liquid parame…
Text truncated - open the document above for the full version.
FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix A - Plots Page 1 of 18 Plot 1 Date/Time: 9/12/2014 12:24:28 PM Test Laboratory: Cetecom Inc. SAR 1 Lab DUT: TZ Medical; Type: ECG Monitor; Serial: IMEI:35150205.018524.809 Communication System: GPRS-FDD (TDMA, GMSK, TN 0-1-2-3); Frequency: 836.6 MHz Medium: MSL900_Batch 100818-1 Medium parameters used: f = 837 MHz; σ = 0.988 mho/m; ε r = 53.352; ρ = 1000 kg/m 3 Phantom section: Flat Section Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007) Procedure Notes: Test Technician: Kathy; Air Temperature: 22.2C; Medium Temperature: 21.7C; Comments: ; DASY Configuration: Probe: ES3DV3 - SN3244; ConvF(6.15, 6.15, 6.15); Calibrated: 3/19/2014; Sensor-Surface: 3mm (Mechanical Surface Detection), Sensor-Surface: 4mm (Mechanical Surface Detection), z = 2.0, 32.0 Electronics: DAE4 Sn1265; Calibrated: 1/29/2014 Phantom: ELI 4.0; Type: QDOVA001BA; Serial: xxxx DASY52 52.8.1(838); Flat-Section 9-12/Leather_Back 0mm/Area Scan (12x14x1): Measurement grid: dx=12mm, dy=12mm Maximum value of SAR (measured) = 0.0611 mW/g Flat-Section 9-12/Leather_Back 0mm/Zoom Scan (10x8x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 7.919 V/m; Power Drift = -0.00 dB Peak SAR (extrapolated) = 0.073 mW/g SAR(1 g) = 0.051 mW/g; SAR(10 g) = 0.035 mW/g Maximum value of SAR (measured) = 0.0547 mW/g 0 dB = 0.0611 mW/g = -24.28 dB mW/g FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix A - Plots Page 2 of 18 Plot 2 Date/Time: 9/18/2014 12:02:14 PM Test Laboratory: Cetecom Inc. SAR 1 Lab DUT: TZ Medical; Type: ECG Monitor; Serial: IMEI:35150205.018524.809 Communication System: GPRS-FDD (TDMA, GMSK, TN 0-1-2-3); Frequency: 836.6 MHz Medium: MSL900_Batch 100818-1 Medium parameters used: f = 837 MHz; σ = 1.01 mho/m; ε r = 55.101; ρ = 1000 kg/m 3 Phantom section: Flat Section Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2011) Procedure Notes: Test Technician: Kathy; Air Temperature: 22C; Medium Temperature: 20.8C; Comments: ; DASY Configuration: Probe: ES3DV3 - SN3244; ConvF(6.15, 6.15, 6.15); Calibrated: 3/19/2014; Sensor-Surface: 3mm (Mechanical Surface Detection), Sensor-Surface: 4mm (Mechanical Surface Detection), z = 2.0, 32.0 Electronics: DAE4 Sn1265; Calibrated: 1/29/2014 Phantom: ELI 4.0; Type: QDOVA001BA; Serial: xxxx DASY52 52.8.1(838); Flat-Section 9-17/Plastic_Back 0mm/Area Scan (12x14x1): Measurement grid: dx=12mm, dy=12mm Maximum value of SAR (measured) = 0.364 mW/g Flat-Section 9-17/Plastic_Back 0mm/Zoom Scan (8x8x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 17.048 V/m; Power Drift = 0.03 dB Peak SAR (extrapolated) = 0.414 mW/g SAR(1 g) = 0.317 mW/g; SAR(10 g) = 0.233 mW/g Maximum value of SAR (measured) = 0.334 mW/g 0 dB = 0.364 mW/g = -8.78 dB mW/g FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix A - Plots Page 3 of 18 Plot 3 Date/Time: 9/18/2014 12:43:12 PM Test Laboratory: Cetecom Inc. SAR 1 Lab DUT: TZ Medical; Type: ECG Monitor; Serial: IMEI:35150205.018524.809 Communication System: GPRS-FDD (TDMA, GMSK, TN 0-1-2-3); Frequency: 836.6 MHz Medium: MSL900_Batch 100818-1 Medium parameters used: f = 837 MHz; σ = 1.01 mho/m; ε r = 55.101; ρ = 1000 kg/m 3 Phantom section: Flat Section Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2011) Procedure Notes: Test Technician: Kathy; Air Temperature: 22C; Medium Temperature: 20.7C; Comments: ; DASY Configuration: Probe: ES3DV3 - SN3244; ConvF(6.15, 6.15, 6.15); Calibrated: 3/19/2014; Sensor-Surface: 3mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 4mm (Mechanical Surface Detection), z = 2.0, 32.0 Electronics: DAE4 Sn1265; Calibrated: 1/29/2014 Phantom: ELI 4.0; Type: QDOVA001BA; Serial: xxxx DASY52 52.8.1(838); Flat-Section 9-17/Plastic_Front 0mm/Area Scan (12x14x1): Measurement grid: dx=12mm, dy=12mm Maximum value of SAR (measured) = 0.350 mW/g Flat-Section 9-17/Plastic_Front 0mm/Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 6.793 V/m; Power Drift = 0.07 dB Peak SAR (extrapolated) = 0.413 mW/g SAR(1 g) = 0.316 mW/g; SAR(10 g) = 0.227 mW/g Maximum value of SAR (measured) = 0.336 mW/g 0 dB = 0.350 mW/g = -9.12 dB mW/g FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix A - Plots Page 4 of 18 Plot 4 Date/Time: 9/18/2014 8:17:09 AM Test Laboratory: Cetecom Inc. SAR 1 Lab DUT: TZ Medical; Type: ECG Monitor; Serial: IMEI:35150205.018524.809 Communication System: GPRS-FDD (TDMA, GMSK, TN 0-1-2-3); Frequency: 1880 MHz Medium: MSL1900_Batch 100824-3 Medium parameters used: f = 1880 MHz; σ = 1.555 mho/m; ε r = 52.213; ρ = 1000 kg/m 3 Phantom section: Flat Section Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007) Procedure Notes: Test Technician: Kathy; Air Temperature: 22C; Medium Temperature: 21.2C; Comments: ; DASY Configuration: Probe: ES3DV3 - SN3244; ConvF(4.66, 4.66, 4.66); Calibrated: 3/19/2014; Sensor-Surface: 3mm (Mechanical Surface Detection (Locations From Previous Scan Used)), Sensor-Surface: 3mm (Mechanical Surface Detection), z = 2.0, 32.0 Electronics: DAE4 Sn1265; Calibrated: 1/29/2014 Phantom: SAM; Type: QD 000 P40 CC; Serial: 1592 DASY52 52.8.1(838); Flat-Section_9-17-2014/Leather_Back 0mm/Area Scan (11x14x1): Measurement grid: dx=12mm, dy=12mm Maximum value of SAR (measured) = 0.0339 mW/g Flat-Section_9-17-2014/Leather_Back 0mm/Zoom Scan (5x5x7)/Cube 0: Measurement grid: dx=8mm, dy=8mm, dz=5mm Reference Value = 3.357 V/m; Power Drift = -0.14 dB Peak SAR (extrapolated) = 0.046 mW/g SAR(1 g) = 0.029 mW/g; SAR(10 g) = 0.018 mW/g Maximum value of SAR (measured) = 0.0342 mW/g 0 dB = 0.0339 mW/g = -29.40 dB mW/g FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix A - Plots Page 5 of 18 Plot 5 Date/Time: 9/17/2014 2:52:33 PM Test Laboratory: Cetecom Inc. SAR 1 Lab DUT: TZ Medical; Type: ECG Monitor; Serial: IMEI:35150205.018524.809 Communication …
Text truncated - open the document above for the full version.
FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 1 of 33 Tissue Parameters Recipe for liquids below 1 GHz: Water 35-58% Sugar 40-60% Salt 0-6% Hydroxyethyl-cellulose <0.3% Preventol-D7 0.1-0.7% Recipe for liquids above 1-3 GHz: Water 52-75% DGBE 25-48% Salt <1.0% SAR measurements were made within 24 hours of the measurement of liquid parameters. Relative permittivity and conductivity are within ±5% of the target. FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 2 of 33 850 MHz Body Liquid Date Temp (°C) Frequency (MHz) Relative Permativity Conductivity (S/m) 82453.57540.9551 82553.53940.9576 82653.53330.9596 82753.50760.9623 83553.35530.9826 83653.35590.9857 83753.35230.9885 84653.33131.0114 84753.33441.0138 84853.34711.0161 84953.35321.0181 82455.29020.9832 82555.27010.9851 82655.26010.9868 82755.2410.9894 83555.12831.0058 83655.11681.0068 83755.10061.0095 84655.08241.0281 84755.07451.0297 84855.08161.0318 84955.09141.0339 2014/09/1221.7 2014/09/1721 0.9 0.92 0.94 0.96 0.98 1 1.02 1.04 820840 Conductivity (S /m) Frequency (MHz) Conductivity Target Liquid V alue 5% Liquid Tolerance -5% Liquid Tolerance 2014/09/12 2014/09/17 51 52 53 54 55 56 57 58 59 820840 Relative Permittivity Frequency (MHz) Permittivity Target Liquid Value 5% Liquid Tolerance -5% Liquid Tolerance 2014/09/12 2014/09/17 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 3 of 33 1900 MHz Body Liquid Date Temp (°C) Frequency (MHz) Relative Permativity Conductivity (S/m) 185052.3321.5177 185152.33061.5188 188052.21321.5554 190052.13951.5771 190752.10941.5841 190852.10761.5851 190952.10491.586 191052.10491.586 185051.21761.4855 185151.21751.4879 188051.28261.5473 190051.34181.5568 190751.30791.5569 190851.30311.5568 190951.30211.5559 191051.28521.5559 2014/09/1721.1 9/30/201421.8 1.42 1.44 1.46 1.48 1.5 1.52 1.54 1.56 1.58 1.6 18451895 Conductivity (S /m) Frequency (MHz) Conductivity Target Liquid V alue 5% Liquid Tolerance -5% Liquid Tolerance 2014/09/17 50 51 52 53 54 55 56 18451895 Relative Permittivity Frequency (MHz) Permittivity Target Liquid V alue 5% Liquid Tolerance -5% Liquid Tolerance 2014/09/17 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 4 of 33 Test Equipment SAR1 Lab Equipment Model Serial Instrument description Supplier / Manufacturer Model Serial No. Calibration (date) Calibration Due (date) Robot Staubli TX90 F10/5D3NA 1/A/01 N/A N/A SAM Twin Phantom SPEAG SM 000 T01 DA 1592 N/A N/A Elliptical Phantom SPEAG QD OVA 001 BB 1092 N/A N/A Software SPEAG Dasy52.6.2.482 N/A N/A N/A Device Holder SPEAG SD 000H01 N/A N/A N/A Data Acquisition Electronics SPEAG DAE4 1265 2014/01/29 2015/01/29 SAR Probe SPEAG ES3DV3 3244 2014/03/19 2015/03/19 Shared Equipment Instrument description Supplier / Manufacturer Model Serial No. Calibration (date) Calibration Due (date) 900 MHz Body Tissue Simulant SPEAG MSL 900 Batch 100818-1 2014/09/12 – 2014/09-17 N/A 1900 MHz Body Tissue Simulant SPEAG MSL 1900 Batch 100824-3 2014/09/17 – 2014/09/30 N/A 835 MHz Dipole SPEAG D835V2 4d113 2014/04/07 2015/04/07 1900 MHz Dipole SPEAG D1900V2 5d135 2014/04/09 2015/04/09 Network Analyzer Agilent FieldFox N9923A MY51491621 2013/06/21 2015/06/21 Directional coupler Werlatone C6529 11249 N/A N/A RF Amplifier Vectawave VTL5400 N/A N/A N/A Dielectric Measurement Kit SPEAG DAK-3.5 1118 2014/04/08 2015/04/08 Synthesized CW Generator Agilent 8371213 US37101255 N/A N/A Power Meter Agilent E4419B MY45101996 2013/06/03 2015/06/03 Power Sensor Agilent E9300A MY41498484 2013/06/04 2015/06/04 Power Sensor Agilent E9300A MY41498492 2013/06/04 2015/06/04 Radio Communications Tester Rohde & Schwarz CMU 200 101821 2013/06 2015/06 Radio Communications Tester Rohde & Schwarz CMU 200 109879 2013/06 2015/06 Radio Communications Tester Rohde & Schwarz CMU 200 110759 2013/06 2015/06 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 5 of 33 Equipment Calibration/Performance Documents: Attached: SAR Probe ES3DV3 Calibration Report 835 MHz Dipole Calibration Report 1900 MHz Dipole Calibration Report FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 6 of 33 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 7 of 33 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 8 of 33 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 9 of 33 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 10 of 33 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 11 of 33 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 12 of 33 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 13 of 33 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 14 of 33 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 15 of 33 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 16 of 33 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 17 of 33 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 18 of 33 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 19 of 33 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 20 of 33 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 21 of 33 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 22 of 33 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 23 of 33 FCC ID: ZIMTZMR3G IC Cert. No.: 9647A-TZMR3G SAR Test Report - Appendix C - Misc Page 24 of 33 FCC ID: ZIMTZMR3G IC …
Text truncated - open the document above for the full version.
FCC: Accredited IC recognized # 3462B-1 CETECOM Inc. 411 Dixon Landing Road Milpitas, CA 95035 U.S.A. Phone: + 1 (408) 586 6200 Fax: + 1 (408) 586 6299 E-mail: [email protected] http://www.cetecom.com CETECOM Inc. is a Delaware Corporation with Corporation number: 2905571 V1.01 2013-07-29 © Copyright by CETECOM FCC/IC Test Report FOR: TZ Medical, Inc. Model: TZMR3G Product Description: The Aera CT is a battery operated ambulatory electrocardiograph recorder for use in mobile cardiovascular telemetry. FCC ID: ZIMTZMR3…
Text truncated - open the document above for the full version.
| # | Rule Parts | Frequency Range | Power Output | Emission | Tolerance |
|---|---|---|---|---|---|
| 6 | 24E | 1.85 GHz - 1.91 GHz | 603.00 mW | 4M17F9W | 2.5 ppm |

Ambulatory electrocardiograph monitor with onboard ECG analysis and wireless connection via Wi-Fi and BLE
Equipment Class
DTS - Digital Transmission System
AERA CT MOBILE CARDIAC MONITOR
Equipment Class
PCT - PCS Licensed Transmitter worn on body