PRODUCTSPECIFICATIONOF BLE-PATIENTWRISTBANDWITHCALL BUTTON No.4/363,Block-3,4 th floor,OMRRoad, Kandanchavadi,Chennai-600096,India. www.trackerwave.com DocumentTitleBLE-PATIENTWRISTBANDWITHCALL BUTTON DocumentNumber VersionNumber1.0 VersionDate17.01.2020 PreparedByBalamurugan.KP ApprovedbyPradeep.V copyright©2019TrackerwavePvt.Ltd. ProductSpecificationV1 TrackerwaveTagi CONTENTS 1.PRODUCTOVERVIEW ......................................................................................................... 1 1.1.Introduction ........................................................................................................... 1 2.SPECIFICATIONS .................................................................................................................. 1 2.1.KeyFeatures ........................................................................................................... 1 2.2.Applications ............................................................................................................ 1 2.3.Parameters ............................................................................................................. 1 2.4.OperatingConditions ............................................................................................. 1 REVISIONHISTORY ........................................................................................................................ 2 ProductSpecificationV1 TrackerwaveTag 1 1.PRODUCTOVERVIEW 1.1.Introduction TrackerwaveOvitagWristbandisaBluetoothLowEnergy5.0baseddeviceforindoortracking.The Wristbandsupportspatientfalldetectionandactivitymonitoring.WristbandButtonhelpspatient callnurse.Wristbandispoweredbysmallsizecoinbatteryanditsreplaceable.Itisdesignedfor healthcaretowithstandE20,GammaandPlasmaserialization. 2.SPECIFICATIONS 2.1.KeyFeatures Bluetooth5.0 SOSbutton IndicationLED’s Replaceablecoinbattery Accelerometer Applicationprocessor o 64MHzArm®Cortex®-M33CPU 2.2.Applications IndoorLocationtracking Patientfalldetection Activitymonitoring Nursecall Geofencing Queuemanagement Nurseroutine 2.3.Parameters Operationvoltage:1.8-3.6V Transmissionrange:Maximum50meters Batterylife:1Year 2.4.OperatingConditions Temperature:-20to+75°C Humidity:80% Dimension:3.8X3.3X1cm Weight:10grams ProductSpecificationV1 TrackerwaveTag 2 REVISIONHISTORY S.NoVer.No.Ver.DateChangedBySection(s) Changed Descriptionofchange 1.1.001.17.2020BalamuruganKP-InitialDraft FCCRequirement changesormodificationsnotexpresslyapprovedbythepartyresponsibleforcompliancecould voidtheuser’sauthoritytooperatetheequipment. ThisdevicecomplieswithPart15oftheFCCRules.Operationissubjecttothefollowingtwoconditions: (1)thisdevicemaynotcauseharmfulinterference,and (2)thisdevicemustacceptanyinterferencereceived,includinginterferencethatmaycauseundesired operation. Note:ThisequipmenthasbeentestedandfoundtocomplywiththelimitsforaClassBdigitaldevice, pursuanttoPart15oftheFCCRules.Theselimitsaredesignedtoprovidereasonableprotectionagainst harmfulinterferenceinaresidentialinstallation.Thisequipmentgenerates,uses,andcanradiateradio frequencyenergy,andifnotinstalledandusedinaccordancewiththeinstructions,maycauseharmful interferencetoradiocommunications.However,thereisnoguaranteethatinterferencewillnotoccurina particularinstallation.Ifthisequipmentdoescauseharmfulinterferencetoradioortelevisionreception, whichcanbedeterminedbyturningtheequipmentoffandon,theuserisencouragedtotrytocorrectthe interferencebyoneormoreofthefollowingmeasures: –Reorientorrelocatethereceivingantenna. –Increasetheseparationbetweentheequipmentandreceiver. –Connecttheequipmentintoanoutletonacircuitdifferentfromthattowhichthereceiverisconnected. –Consultthedealeroranexperiencedradio/TVtechnicianforhelp.
Date:04/17/2018VCB_F-01:OperatingProceduresForm–FCCConfidentialityv02Page1of1 TRACKERWAVEPRIVATELlMITED FederalCommunicationsCommission AuthorizationandEvaluationDivision ConfidentialityRequestregardingapplicationforcertificationofFCCID:2ATS5-TWBL2 PursuanttoSections0.457and0.459oftheCommission’sRules,weherebyrequest confidentialtreatmentofinformationaccompanyingthisapplicationasoutlinedbelow: ExhibitTypeFileName 1.BlockDiagram, 2.Schematics, 3.OperationalDescription 1.2ATS5-TWBL2-Blockdiagram 2.2ATS5-TWBL2-SCH 3.2ATS5-TWBL2-Operationaldescription Theabovematerialscontaintradesecretsandproprietaryinformationnotcustomarily releasedtothepublic.Thepublicdisclosureofthesematerialsmaybeharmfultothe applicantandprovideunjustifiedbenefitstoitscompetitors. TheapplicantunderstandsthatpursuanttoSection0.457oftheRules,disclosureofthis applicationandallaccompanyingdocumentationwillnotbemadebeforethedateofthe Grantforthisapplication. PursuanttoDA04-1705June15,2004oftheCommission’spublicnotice,wealso requesttemporaryconfidentialtreatmentofinformationaccompanyingthisapplication asoutlinedbelowforaninitialperiodofxxxx(ie-“45”,“90”or“180”)days: ExhibitTypeFileName N/AN/A Signature: Contactname:PradeepV. Companyname:TRACKERWAVEPRIVATELlMITED Address: No.4,363,BIockC,3rdFloor,OMRRoad,Kanthanchavadi,Chennai,India600096
FCC ID:2ATS5-TWBL2 FCC ID:2ATS5-TWBL2 FCC ID:2ATS5-TWBL2
FCC ID:2ATS5-TWBL2 FCC ID:2ATS5-TWBL2 FCC ID:2ATS5-TWBL2
Shenzhen Huaxia Testing Technology Co., Ltd 1F., Block A of Tongsheng Technology Building,Huahui Road, Dalang Street, Longhua District, Shenzhen, China Telephone: +86-755-26648640 Fax:+86-755-26648637 Website:www.cqa-cert.com Report Template Version: V04 Repor tTemplate Revision Date: 2018-07-06 RF Exposure Evaluation Report The test report is effective only with both signature and specialized stamp, The result(s) shown in this report refer only to the sample(s) tested. Without written approval of CQA, this report can’t be reproduced except in full. Report No. : CQASZ20200200074E-02 Applicant: Trackerwave Private Limited Address of Applicant:4/363, Block C, 3rd Floor, OMR Road, Kanthanchavadi, Chennai, India – 600 096 Equipment Under Test (EUT): EUT Name: BLE - Patient Wrist Band with Call Button Mode No.: TWBL2 Brand Name: OVITAG FCC ID: 2ATS5-TWBL2 Standards: 47 CFR Part 1.1307 47 CFR Part 2.1093 KDB447498D01 General RF Exposure Guidance v06 Date of Receipt:2020-02-21 Date of Test:2020-02-21 to 2020-02-25 Date of Issue:2020-02-25 Test Result :PASS* * In the configuration tested, the EUT complied with the standards specified above. Tested By: (Tom Chen) Reviewed By: (Aaron Ma) Approved By: (Jack Ai) Shenzhen Huaxia Testing Technology Co., Ltd Report No.: CQASZ20200200074E-02 Page: 2 of 6 1Version Revision History Of Report ReportNo.Version Description Issue Date CQASZ20200200074E- 02 Rev.01 Initial report 2020-02-25 Shenzhen Huaxia Testing Technology Co., Ltd Report No.: CQASZ20200200074E-02 Page: 3 of 6 2 Contents Page 1 VERSION......................................................................................................................................................................2 2CONTENTS..................................................................................................................................................................3 3 GENERAL INFORMATION........................................................................................................................................4 3.1 C LIENTINFORMATION................................................................................................................................................4 3.2 G ENERALDESCRIPTION OFEUT................................................................................................................................4 4 SAR EVALUATION.....................................................................................................................................................5 4.1 RF E XPOSURECOMPLIANCEREQUIREMENT..............................................................................................................5 4.1.1 Standard Requirement..................................................................................................................................5 4.1.2 Limits................................................................................................................................................................5 4.1.3 EUT RF Exposure..........................................................................................................................................6 Shenzhen Huaxia Testing Technology Co., Ltd Report No.: CQASZ20200200074E-02 Page: 4 of 6 3 General Information 3.1 Client Information Applicant: Trackerwave Private Limited Address of Applicant: 4/363, Block C, 3rd Floor, OMR Road, Kanthanchavadi, Chennai, India – 600 096 Manufacturer: Shenzhen Minew Technolo gies Co., Ltd Address of Manufacturer: 3 rd Floor, Bldg I, Gangzhilong Science Park, Qinglong Rd., Longhua Dist., Shenzhen 518109, China 3.2 General Description of EUT Product Name: BLE - Patient Wrist Band with Call Button Model No.: TWBL2 Trade Mark: OVITAG Hardware Version: 1.2 Software Version: 1.0 Operation Frequency:2402MHz~2480MHz Bluetooth Version:V5.0 Modulation Type:GFSK Transfer Rate:1Mbps/2Mbps Number of Channel:40 Product Type:MobilePortableFix Location Test Software of EUT:nRFgo Studio (manufacturer declare ) Antenna Type: PCB antenna Antenna Gain: -7.6dBi EUT Power Supply: button battery:DC3.0V Shenzhen Huaxia Testing Technology Co., Ltd Report No.: CQASZ20200200074E-02 Page: 5 of 6 4 SAR Evaluation 4.1 RF Exposure Compliance Requirement 4.1.1 Standard Requirement According to KDB447498D01 General RF Exposure Guidance v06 4.3.1. Standalone SAR test exclusion considerations Unless specifically required by the published RF exposure KDB procedures, standalone 1-g head or body and 10-g extremity SAR evaluation for general population exposure conditions, by measurement or numerical simulation, is not required when the corresponding SAR Exclusion Threshold condition, listed below, is satisfied. 4.1.2 Limits The 1-g and 10-g SAR test exclusion thresholds for 100 MHz to 6 GHz at test separation distances ≤50 mm are determined by: [(max. power of channel, including tune-up tolerance, mW)/(min. test separation distance, mm)] · [√f(GHz)]≤3.0 for 1-g SAR and≤7.5 for 10-g extremity SAR, where f(GHz) is the RF channel transmit frequency in GHz Power and distance are rounded to the nearest mW and mm before calculation 17 The result is rounded to one decimal place for comparison The test exclusions are applicable only when the minimum test separation distance is≤50 mm and for transmission frequencies between 100 MHz and 6 GHz. When the minimum test separation distance is < 5 mm, a distance of 5 mm is applied to determine SAR test exclusion Shenzhen Huaxia Testing Technology Co., Ltd Report No.: CQASZ20200200074E-02 Page: 6 of 6 4.1.3EUT RF Exposure For BLE Measurement Data GFSK(1Mbps) mode Test channelPeak Output Power (dBm) Tune up tolerance (dBm) Maximum tune-up Power (dBm)(mW) Lowest(2402MHz)-0.32-1.0±101.000 Middle (2440MHz)-0.55-1.0±101.000 Highest(2480MHz)-0.42-1.0±101.000 GFSK(2Mbps) mode Test channelPeak Output Power (dBm) Tune up tolerance (dBm) Maximum tune-up Power (dBm)(mW) Lowest(2402MHz)-0.42-1.0±101.000 Middle (2440MHz)-0.65-1.0±101.000 H…
Text truncated - open the document above for the full version.
Shenzhen Huaxia Testing Technology Co., Ltd 1F., Block A of Tongsheng Technology Building,Huahui Road, Dalang Street, Longhua District, Shenzhen, China Telephone: +86-755-26648640 Fax:+86-755-26648637 Website:www.cqa-cert.com Report Template Version: V04 Report Template Revision Date: 2018-07-06 TEST REPORT The test report is effective only with both signature and specialized stamp, The result(s) shown in this report refer only to the sample(s) tested. Without written approval of CQA, this report can’t be reproduced except in full. Report No. : CQASZ20200200074E-01 Applicant: Trackerwave Private Limited Address of Applicant:4/363, Block C, 3rd Floor, OMR Road, Kanthanchavadi, Chennai, India –600 096 Equipment Under Test (EUT): EUT Name: BLE - Patient Wrist Band with Call Button Model No.: TWBL2 Brand Name: OVITAG FCC ID: 2ATS5-TWBL2 Standards:47 CFR Part 15, Subpart C Date of Receipt:2020-02-21 Date of Test:2020-02-21 to 2020-02-25 Date of Issue:2020-02-25 Test Result :PASS* *In the configuration tested, the EUT complied with the standards specified above Tested By: (Tom Chen) Reviewed By: (Aaron Ma) Approved By: (Jack Ai) Shenzhen Huaxia Testing Technology Co., Ltd Report No.: CQASZ20200200074E-01 Page: 2 of 44 1Version Revision History Of Report ReportNo.Version Description Issue Date CQASZ20200200074E-01Rev.01 Initial report 2020-02-25 Shenzhen Huaxia Testing Technology Co., Ltd Report No.: CQASZ20200200074E-01 Page: 3 of 44 2 Test Summary Test ItemTest RequirementTest methodResult Antenna Requirement 47 CFR Part 15, Subpart C Section 15.203/15.247 (c) ANSI C63.10 2013PASS AC Power Line Conducted Emission 47 CFR Part 15, Subpart C Section 15.207 ANSI C63.10 2013N/A Conducted Peak Output Power 47 CFR Part 15, Subpart C Section 15.247 (b)(3) ANSI C63.10 2013PASS 6dB Occupied Bandwidth 47 CFR Part 15, Subpart C Section 15.247 (a)(2) ANSI C63.10 2013PASS Power Spectral Density 47 CFR Part 15, Subpart C Section 15.247 (e) ANSI C63.10 2013PASS Band-edge for RF Conducted Emissions 47 CFR Part 15, Subpart C Section 15.247(d) ANSI C63.10 2013PASS RF Conducted Spurious Emissions 47 CFR Part 15, Subpart C Section 15.247(d) ANSI C63.10 2013PASS Radiated Spurious Emissions 47 CFR Part 15, Subpart C Section 15.205/15.209 ANSI C63.10 2013PASS Restricted bands around fundamental frequency (Radiated Emission) 47 CFR Part 15, Subpart C Section 15.205/15.209 ANSI C63.10 2013PASS N/A: The EUT powered by button battery, So Not Applicable Shenzhen Huaxia Testing Technology Co., Ltd Report No.: CQASZ20200200074E-01 Page: 4 of 44 3 Contents Page 1 VERSION....................................................................................................................................................................2 2 TEST SUMMARY......................................................................................................................................................3 3CONTENTS................................................................................................................................................................4 4 GENERAL INFORMATION.....................................................................................................................................5 4.1 C LIENTINFORMATION........................................................................................................................................... 5 4.2 G ENERALDESCRIPTION OFEUT...........................................................................................................................5 4.3 A DDITIONALINSTRUCTIONS..................................................................................................................................7 4.4 T ESTENVIRONMENT..............................................................................................................................................8 4.5 D ESCRIPTION OFSUPPORTUNITS..........................................................................................................................8 4.6 S TATEMENT OF THE MEASUREMENT UNCERTAINTY..............................................................................................9 4.7 T ESTLOCATION...................................................................................................................................................10 4.8 T ESTFACILITY.....................................................................................................................................................10 4.9 D EVIATION FROMSTANDARDS............................................................................................................................10 4.10 O THERINFORMATIONREQUESTED BY THECUSTOMER......................................................................................10 4.11 E QUIPMENTLIST................................................................................................................................................. 11 5 TEST RESULTS AND MEASUREMENT DATA............................................................................................... 12 5.1 A NTENNAREQUIREMENT.................................................................................................................................... 12 5.2 C ONDUCTEDPEAKOUTPUTPOWER....................................................................................................................13 5.3 6 DBOCCUPYBANDWIDTH..................................................................................................................................17 5.4 P OWERSPECTRALDENSITY................................................................................................................................20 5.5 B AND-EDGE FORRF CONDUCTEDEMISSIONS.................................................................................................... 23 5.6 S PURIOUSRF CONDUCTEDEMISSIONS....................................................................................................…
Text truncated - open the document above for the full version.
FCC ID:2ATS5-TWBL2 1 Photographs - EUT Test Setup 1.1 Radiated Spurious Emission 9KHz~30MHz: 30MHz~1GHz: FCC ID:2ATS5-TWBL2 Above 1GHz:
1F Block A of Tongsheng Technology Bldg, Huahui Rd · Shenzhen · China
| # | Rule Parts | Frequency Range | Power Output |
|---|---|---|---|
| 1 | 15C | 2.40 GHz - 2.48 GHz | 900.00 µW |

DIGITRACE ID CARD-U
Equipment Class
DTS - Digital Transmission System
DIGITRACE ID CARD
Equipment Class
DTS - Digital Transmission System
OVITAG READER GATEWAY
Equipment Class
DTS - Digital Transmission System