Text extracted from the exhibit documents filed with the FCC. Open a document above to read the original.
Your Custom Non-Rechargeable Hearing Aid User Manual Non-Rechargeable Products 2.4 GHz CIC (Completely-In-Canal) 312 2 | Warnings, Cautions, & Notices3 | Warnings, Cautions, & Notices REQUIRED HEARING AID INFORMATION The following additional information is provided in compliance with U.S. Food and Drug Administration (FDA) regulations: WARNING: People younger than 18 should go to a doctor before using this. People younger than 18 years old need specialized care, and using this without a medical evaluation may worsen impairment or disability. A hearing aid user who is younger than 18 should have a recent medical evaluation from a doctor, preferably an ear-nose- throat doctor (an ENT). Before using this, a doctor should determine that the use of a hearing aid is appropriate. WARNING to Hearing Aid Dispensers: You should advise a prospective hearing aid user to consult promptly with a doctor, preferably an ear specialist such as an ENT, before dispensing a hearing aid if you determine through inquiry, actual observation, or review of any other available information concerning the prospective user, that the prospective user has any of the following conditions: • Visible deformity of the ear, either congenital or traumatic • Fluid, pus, or blood coming out of the ear within the previous 6 months • Pain or discomfort in the ear • History of excessive ear wax or suspicion that something is in the ear canal • Dizziness, either recent or long-standing • Sudden, quickly worsening, or fluctuating hearing loss within the previous 6 months • Hearing loss or ringing (tinnitus) only in one ear or a noticeable difference in hearing between ears • Audiometric air-bone gap equal to or greater than 15 dB at 500 Hz, 1000 Hz, and 2000 Hz WARNING to Hearing Aid Dispenser, Outputs over 132 dB SPL: You should exercise special care in selecting and fitting a hearing aid with a maximum output that exceeds 132 dB SPL because it may impair the remaining hearing of the hearing aid user. International Use Your hearing aids are approved to operate at a radio frequency that is specific to your country or region and might not be approved for use outside your country or region. Be aware that operation during international travel may cause interference to other electronic instruments, or other electronic instruments may cause interference to your hearing aids. Warnings, Cautions, & Notices We are required by regulations to provide the following warnings: WARNING: Use of wireless hearing aids directly next to other electronic equipment should be avoided because it could result in improper performance. If such use is necessary, note as to whether your hearing aids and the other equipment are operating normally. WARNING: Use of accessories, components or replacement parts other than those provided by the manufacturer of your hearing aids could result in increased electromagnetic emissions and decreased electromagnetic immunity and could result in degradation of performance. WARNING: If Portable Radio Frequency communications equipment is used closer than 30 cm (12 inches) from your hearing aid, degradation of the performance of your hearing aid could result. If this occurs, move away from the communications equipment. WARNING: The stability of the hearing aid is designed for normal use. Avoid physical impacts to the ear when wearing a hearing aid that may cause the device to break or component of the device to become detached. This may lead to lacerations of the ear canal or perforation of the ear drum. If this occurs, it is strongly recommended to see a physician for evaluation and safe removal. WARNING: If the hearing aid drops onto a hard surface, it could cause the hearing aid to break or become damaged. This includes mechanical stress or shock of the device. Ensure the hearing aid is intact before placing it in the ear. If you find the device is damaged, discontinue use and see your hearing professional. WARNING: Repeated stress to the removal handle may cause it to break. If the removal handle breaks and you’re unable to remove the device carefully, it is recommended to see a physician for safe removal. WARNING TO HEARING CARE PROFESSIONAL A hearing care professional should advise a prospective sound generator (tinnitus masker) user to consult promptly with a licensed physician (preferably an ear specialist) before using a sound generator if the hearing care professional determines through inquiry, actual observation or review or any other available information concerning the prospective user that the prospective user has any of the following conditions: • Visible congenital or traumatic deformity of the ear • History of active drainage from the ear within the previous 90 days • History of sudden or rapidly progressive hearing loss within the previous 90 days • Acute or chronic dizziness • Unilateral hearing loss of sudden or recent onset within the previous 90 days Warnings, Cautions, & Notices 4 | Warnings, Cautions, & Notices5 | Warnings, Cautions, & Notices Warnings, Cautions, & Notices Warnings, Cautions, & Notices WARNING: There are some potential concerns associated with the use of any sound generating tinnitus therapy instrument. Among them are the potential for worsening of tinnitus, a possible change in hearing thresholds, and possible skin irritation at the point of contact with the hearing aid. Multiflex Tinnitus Technology has been designed to minimize these concerns. However, should you experience or notice any of the above conditions or any dizziness, nausea, headaches or heart palpitations, you should immediately discontinue use of the hearing aid and seek a consultation with a medical, audiology or other hearing care professional. As with any hearing aid, misuse of the tinnitus therapy instrument could present some potentially harmful effects. Care should be taken to prevent the unauthorized use and to keep the hearing aid out of the reach of children and pets. FALL ALERT FEATURE The Fall Alert feature…
Text truncated - open the document above for the full version.
Starkey | 6700 Washington Avenue S Eden Prairie, Minnesota 55344 T: 1.800.328.8602 Hear better. Live better. Starkey.com March 30, 2023 Federal Communications Commission Authorization and Evaluation Division 7435 Oakland Mills Road Columbia, Maryland 21046 Re: section 2.911(d)(7) filing To whom it may concern, This letter is to confirm that Starkey Laboratories, Inc. has accepted responsibility to act as the agent for service of process as acknowledged by the applicant, Starkey Laboratories, Inc. William J. Mitchell confirms he is not associated with or part of any TCB or test lab. U.S. Agent: William J. Mitchell Title: Principal Wireless and Safety Regulatory Engineer Company Name: Starkey Laboratories, Inc. Address: | 6600 Washington Avenue, South, Eden Prairie, MN 55344 Telephone No: | 952-918-5683 Email: | [email protected] FRN (FCC Registration | 0015964463 Number): The applicant accepts and acknowledges the obligation to maintain an agent for no less than one year after the grantee has terminated all marketing and importation or the conclusion of any Commission-related proceeding involving the equipment. Signature of Applicant (must be signed by the contact of record in the FCC database for the applicant grantee code): luc J Duthil | Feb Signature of U.S. Agent (if different from applicant): Dp bet! Name: William J. Mitchell Name: Date: March 30, 2023 Date: Starkey logo and Hear better. Live better. are registered trademarks of Starkey Laboratories, Inc.
Sta rkey 6700 Washington Avenue S Eden Prairie, Minnesota 55344 T: 1.800.328.8602 Hear better. Live better. Starkey.com April 13, 2023 Federal Communications Commission Authorization and Evaluation Division 7435 Oakland Mills Road Columbia, Maryland 21046 Re: Attestation regarding ‘covered’ equipment and the Covered List FCC ID: EOA-24GENCICZN To whom it may concern, Per 2.911(d)(5)(i), Starkey Laboratories, Inc. certifies that the equipment for which authorization is sought is not “covered” equipment prohibited from receiving an equipment authorization pursuant to section 2.903 of the FCC rules. Per 2.911(d)(5) (ii), Starkey Laboratories, Inc. certifies that, as of the date of the filing of the application, the applicant is not identified on the Covered List as an entity producing “covered” equipment. Yi William J. Bie fh G Principal Wireless and Safety Regulatory Engineer 952-918-5683 [email protected] Starkey logo and Hear better. Live better. are registered trademarks of Starkey Laboratories, Inc.
Sta rkey 6700 Washington Avenue S Eden Prairie, Minnesota 55344 T: 1.800.328.8602 Hear better. Live better. Starkey.com April 13, 2023 Federal Communications Commission 45 L Street, NE Washington, DC 20554 Attention: Application Examiner Reference: Certification: | Part 15.247 Intentional Radiator Applicant: Starkey Laboratories, Inc. Equipment: Low Power Device (2400-2483.5 MHz) Model #: Genesis AI Custom CIC Hearing Aid FCC ID: EOA-24GENCICZN Dear Sir/Madam: Enclosed please find an FCC EOA-24GENCICZN certification application. The device is a wireless hearing aid intended for use by the hearing impaired. The wireless features are used during the initial fitting of a hearing aid to program the hearing aid and used during normal usage to communicate control information and digital audio to and from any properly configured Bluetooth Low Energy compliant accessory and between a binaural pair of Genesis hearing aids. Further details and supporting documents are attached. Should you have any questions, please feel free to contact the undersigned. Sincerely, i 7] LL, William J. Mitchell/PE Principal Wireless and Safety Regulatory Engineer 952-918-5683 [email protected] Starkey logo and Hear better. Live better. are registered trademarks of Starkey Laboratories, Inc.
Sta rkey 6700 Washington Avenue S Eden Prairie, Minnesota 55344 T: 1.800.328.8602 Hear better. Live better. Starkey.com April 13, 2023 Federal Communications Commission Authorization and Evaluation Division 7435 Oakland Mills Road Columbia, Maryland 21046 Re: Request for confidentiality Applicant: Starkey Laboratories, Inc. FCC ID: EOA-24GENCICZN To whom it may concern: Permanent Confidentiality Request is hereby submitted by Starkey Laboratories, Inc. to withhold permanently from public review certain portions of the application for equipment certification for the referenced FCC identifiers. This request for confidentiality is made pursuant to 47 CFR 0.457(d) and 0.459 of the FCC Rules. In particular, the following sections of the application are to be kept permanently confidential: e Schematics e Detailed Block diagrams e Detailed Operational/Functional Description Short-term Confidentiality Request is hereby submitted by Starkey Laboratories, Inc. to withhold from public review for a period of 180 days from the date of the Grant of Equipment Authorization and prior to marketing, certain portions of the application for equipment certification for the referenced FCC identifiers. This request for confidentiality is made pursuant to 47 CFR 0.457(d) and 0.459 of the FCC Rules. In particular, the following sections of the application are to be kept confidential for a period of 180 days from the date of Authorization: e Test setup photos e External photos e Internal photos e User manual Rationale for request for confidentiality: Starkey Laboratories, Inc. has invested considerable time and materials in research and development to produce the referenced product: Disclosure of the permanently confidential portions of this application to competitors would not only give them significant competitive advantages in developing similar products, but would also disclose successful implementation of unpublished, leading edge technology developed by us. Disclosure of the short-term confidential portions of this application during the period of importation and/or distribution would reveal key aspects of proprietary technology to competitors and diminish the value of our investment in research and development. Sincerely, Y l K 2 ) . h . l i y William J. Mitchell, PE Principal Wireless and Safety Regulatory Engineer 952-918-5683 [email protected] Starkey logo and Hear better. Live better. are registered trademarks of Starkey Laboratories, Inc.
AGXs Genesis AI Custom wireless CIC Hearing Aid (scale in cm) Top view Side 1 Side 2 Side 3 Side 4 Bottom view
Intrigue AI Custom wireless CIC Hearing Aid (scale in cm) Top view Side 1 Side 2 Side 3 Side 4 Bottom view
MicroTech MT NXG AI Custom wireless CIC Hearing Aid (scale in cm) Top view Side 1 Side 2 Side 3 Side 4 Bottom view
NuEar NE NXG AI Custom wireless CIC Hearing Aid (scale in cm) Top view Side 1 Side 2 Side 3 Side 4 Bottom view
Genesis AI Custom wireless CIC Hearing Aid (scale in cm) Top view Side 1 Side 2 Side 3 Side 4 Bottom view
Sta rkey 6700 Washington Avenue S Eden Prairie, Minnesota 55344 T: 1.800.328.8602 Hear better. Live better. Starkey.com April 13, 2023 Federal Communications Commission 45 L Street, NE Washington, DC 20554 Subject: FCC ID labeling on EOA-24GENCICZN wireless hearing aids Re: Response to Inquiry to FCC (Tracking Number 775017) To whom it may concern, Consistent to the direction provided on a previous similar product via your July 12, 2010 Response to Inquiry to FCC (Tracking Number 775017) concerning approval by the FCC for alternate labeling to our hearing aid under CFR 2.925(f), we are placing the FCC ID label for the EOA-24GENCICZN wireless hearing aids in the User Manual. Consistent to the direction provided on the earlier product, we are including the following documents with our submission for certification of Starkey’s wireless hearing aids having an FCC ID of EOA-24GENCICZN. ° The FCC Response to Inquiry to FCC (Tracking Number 775017) ° Supporting documents similar to those associated with that Inquiry ° Figure 1 below showing curved nature, relative size, and available printed area and FCC ID of EOA-24GENCICZN hearing aids. ° The relevant page from the User Manual showing the alternate FCC ID marking location and other Part 15 required wording. If you have any questions or comments, please do not hesitate to contact me. Sincerely, L p l i . - } C p e t e ll William J. Mitchell, P Principal Wireless and Safety Regulatory Engineer 952-918-5683 [email protected] |GENCICZN Figure 1 (scale in inches and mm) Starkey logo and Hear better. Live better. are registered trademarks of Starkey Laboratories, Inc.
Genesis AI Custom wireless CIC Hearing Aid Right side (2 views) Left side End 1 view End 2 view Bottom view Top view PCB front side PCB back side
AGXs Genesis AI Custom wireless CIC Hearing Aid Right side (2 views) Left side End 1 view End 2 view Bottom view Top view PCB front side PCB back side
MicroTech MT NXG AI Custom wireless CIC Hearing Aid Right side (2 views) Left side End 1 view End 2 view Bottom view Top view PCB front side PCB back side
NuEar NE NXG AI Custom wireless CIC Hearing Aid Right side (2 views) Left side End 1 view End 2 view Bottom view Top view PCB front side PCB back side
Audibel Intrigue AI Custom wireless CIC Hearing Aid Right side (2 views) Left side End 1 view End 2 view Bottom view Top view PCB front side PCB back side
This report must not be used to claim product certification, approval, or endorsement by A2LA, NIST, or any agency of the U.S. Government. This Report shall not be reproduced, except in full without written approval of the laboratory. Starkey Laboratories, Inc. CIC in Genesis AI family FCC 2.1093:2023 Bluetooth Low Energy Radio Report: STAK0289.4, Issue Date: July 10, 2023 TABLE OF CONTENTS 2023.04.17 Section Page Number Certificate of Evaluation .................................................................................. 3 Revision History .............................................................................................. 4 Accreditations .................................................................................................. 5 Facilities .......................................................................................................... 6 Product Description ......................................................................................... 7 Exposure Condition ......................................................................................... 8 SAR Test Exclusion ........................................................................................ 9 End of Report .................................................................................................. 11 Report No. STAK0289.42/11 CERTIFICATE OF EVALUATION 2018-09-14 Product compliance is the responsibility of the client; therefore, the tests and equipment modes of operation represented in this report were agreed upon by the client, prior to testing. The results of this test pertain only to the sample(s) tested. The specific description is noted in each of the individual sections of the test report supporting this certificate of test. This report reflects only those tests from the referenced standards shown in the certificate of test. It does not include inspection or verification of labels, identification, marking or user information. As indicated in the Statement of Work sent with the quotation, Element’s standard process is to always use the latest published version of the test methods even when earlier versions are cited in the test specification. Issuance of a purchase order was de facto acceptance of this approach. Otherwise, the client would have advised Element in writing of the specific version of the test methods they wanted applied to the subject testing Last Date of Evaluation: July 3, 2023 Starkey Laboratories, Inc. EUT: CIC in Genesis AI family RF Exposure Evaluation Standards Specification Method FCC 2.1093:2023 FCC 447498 D01 General RF Exposure Guidance v06 Results Method Clause Description Applied Results Comments 4.3.1 SAR Test Exclusion Yes Pass None Deviations From Evaluation Standards None Approved By: Donald Facteau, Process Architect Report No. STAK0289.43/11 REVISION HISTORY 2018.07.17 Revision Number Description Date (yyyy-mm-dd) Page Number 00 None Report No. STAK0289.44/11 ACCREDITATIONS AND AUTHORIZATIONS 2021.09.01 United States FCC - Designated by the FCC as a Telecommunications Certification Body (TCB). Certification chambers, Open Area Test Sites, and conducted measurement facilities are listed with the FCC. A2LA - Each laboratory is accredited by A2LA to ISO / IEC 17025, and as a product certifier to ISO / IEC 17065 which allows Element to certify transmitters to FCC and IC specifications. Canada ISED - Recognized by Innovation, Science and Economic Development Canada as a Certification Body (CB) and as a CAB for the acceptance of test data. European Union European Commission – Recognized as an EU Notified Body validated for the EMCD and RED Directives. United Kingdom BEIS – Recognized by the UK as an Approved Body under the UK Radio Equipment and UK EMC Regulations. Australia/New Zealand ACMA - Recognized by ACMA as a CAB for the acceptance of test data. Korea MSIT / RRA - Recognized by KCC’s RRA as a CAB for the acceptance of test data. Japan VCCI - Associate Member of the VCCI. Conducted and radiated measurement facilities are registered. Taiwan BSMI – Recognized by BSMI as a CAB for the acceptance of test data. NCC - Recognized by NCC as a CAB for the acceptance of test data. Singapore IDA – Recognized by IDA as a CAB for the acceptance of test data. Israel MOC – Recognized by MOC as a CAB for the acceptance of test data. Hong Kong OFCA – Recognized by OFCA as a CAB for the acceptance of test data. Vietnam MIC – Recognized by MIC as a CAB for the acceptance of test data. SCOPE For details on the Scopes of our Accreditations, please visit: California Minnesota Oregon Texas Washington Report No. STAK0289.45/11 FACILITIES 2020.12.15 California Labs OC01-17 41 Tesla Irvine, CA 92618 (949) 861-8918 Minnesota Labs MN01-11 9349 W Broadway Ave. Brooklyn Park, MN 55445 (612) 638-5136 Oregon Labs EV01-12 6775 NE Evergreen Pkwy #400 Hillsboro, OR 97124 (503) 844-4066 Texas Labs TX01-09 3801 E Plano Pkwy Plano, TX 75074 (469) 304-5255 Washington Labs NC01-05 19201 120 th Ave NE Bothell, WA 98011 (425) 984-6600 A2LA Lab Code: 3310.04 Lab Code: 3310.05 Lab Code: 3310.02 Lab Code: 3310.03 Lab Code: 3310.06 Innovation, Science and Economic Development Canada 2834B-1, 2834B-3 2834E-1, 2834E-3 2834D-1 2834G-1 2834F-1 BSMI SL2-IN-E-1154R SL2-IN-E-1152R SL2-IN-E-1017 SL2-IN-E-1158R SL2-IN-E-1153R VCCI A-0029 A-0109 A-0108 A-0201 A-0110 Recognized Phase I CAB for ISED, ACMA, BSMI, IDA, KCC/RRA, MIC, MOC, NCC, OFCA US0158 US0175 US0017 US0191 US0157 Report No. STAK0289.46/11 PRODUCT DESCRIPTION 2018-07-17 Client and Equipment Under Evaluation Information Company Name: Starkey Laboratories, Inc. Address: 6600 Washington Ave S City, State, Zip: Eden Prairie, MN 55344-3404 Evaluation Requested By: Bill Mitchell EUT: CIC in Genesis AI family Date of Evaluation: 7/3/2023 Information Provided by the Party Requesting the Evaluation Functional Description of the Equipment: CIC hearing aid containing Bluetooth LE radio (FCC ID: EOA-24GENCICZN). Objective: To demonstrate co…
Text truncated - open the document above for the full version.
Genesis AI Custom wireless CIC Antenna Description The Bluetooth 2.4 GHz antenna is a center fed monopole. The peak gain of the antenna in the assembled DUT is -4 dBi. Figure 1 2.4 GHz Antenna (scale in mm) Antenna shown on faceplate before hearing aid is removed from excess plastic around faceplate and custom case molded to patient ear canal is placed on it. Half of antenna shown above faceplate in photo, other half is wire below faceplate. Date of antenna pattern measurement: April 18, 2023 Figure 2 Hearing Aid Antenna Pattern Figure 3 Hearing Aid Elevation cut @ 0 degrees from X axis. Scale in dBm Figure 4 Hearing Aid Antenna Azimuth Cut @ 90 degrees from Z axis. Scale in dBm. Antenna Pattern Measurement Information The antenna pattern shown in Figure 2 was measured using a MVG SGL24L antenna test system, serial number ATL0232S located at Starkey Laboratories, Inc., 6600 Washington Avenue, South, Eden Prairie, MN 55344 System was calibrated on September 9, 2021 and September 16, 2022, due for calibration in September 2023. Signal levels were measured using an Agilent N9020A MXA Signal Analyzer (Spectrum Analyzer). serial number MY50410289, calibrated on July 19, 2021 and October 26, 2022, due for calibration on October 31, 2024. The antenna pattern plot in Figure 2 is generated by the SG24L test system software. Figure 5a Overall view of SG24L test chamber, showing ring of receiving antennas Figure 5b Test stand in SG24L test chamber Figure 5c close-up of unit under test in test chamber Antenna Gain Measurement Information The MVG SGL24L antenna test system runs internal scripts that yield the maximum EIRP from each radiated power measurement. From there, the following equation could be used to calculate the antenna gain in dBi. Max Antenna Gain = Max EIRP – Power at antenna pads Where, Power at antenna pads = BLE Chipset Power Setting – Simulated PCB Insertion Loss Subtracting the conducted power at the antenna pads from the EIRP value, yields the antenna gain as follows: • Max Antenna Gain = -4.0 dBi
This report must not be used to claim product certification, approval, or endorsement by A2LA or any agency of the U.S. Government. This Report shall not be reproduced, except in full without written approval of the laboratory. Starkey Laboratories, Inc. CIC in Genesis AI family FCC 15.247:2023 RSS-247 Issue 2:2017 Bluetooth Radio Report: STAK0289.1 Rev. 1, Issue Date: July 11, 2023 TABLE OF CONTENTS 2023.04.17 Section Page Number Cover Page ..................................................................................................... 1 Certificate of Test ............................................................................................ 2 Revision History .............................................................................................. 3 Accreditations .................................................................................................. 4 Facilities .......................................................................................................... 5 Measurement Uncertainty ............................................................................... 6 Test Setup Block Diagrams ............................................................................. 7 Product Description ......................................................................................... 10 Power Settings and Antenna Information ........................................................ 11 Configurations ................................................................................................. 12 Modifications ................................................................................................... 13 Duty Cycle ....................................................................................................... 14 Duty Cycle (SRE Test Mode) .......................................................................... 19 DTS Bandwidth (6 dB) .................................................................................... 24 Occupied Bandwidth (99%) ............................................................................. 29 Output Power .................................................................................................. 34 Equivalent Isotropic Radiated Power .............................................................. 39 Power Spectral Density ................................................................................... 44 Band Edge Compliance .................................................................................. 49 Spurious Conducted Emissions ...................................................................... 53 Spurious Radiated Emissions ......................................................................... 64 End of Report .................................................................................................. 71 Report No. STAK0289.1 Rev 12/72 CERTIFICATE OF TEST 2022.04.01 Product compliance is the responsibility of the client; therefore, the tests and equipment modes of operation represented in this report were agreed upon by the client, prior to testing. The results of this test pertain only to the sample(s) tested. The specific description is noted in each of the individual sections of the test report supporting this certificate of test. This report reflects only those tests from the referenced standards shown in the certificate of test. It does not include inspection or verification of labels, identification, marking or user information. As indicated in the Statement of Work sent with the quotation, Element’s standard process is to always use the latest published version of the test methods even when earlier versions are cited in the test specification. Issuance of a purchase order was de facto acceptance of this approach. Otherwise, the client would have advised Element in writing of the specific version of the test methods they wanted applied to the subject testing. Last Date of Test: April 20, 2023 Starkey Laboratories, Inc. EUT: CIC in Genesis AI family Radio Equipment Testing Standards Specification Method FCC 15.247:2023 ANSI C63.10:2013, FCC KDB 558074 v05r02:2019 RSS-247 Issue 2:2017 RSS-Gen Issue 5:2018+A1:2019+A2:2021 Results Test Description Result FCC Section(s) RSS Section(s) ANSI C63.10 Section(s) Comments Powerline Conducted Emissions N/A 15.207 RSS-Gen 8.8 6.2 Not required for battery powered device Duty Cycle Pass KDB 558074 -6.0 RSS-Gen 3.2 11.6 DTS Bandwidth (6 dB) Pass 15.247(a)(2), KDB 558074 -8.2 RSS-247 5.2(a) 11.8.2 Occupied Bandwidth (99%) Pass KDB 558074 -2.1 RSS-Gen 6.7 6.9.3 Output Power Pass 15.247(b)(3), KDB 558074 - 8.3.1 RSS-247 5.4(d, f), RSS-Gen 6.12 11.9.1.1 Equivalent Isotropic Radiated Power Pass 15.247(b)(3), KDB 558074 - 8.3.1 RSS-247 5.4(d, f), RSS-Gen 6.12 11.9.1.1 Power Spectral Density Pass 15.247(e), KDB 558074 -8.4 RSS-247 5.2(b) 11.10.2 Band Edge Compliance Pass 15.247(d), KDB 558074 -8.5 RSS-247 5.5 11.11 Spurious Conducted Emissions Pass 15.247(d), KDB 558074 -8.5 RSS-247 5.5 11.11 Spurious Radiated Emissions Pass 15.247(d), KDB 558074 - 8.6, 8.7 RSS-247 5.5, RSS-Gen 6.13, 8.10 11.12.1, 11.13.2, 6.5, 6.6 Deviations From Test Standards None Approved By: Eric Brandon, Department Manager Report No. STAK0289.1 Rev 13/72 REVISION HISTORY 2018.07.17 Revision Number Description Date (yyyy-mm-dd) Page Number 00 None 01 Added Contents, updated CoT, corrected OBW 2023-07-11 All Report No. STAK0289.1 Rev 14/72 ACCREDITATIONS AND AUTHORIZATIONS 2021.09.01 United States FCC - Designated by the FCC as a Telecommunications Certification Body (TCB). Certification chambers, Open Area Test Sites, and conducted measurement facilities are listed with the FCC. A2LA - Each laboratory is accredited by A2LA to ISO / IEC 17025, and as a product certifier to ISO / IEC 17065 which allows Element to certify transmitters to FCC and IC specifications. Canada ISED - Recognized by Innovation, Science and Economic Development Canada as a Certification Body (CB) an…
Text truncated - open the document above for the full version.
XMit 2023.02.14.0 LAB 001.JPG LAB 002.JPG DIRECT CONNECT SPURIOUS RADIATED EMISISONS WTD.2023.02.14.0 EmiR5 2022.07.06.0 PSA-ESCI 2023.02.14.0 SPURIOUS RADIATED EMISISONS WTD.2023.02.14.0 EmiR5 2022.07.06.0 PSA-ESCI 2023.02.14.0 SPURIOUS RADIATED EMISISONS WTD.2023.02.14.0 EmiR5 2022.07.06.0 PSA-ESCI 2023.02.14.0
| # | Rule Parts | Frequency Range | Power Output |
|---|---|---|---|
| 1 | 15C | 2.40 GHz - 2.48 GHz | 3.00 mW |

Starlink Edge TV
Equipment Class
DTS - Digital Transmission System
Edge AI ITE right
Equipment Class
DTS - Digital Transmission System
Edge AI CIC
Equipment Class
DTS - Digital Transmission System
Edge AI mRIC R
Equipment Class
DTS - Digital Transmission System
Edge AI ITE left
Equipment Class
DTS - Digital Transmission System