Text extracted from the exhibit documents filed with the FCC. Open a document above to read the original.
Radical-7 Operator's Manual www.masimo.com 1 Masimo The Radical-7 operating instructions provide the necessary information for proper operation of all models of the Radical-7 Pulse CO-Oximeter system. There may be information provided in this manual that is not relevant for your system. General knowledge of pulse oximetry and an understanding of the features and functions of the Radical-7 are prerequisites for its proper use. Do not operate the Radical-7 without completely reading and understanding the instructions in this manual. NOTICE: Purchase or possession of this instrument does not carry any express or implied license to use this instrument with replacement parts which would, alone or in combination with this instrument, fall within the scope of one of the patents relating to this instrument. Caution Federal law restricts this device to sale by or on the order of a physician. Wireless Radio FCC ID: VKF-RAD7CA, IC: 7362A-RAD7CA Masimo Corporation 40 Parker Irvine, CA 92618, USA Tel.: 949-297-7000 Fax.: 949-297-7001 www.masimo.com EU authorized representative for Masimo Corporation: MDSS GmbH Schiffgraben 41 D-30175 Hannover, Germany Medical electrical equipment with respect to electric shock, fire and mechanical hazards only in accordance with UL 60601-1/CAN/CSA C22.2 No. 601.1 Patents: www.masimo.com/patents.htm © 2012 Masimo Corporation ®, Adaptive Probe Off Detection®, APOD®, Discrete Saturation Transform®, DST®, FastSat®, FST®, Masimo®, Pulse Oximeter®, PVI®, rainbow®, SatShare®, SET®, Signal Extraction Technology®, Signal IQ®, SpCO®, SpHb®, SpMet® are federally registered trademarks of Masimo Corporation. Adaptive Threshold Alarm™, In Vivo Adjustment™, Pleth Variability Index™, Radical-7™, Rainbow Acoustic Monitoring™, rainbow Resposable™, RDS™, RRa™, RRp™,SafetyNet™, SpOC™ are trademarks of Masimo Corporation. All other trademarks and registered trademarks are property of their respective owners. The use of the trademarks PATIENT SAFETYNET and PSN is under license from University HealthSystem Consortium. www.masimo.com 3 Masimo Contents About this Manual ----------------------------------------------------------------------------------------------- 7 Product Description, Indications for Use, Contraindications, and Features ------------------------9 Product Description -----------------------------------------------------------------------------------------9 Indications for Use ---------------------------------------------------------------------------------------- 10 Contraindications ------------------------------------------------------------------------------------------ 10 Safety Information, Warnings, and Cautions -------------------------------------------------------------11 Parameter Related Safety Information, Warnings, and Cautions --------------------------------11 Device Related Safety Information, Warnings, and Cautions ------------------------------------ 15 Electrical Safety Information, Warnings, and Cautions ------------------------------------------- 18 Alarm Related Safety Information, Warnings, and Cautions------------------------------------- 20 Sensor Related Safety Information, Warnings, and Cautions------------------------------------ 21 Chapter 1: Technology Overview --------------------------------------------------------------------------- 23 Signal Extraction Technology (SET) ------------------------------------------------------------------- 23 rainbow Pulse CO-Oximetry Technology -------------------------------------------------------------- 26 rainbow Acoustic Monitoring (RAM) Technology--------------------------------------------------- 30 Chapter 2: Radical-7 Descriptions ------------------------------------------------------------------------- 33 General System Description ----------------------------------------------------------------------------- 33 Functionality of the Radical-7 --------------------------------------------------------------------------- 34 Handheld ---------------------------------------------------------------------------------------------------- 35 Standalone -------------------------------------------------------------------------------------------------- 37 Monitor Interface ------------------------------------------------------------------------------------------ 40 Chapter 3: Setup ----------------------------------------------------------------------------------------------- 41 Unpacking and Inspection ------------------------------------------------------------------------------- 41 Docking Station Power Requirements----------------------------------------------------------------- 41 Setting Up the Docking Station ------------------------------------------------------------------------ 42 Initial Battery Charging ---------------------------------------------------------------------------------- 42 Setting Up for Philips, Agilent, or HP VueLink ------------------------------------------------------ 42 Setting Up for SpaceLabs Flexport--------------------------------------------------------------------- 43 Setting Up and Using SatShare ------------------------------------------------------------------------ 44 Chapter 4: Operation------------------------------------------------------------------------------------------ 45 Using the Touchscreen and Buttons ------------------------------------------------------------------ 45 Radical-7 Contents www.masimo.com 4 Masimo Using Screenlock ------------------------------------------------------------------------------------------- 46 Using the Home Button ---------------------------------------------------------------------------------- 46 Navigating the Radical-7 --------------------------------------------------------------------------------- 47 About the Display View ----------------------------------------------------------------------------------- 48 Accessing the Main Menu -------------------------------------------------------------------------------- 55 Navigating t…
Text truncated - open the document above for the full version.
Attestation Letter 5-04-2012 To whom it may concern, The referenced Model AW-NH610 in the radio schematic is identical to Model BCM4330 in the frequency block diagram. Sincerely, Michael Clark Engineer EMC Test
Northwest EMC TCB Compliance Opinion Rev. 7/1/10 MASI0088 Information about the Applicant Grantee (Company Name) Masimo Corporation FCC Contact Person Eugene Kim Address 40 Parker City, State, Zip Irvine, CA 92618 Job Number MASI0088 Model RAD7CA FCC ID VKF-RAD7CA Agent Michael Clark Approval Type Original Equipment Class DTS Digital Transmission System NII National Information Infrastructure Rule Part 15.247 15.407 Overview Model RAD7CA is a handheld Pulse CO-Oximeter containing a radio interface to allow wireless connection of a Patient Safety Network and other wireless systems. The device operated under FCC Cfr. 15.247 and 15.407. Confidentiality requested for allowed files. Request is properly addressed and referenced. Recommendation All items have been resolved and completed to my satisfaction; therefore I recommend this application for approval. Signature Dave Tolman, TCB Committee 5/23/2012
April-18-2012 Dear Application Examiner: Masimo Corporation is submitting this application for the certification of their model RAD7CA Pulse CO-Oximeter. It is an LCD based display handheld pulse CO-Oximeter with a model RDS series docking station. It contains a radio interface to allow wireless connection to Patient Safety Net and other wireless systems. The technical reports and exhibits demonstrate compliance with FCC Part 15.247, 15.407. The applicant’s radio is compliant with FCC Part 15.247 and 15.407. FCC ID: VKF-RAD7CA Please treat the contents of this application in a confidential manner. Your efforts in reviewing this application are greatly appreciated. Regards, Michael Clark
Request for Confidentiality Letter-FCC 3-14-2012 Federal Communications Commission Authorization and Evaluation Division 7435 Oakland Mills Road Columbia, Maryland 21046 To whom it may concern, We hereby respectfully request that under the provision of 47 CFR 0.459 and 0.457(d) the documents listed below and attached with this application for certification be provided with confidential status. Schematics.pdf Operational Description.pdf Frequency Block Diagrams.pdf Any exhibit / information for which we have requested confidentiality, but which may not be accorded such treatment by the FCC, should be returned to us. The documents listed above contain trade secrets that are treated as confidential by us. Substantial competitive harm to us could result should they be made available to the public. Sincerely, Michael Clark Engineer EMC Test
External Photos Model RAD7CA
Internal Photos Radio Board – Top view Radio Board – Bottom view
Masimo Corporation Model: RAD7CA Evaluated to the following SAR Specifications: FCC 2.1093:2012 Health Safety Code 6:2009 Report No. MASI0103 Report Prepared By www.nwemc.com 1-888-EMI-CERT © 2012 Northwest EMC, Inc SAR Evaluation Report 22975 NW Evergreen Parkway Suite 400 Hillsboro, Oregon 97124 Certificate of Evaluation Last Date of Test: May 21, 2012 Masimo Corporation Model: RAD7CA Applicable Standards Test Description Specification Test Method Pass/Fail FCC OET 65C:2001 IEEE Std 1528:2003 FCC KDB 447498 D01 v04 FCC KDB 248227 D01 v01r02 FCC 2.1093:2012 FCC 15.247:2012 FCC 15.407:2012 FCC KDB 865664 Pass SAR Evaluation Health Safety Code 6:2009 RSS-102, Issue 4:2010 Pass Highest SAR Values Frequency Band (GHz) Head 1g (W/kg) Body 1g (W/kg) Limit 1g (W/kg) Exposure Environment 2.4 N/A 0.113 5 .2 & 5.8 N/A 0.511 1.6 General Population Uncontrolled Modifications made to the product See the Modifications section of this report Test Facility The measurement facility used to collect the data is located at: Northwest EMC, Inc. 22975 NW Evergreen Parkway, Suite 400 Hillsboro, OR 97124 Phone: (503) 844-4066 Fax: 844-3826 Approved By: Don Facteau, IS Manager NVLAP Lab Code: 200630-0 This report must not be used to claim product certification, approval, or endorsement by NVLAP, NIST, or any agency of the federal government of the United States of America. Product compliance is the responsibility of the client, therefore the tests and equipment modes of operation represented in this report were agreed upon by the client, prior to testing. This Report may only be duplicated in its entirety. The results of this test pertain only to the sample(s) tested. The specific description is noted in each of the individual sections of the test report supporting this certificate of test. Northwest EMC Revision History Revision 06/29/09 Barometric Pressure The recorded barometric pressure has been normalized to sea level. Revision Number Description Date Page Number 00 Accreditations and Authorizations Revision 02/21/2012 United States FCC - Designated by the FCC as a Telecommunications Certification Body (TCB). Certification chambers, Open Area Test Sites, and conducted measurement facilities are listed with the FCC. A2LA - Accredited by A2LA to ISO / IEC Guide 65 as a product certifier. This allows Northwest EMC to certify transmitters to FCC and IC specifications. NVLAP - Each laboratory is accredited by NVLAP to ISO 17025. The scope includes radio, ITE, and medical standards from around the world. See: http://www.nwemc.com/accreditations/ Canada IC - Recognized by Industry Canada as a Certification Body (CB). Certification chambers and Open Area Test Sites are filed with IC. European Union European Commission – Validated by the European Commission as a Conformity Assessment Body (CAB) under the EMC directive and as a Notified Body under the R&TTE Directive. Australia/New Zealand ACMA - Recognized by ACMA as a CAB for the acceptance of test data. Korea KCC / RRA - Recognized by KCC’s RRA as a CAB for the acceptance of test data. Japan VCCI - Associate Member of the VCCI. Conducted and radiated measurement facilities are registered. Taiwan BSMI – Recognized by BSMI as a CAB for the acceptance of test data. NCC - Recognized by NCC as a CAB for the acceptance of test data. Singapore IDA – Recognized by IDA as a CAB for the acceptance of test data. Hong Kong OFTA – Recognized by OFTA as a CAB for the acceptance of test data. Vietnam MIC – Recognized by MIC as a CAB for the acceptance of test data. Russia GOST – Accredited by Certinform VNIINMASH, CERTINFO, SAMTES, and Federal CHEC to perform EMC and Hygienic testing for Information Technology products to GOST standards. Locations Revision 2/23/12 Oregon Labs EV01-EV12 22975 NW Evergreen Pkwy, #400 Hillsboro, OR 97124 (503) 844-4066 California Labs OC01-OC13 41 Tesla Irvine, CA 92618 (949) 861-8918 New York Labs WA01-WA04 4939 Jordan Rd. Elbridge, NY 13060 (315) 685-0796 Minnesota Labs MN01-MN08 9349 W Broadway Ave. Brooklyn Park, MN 55445 (763) 425-2281 Washington Labs SU01-SU07 14128 339 th Ave. SE Sultan, WA 98294 (360) 793-8675 VCCI C-1071, R-1025, G-84, C-2687, T-1658, R-2318 R-1943, G-85, C-2766, T-1659, G-548 R-3125, G-86, G-141, C-3464, T-1634 R-871, G-83, C-3265, T-1511 Industry Canada 2834D-1, 2834D-2 2834B-1, 2834B-2, 2834B-3 2834E-1 2834C-1 Northwest EMC Product Description Rev 11/17/06 Party Requesting the Test Company Name: Masimo Corporation Address: 40 Parker City, State, Zip: Irvine, CA 92618 Test Requested By: Michael Clark Model: RAD7CA First Date of Test: May 8, 2012 Last Date of Test: May 21, 2012 Receipt Date of Samples: May 8, 2012 Equipment Design Stage: Production Equipment Condition: No Damage Information Provided by the Party Requesting the Test Functional Description of the EUT (Equipment Under Test) The EUT is the Masimo Model RAD7CA Pulse CO-Oximeter containing an 802.11 a/b/g radio. The RAD7CA is a handheld pulse CO-Oximeter with a LCD display. The radio interface allows wireless connection to Patient Safety Net and other wireless systems. The RAD7CA is only used in a clinic or hospital environment. The hardware design limits wireless transmission to patient telemetry data only. No other type of data transmission is possible. Kirby Dotson, Manager Instrument Hardware at Masimo, attests that the maximum duty cycle is about 6%: “ The software sends 120bytes @ 62.5 Hz for 7500 bytes per second or 60K bits/sec. Worst duty cycle will be when in 1.1Mbit/sec mode which is 60K/1.1M for a duty cycle of about 6%.” The RAD7CA uses two transmitting chip antennas, one each side of the unit. Kirby Dotson of Masimo also attests that the RAD7CA uses “"Switched Antenna Diversity" as defined by FCC KDB 248227, Foot note 1 on page 2: "Switched diversity used in 802.11 a/b/g devices typically allows transmission from one of two antennas in a random manner, according to the received signal quality of each antenna and the diversity algorithm implemente…
Text truncated - open the document above for the full version.
Masimo Corporation RAD7CA Report #: MASI0095.1 Report Prepared By Northwest EMC Inc. NORTHWEST EMC – (888) 364-2378 – www.nwemc.com California – Minnesota – Oregon – New York – Washington 22975 NW Evergreen Parkway Suite 400 Hillsboro, Oregon 97124 Certificate of Test Last Date of Test: April 27, 2012 Masimo Corporation Model: RAD7CA Emissions Test Description Specification Test Method Pass/Fail Emission Bandwidth FCC 15.407:2012 ANSI C63.10:2009 Pass Peak Transmit Power FCC 15.407:2012 ANSI C63.10:2009 Pass Peak Power Spectral Density FCC 15.407:2012 ANSI C63.10:2009 Pass Peak Excursion FCC 15.407:2012 ANSI C63.10:2009 Pass Frequency Stability FCC 15.407:2012 ANSI C63.10:2009 Pass Spurious Radiated Emissions FCC 15.407:2012 ANSI C63.10:2009 Pass AC Powerline Conducted Emissions FCC 15.407:2012 ANSI C63.10:2009 Pass Deviations From Test Standards None Approved By: Tim O'Shea, Operations Manager NVLAP Lab Code: 200676-0 Test Facility The measurement facility used to collect the data is located at: Northwest EMC, Inc. 41 Tesla Ave. Irvine, CA 92618 Phone: (503) 844-4066 Fax: 844-3826 This site has been fully described in a report filed with and accepted by the FCC (Federal Communications Commission) and Industry Canada (Site filing #2834B-1). This report must not be used to claim product certification, approval, or endorsement by NVLAP, NIST, or any agency of the federal government of the United States of America. Product compliance is the responsibility of the client, therefore the tests and equipment modes of operation represented in this report were agreed upon by the client, prior to testing. This Report may only be duplicated in its entirety. The results of this test pertain only to the sample(s) tested. The specific description is noted in each of the individual sections of the test report supporting this certificate of test. Revision History Revision 09/01/11 Revision Number Description Date Page Number 00 None Accreditations and Authorizations Revision 02/21/2012 United States FCC - Designated by the FCC as a Telecommunications Certification Body (TCB). Certification chambers, Open Area Test Sites, and conducted measurement facilities are listed with the FCC. A2LA - Accredited by A2LA to ISO / IEC Guide 65 as a product certifier. This allows Northwest EMC to certify transmitters to FCC and IC specifications. NVLAP - Each laboratory is accredited by NVLAP to ISO 17025. The scope includes radio, ITE, and medical standards from around the world. See: http://www.nwemc.com/accreditations/ Canada IC - Recognized by Industry Canada as a Certification Body (CB). Certification chambers and Open Area Test Sites are filed with IC. European Union European Commission – Validated by the European Commission as a Conformity Assessment Body (CAB) under the EMC directive and as a Notified Body under the R&TTE Directive. Australia/New Zealand ACMA - Recognized by ACMA as a CAB for the acceptance of test data. Korea KCC / RRA - Recognized by KCC’s RRA as a CAB for the acceptance of test data. Japan VCCI - Associate Member of the VCCI. Conducted and radiated measurement facilities are registered. Taiwan BSMI – Recognized by BSMI as a CAB for the acceptance of test data. NCC - Recognized by NCC as a CAB for the acceptance of test data. Singapore IDA – Recognized by IDA as a CAB for the acceptance of test data. Hong Kong OFTA – Recognized by OFTA as a CAB for the acceptance of test data. Vietnam MIC – Recognized by MIC as a CAB for the acceptance of test data. Russia GOST – Accredited by Certinform VNIINMASH, CERTINFO, SAMTES, and Federal CHEC to perform EMC and Hygienic testing for Information Technology products to GOST standards. Locations Revision 2/23/12 Oregon Labs EV01-EV12 22975 NW Evergreen Pkwy, #400 Hillsboro, OR 97124 (503) 844-4066 California Labs OC01-OC13 41 Tesla Irvine, CA 92618 (949) 861-8918 New York Labs WA01-WA04 4939 Jordan Rd. Elbridge, NY 13060 (315) 685-0796 Minnesota Labs MN01-MN08 9349 W Broadway Ave. Brooklyn Park, MN 55445 (763) 425-2281 Washington Labs SU01-SU07 14128 339 th Ave. SE Sultan, WA 98294 (360) 793-8675 VCCI C-1071, R-1025, G-84, C-2687, T-1658, R-2318 R-1943, G-85, C-2766, T-1659, G-548 R-3125, G-86, G-141, C-3464, T-1634 R-871, G-83, C-3265, T-1511 Industry Canada 2834D-1, 2834D-2 2834B-1, 2834B-2, 2834B-3 2834E-1 2834C-1 Product Description Revision 09/01/11 Client and Equipment Under Test (EUT) Information Company Name: Masimo Corporation Address: 40 Parker City, State, Zip: Irvine, CA 92618 Test Requested By: Michael Clark Model: RAD7CA First Date of Test: April 18, 2012 Last Date of Test: April 27, 2012 Receipt Date of Samples: April 18, 2012 Equipment Design Stage: Production Equipment Condition: No Damage Information Provided by the Party Requesting the Test Functional Description of the EUT (Equipment Under Test): 802.11a/b/g radio Testing Objective: Seeking to demonstrate compliance under FCC 15.407 for operation in the 5.2 GHz band only Configurations Revision 09/01/11 Configuration 1 MASI0095 Software/Firmware Running during test Description Version Tera Term 4.73 (Linux) base E 0.0.1.6 EUT Description Manufacturer Model/Part Number Serial Number Pulse Co-Oximeter Masimo Corporation RAD7CA 34996 Rev C Peripherals in test setup boundary Description Manufacturer Model/Part Number Serial Number Remote Laptop Hewlett Packard Compaq 6515b CNU7300W4L Cables Cable Type Shield Length (m) Ferrite Connection 1 Connection 2 AC Cable No 1.8m No RAD7CA AC Mains USB Cable No 1.0m No RAD7CA Remote Laptop PA = Cable is permanently attached to the device. Shielding and/or presence of ferrite may be unknown. Configuration 2 MASI0095 Software/Firmware Running during test Description Version Tera Term 4.73 (Linux) base E 0.0.1.6 EUT Description Manufacturer Model/Part Number Serial Number Pulse Co-Oximeter Masimo Corporation RAD7CA 34996 Rev C Peripherals in test setup boundary Description Manufacturer Model/Part Number Serial Number Remote Laptop Hewlett Packard Compaq 6…
Text truncated - open the document above for the full version.
Masimo Corporation RAD7CA Report #: MASI0095 Report Prepared By Northwest EMC Inc. NORTHWEST EMC – (888) 364-2378 – www.nwemc.com California – Minnesota – Oregon – New York – Washington 22975 NW Evergreen Parkway Suite 400 Hillsboro, Oregon 97124 Certificate of Test Last Date of Test: April 27, 2012 Masimo Corporation Model: RAD7CA Emissions Test Description Specification Test Method Pass/Fail Occupied Bandwidth FCC 15.247:2012 ANSI C63.10:2009 Pass Output Power FCC 15.247:2012 ANSI C63.10:2009 Pass Band Edge Compliance FCC 15.247:2012 ANSI C63.10:2009 Pass Spurious Conducted Emissions FCC 15.247:2012 ANSI C63.10:2009 Pass Power Spectral Density FCC 15.247:2012 ANSI C63.10:2009 Pass Spurious Radiated Emissions FCC 15.247:2012 ANSI C63.10:2009 Pass AC Powerline Conducted Emissions FCC 15.207:2012 ANSI C63.10:2009 Pass Deviations From Test Standards None Approved By: Tim O'Shea, Operations Manager NVLAP Lab Code: 200676-0 Test Facility The measurement facility used to collect the data is located at: Northwest EMC, Inc. 41 Tesla Ave. Irvine, CA 92618 Phone: (503) 844-4066 Fax: 844-3826 This site has been fully described in a report filed with and accepted by the FCC (Federal Communications Commission) and Industry Canada (Site filing #2834B-1). This report must not be used to claim product certification, approval, or endorsement by NVLAP, NIST, or any agency of the federal government of the United States of America. Product compliance is the responsibility of the client, therefore the tests and equipment modes of operation represented in this report were agreed upon by the client, prior to testing. This Report may only be duplicated in its entirety. The results of this test pertain only to the sample(s) tested. The specific description is noted in each of the individual sections of the test report supporting this certificate of test. Revision History Revision 09/01/11 Revision Number Description Date Page Number 00 None Accreditations and Authorizations Revision 02/21/2012 United States FCC - Designated by the FCC as a Telecommunications Certification Body (TCB). Certification chambers, Open Area Test Sites, and conducted measurement facilities are listed with the FCC. A2LA - Accredited by A2LA to ISO / IEC Guide 65 as a product certifier. This allows Northwest EMC to certify transmitters to FCC and IC specifications. NVLAP - Each laboratory is accredited by NVLAP to ISO 17025. The scope includes radio, ITE, and medical standards from around the world. See: http://www.nwemc.com/accreditations/ Canada IC - Recognized by Industry Canada as a Certification Body (CB). Certification chambers and Open Area Test Sites are filed with IC. European Union European Commission – Validated by the European Commission as a Conformity Assessment Body (CAB) under the EMC directive and as a Notified Body under the R&TTE Directive. Australia/New Zealand ACMA - Recognized by ACMA as a CAB for the acceptance of test data. Korea KCC / RRA - Recognized by KCC’s RRA as a CAB for the acceptance of test data. Japan VCCI - Associate Member of the VCCI. Conducted and radiated measurement facilities are registered. Taiwan BSMI – Recognized by BSMI as a CAB for the acceptance of test data. NCC - Recognized by NCC as a CAB for the acceptance of test data. Singapore IDA – Recognized by IDA as a CAB for the acceptance of test data. Hong Kong OFTA – Recognized by OFTA as a CAB for the acceptance of test data. Vietnam MIC – Recognized by MIC as a CAB for the acceptance of test data. Russia GOST – Accredited by Certinform VNIINMASH, CERTINFO, SAMTES, and Federal CHEC to perform EMC and Hygienic testing for Information Technology products to GOST standards. Locations Revision 2/23/12 Oregon Labs EV01-EV12 22975 NW Evergreen Pkwy, #400 Hillsboro, OR 97124 (503) 844-4066 California Labs OC01-OC13 41 Tesla Irvine, CA 92618 (949) 861-8918 New York Labs WA01-WA04 4939 Jordan Rd. Elbridge, NY 13060 (315) 685-0796 Minnesota Labs MN01-MN08 9349 W Broadway Ave. Brooklyn Park, MN 55445 (763) 425-2281 Washington Labs SU01-SU07 14128 339 th Ave. SE Sultan, WA 98294 (360) 793-8675 VCCI C-1071, R-1025, G-84, C-2687, T-1658, R-2318 R-1943, G-85, C-2766, T-1659, G-548 R-3125, G-86, G-141, C-3464, T-1634 R-871, G-83, C-3265, T-1511 Industry Canada 2834D-1, 2834D-2 2834B-1, 2834B-2, 2834B-3 2834E-1 2834C-1 Product Description Revision 09/01/11 Client and Equipment Under Test (EUT) Information Company Name: Masimo Corporation Address: 40 Parker City, State, Zip: Irvine, CA 92618 Test Requested By: Michael Clark Model: RAD7CA First Date of Test: April 18, 2012 Last Date of Test: April 27, 2012 Receipt Date of Samples: April 18, 2012 Equipment Design Stage: Production Equipment Condition: No Damage Information Provided by the Party Requesting the Test Functional Description of the EUT (Equipment Under Test): 802.11a/b/g radio Testing Objective: To demonstrate compliance under FCC 15.247 for operation in the 2.4 and 5.8 GHz bands Configurations Revision 09/01/11 Configuration 1 MASI0095 Software/Firmware Running during test Description Version Tera Term 4.73 (Linux) base E 0.0.1.6 EUT Description Manufacturer Model/Part Number Serial Number Pulse Co-Oximeter Masimo Corporation RAD7CA 34996 Rev C Peripherals in test setup boundary Description Manufacturer Model/Part Number Serial Number Remote Laptop Hewlett Packard Compaq 6515b CNU7300W4L Cables Cable Type Shield Length (m) Ferrite Connection 1 Connection 2 AC Cable No 1.8m No RAD7CA AC Mains USB Cable No 1.0m No RAD7CA Remote Laptop PA = Cable is permanently attached to the device. Shielding and/or presence of ferrite may be unknown. Configuration 2 MASI0095 Software/Firmware Running during test Description Version Tera Term 4.73 (Linux) base E 0.0.1.6 EUT Description Manufacturer Model/Part Number Serial Number Pulse Co-Oximeter Masimo Corporation RAD7CA 34996 Rev C Peripherals in test setup boundary Description Manufacturer Model/Part Number Serial Number Remote Laptop Hewlett Packard Compaq 6515…
Text truncated - open the document above for the full version.
NORTHWEST XMit 2011.01.18 EMC SAR TEST SETUP PHOTOS - BODY Front Front NORTHWEST XMit 2011.01.18 EMC SAR TEST SETUP PHOTOS - BODY Back Back NORTHWEST XMit 2011.01.18 EMC SAR TEST SETUP PHOTOS - BODY Left Side Left Side NORTHWEST XMit 2011.01.18 EMC SAR TEST SETUP PHOTOS - BODY Right Side Right Side
Emission Bandwidth XMit 2012.04.06 NORTHWEST Emission Bandwidth PsaTx 2012.01.20 EMC IMG_0136.JPG IMG_0138.JPG Emission Bandwidth Peak Transmit Power XMit 2012.04.06 NORTHWEST Peak Transmit Power PsaTx 2012.01.20 EMC IMG_0136.JPG IMG_0138.JPG Peak Transmit Power Peak Transmit Power1 XMit 2011.12.23 NORTHWEST Peak Transmit Power PsaTx 2012.01.20 IMG_0136.JPG IMG_0138.JPG Peak Transmit Power1 Peak Power Spectral Density XMit 2012.04.06 NORTHWEST Peak Power Spectral PsaTx 2012.01.20 EMC IMG_0136.JPG IMG_0138.JPG Peak Power Spectral Density Peak Excursion XMit 2012.04.06 NORTHWEST Peak Excursion PsaTx 2012.01.20 EMC IMG_0136.JPG IMG_0138.JPG Peak Excursion Frequency Stability XMit 2011.12.23 NORTHWEST Frequency Stability PsaTx 2012.01.20 SAM_0138.JPG SAM_0139.JPG Frequency Stability photo SAM_0001.JPG SAM_0002.JPG SPURIOUS RADIATED EMISSIONS PSA-ESCI 2012.03.23 photo SAM_0003.JPG SAM_0208.JPG SPURIOUS RADIATED EMISSIONS PSA-ESCI 2012.03.23 photo SAM_0209.JPG SAM_0210.JPG SPURIOUS RADIATED EMISSIONS PSA-ESCI 2012.03.23 photo CE (1).JPG CE (2).JPG AC POWERLINE CONDUCTED EMISSIONS PSA-ESCI 2012.03.08
Occupied Bandwidth XMit 2012.04.06 NORTHWEST Occupied Bandwidth PsaTx 2012.01.25 EMC IMG_0136.JPG IMG_0138.JPG Occupied Bandwidth Output Power XMit 2012.04.06 NORTHWEST Output Power PsaTx 2012.01.25 EMC IMG_0136.JPG IMG_0138.JPG Output Power Output Power XMit 2012.04.06 NORTHWEST Output Power PsaTx 2012.01.25 EMC IMG_0136.JPG IMG_0138.JPG Output Power Band Edge Compliance XMit 2012.04.06 NORTHWEST Band Edge Complian PsaTx 2012.01.25 EMC IMG_0136.JPG IMG_0138.JPG Band Edge Compliance Spurious Conducted Emissions XMit 2012.04.06 NORTHWEST Spurious Conducted PsaTx 2012.01.25 EMC IMG_0136.JPG IMG_0138.JPG Spurious Conducted Emissions Power Spectral Density XMit 2012.04.06 NORTHWEST Power Spectral Densi PsaTx 2012.01.25 EMC IMG_0136.JPG IMG_0138.JPG Power Spectral Density photo SAM_0001.JPG SAM_0002.JPG SPURIOUS RADIATED EMISSIONS PSA-ESCI 2012.03.23 photo SAM_0003.JPG SAM_0208.JPG SPURIOUS RADIATED EMISSIONS PSA-ESCI 2012.03.23 photo SAM_0209.JPG SAM_0210.JPG SPURIOUS RADIATED EMISSIONS PSA-ESCI 2012.03.23 photo CE (1).JPG CE (2).JPG AC POWERLINE CONDUCTED EMISSIONS PSA-ESCI 2012.03.23
| # | Rule Parts | Frequency Range | Power Output |
|---|---|---|---|
| 1 | 15E | 5.18 GHz - 5.24 GHz | 15.70 mW |
Masimo Freedom Watch
Equipment Class
NII - Unlicensed National Information Infrastructure TX
Stork
Equipment Class
DTS - Digital Transmission System
Baby Sensor
Equipment Class
DTS - Digital Transmission System
Masimo W1
Equipment Class
DTS - Digital Transmission System
MightySat
Equipment Class
DTS - Digital Transmission System